| Line 1: | Line 1: | ||
We used | We used: | ||
=ADC 1D= | -detector A to calibrate all the TDCs; | ||
-LeCroy octal discriminator to produce standard negative pulses; | |||
-passive delay line to calibrate TDCs in the region < 100 ns; | |||
-delay amplifier TENNELEC TC215 to calibrate TDCs in the region > 250 ns; | |||
Range of all three TDCs was set to 50nsx100. | |||
=ORTEC 567-02 and ADC 1D= | |||
[[Image:567_02.jpg |500px]] | [[Image:567_02.jpg |500px]] | ||
=ADC 1A= | =ORTEC 567-01 and ADC 1A= | ||
[[Image:567_01.jpg |500px]] | [[Image:567_01.jpg |500px]] | ||
=ADC 1C= | =ORTEC 566-01 and ADC 1C= | ||
[[Image:566_01.jpg |500px]] | [[Image:566_01.jpg |500px]] | ||
Revision as of 03:29, 7 November 2008
We used:
-detector A to calibrate all the TDCs;
-LeCroy octal discriminator to produce standard negative pulses;
-passive delay line to calibrate TDCs in the region < 100 ns;
-delay amplifier TENNELEC TC215 to calibrate TDCs in the region > 250 ns;
Range of all three TDCs was set to 50nsx100.