Difference between revisions of "Neutron Time of Flight Test"
Jump to navigation
Jump to search
| Line 1: | Line 1: | ||
| − | + | ==Verifying Neutron Detection== | |
| − | + | ||
| + | ---- | ||
| + | This test included one detector and a collimated Cf-252 source. The detector was placed 16 inches away from the source collimator. Tests were performed with various thicknesses of poly in between the two. Each count was 5 minutes long and three counts were taken for each test. Detector D was used at 1618 volts through channel 3 of LeCroy power supply and a threshold of 77mV set through the blue faced CFD. | ||
| + | |||
| + | ---- | ||
| + | {| border="1" cellpadding="20" cellspacing="0" | ||
| + | |||
| + | |- | ||
| + | |setup | ||
| + | |1 poly block (4") | ||
| + | |2 poly blocks (8") | ||
| + | |3 ply blocks (12") | ||
| + | |4 poly blocks (16") | ||
| + | |no poly | ||
| + | |background | ||
| + | |background with detector shielded with 4" poly | ||
| + | |- | ||
| + | |test 1 counts | ||
| + | |7180 | ||
| + | |5142 | ||
| + | |3945 | ||
| + | |3235 | ||
| + | |11793 | ||
| + | |1620 | ||
| + | |1314 | ||
| + | |- | ||
| + | |test 2 counts | ||
| + | |7291 | ||
| + | |5213 | ||
| + | |3994 | ||
| + | |3084 | ||
| + | |11694 | ||
| + | |1648 | ||
| + | |1336 | ||
| + | |- | ||
| + | |test 3 counts | ||
| + | |7319 | ||
| + | |5205 | ||
| + | |4014 | ||
| + | |3150 | ||
| + | |11872 | ||
| + | |1629 | ||
| + | |1352 | ||
| + | |- | ||
| + | |averages | ||
| + | |7263 | ||
| + | |5187 | ||
| + | |3984 | ||
| + | |3156 | ||
| + | |11786 | ||
| + | |1632 | ||
| + | |1334 | ||
| + | |-|} | ||
| + | |||
| + | ---- | ||
| + | ==Test for Time of Flight== | ||
Revision as of 15:16, 19 September 2008
Verifying Neutron Detection
This test included one detector and a collimated Cf-252 source. The detector was placed 16 inches away from the source collimator. Tests were performed with various thicknesses of poly in between the two. Each count was 5 minutes long and three counts were taken for each test. Detector D was used at 1618 volts through channel 3 of LeCroy power supply and a threshold of 77mV set through the blue faced CFD.
Test for Time of Flight
| setup | 1 poly block (4") | 2 poly blocks (8") | 3 ply blocks (12") | 4 poly blocks (16") | no poly | background | background with detector shielded with 4" poly |
| test 1 counts | 7180 | 5142 | 3945 | 3235 | 11793 | 1620 | 1314 |
| test 2 counts | 7291 | 5213 | 3994 | 3084 | 11694 | 1648 | 1336 |
| test 3 counts | 7319 | 5205 | 4014 | 3150 | 11872 | 1629 | 1352 |
| averages | 7263 | 5187 | 3984 | 3156 | 11786 | 1632 | 1334 |