Line 4: |
Line 4: |
| CH_High_And_Unknown_Activity_Source.png|<span style="color:#FF0000">'''Red:'''</span> Log plot of scalar measurements using the high activity source with <small><math>Ae^Bx+C</math></small> fit. <span style="color:#0000ff">'''Blue:'''</span> Log plot of scalar measurements using the undetermined activity source with <small><math>Ae^Bx+C</math></small> | | CH_High_And_Unknown_Activity_Source.png|<span style="color:#FF0000">'''Red:'''</span> Log plot of scalar measurements using the high activity source with <small><math>Ae^Bx+C</math></small> fit. <span style="color:#0000ff">'''Blue:'''</span> Log plot of scalar measurements using the undetermined activity source with <small><math>Ae^Bx+C</math></small> |
| | | |
− | CH NoPolyScaled Borated Poly Atten He3 500uCi ADC 05192022.png|<span style="color:#800080">'''Purple:'''</span> Input signal from spec amp. <span style="color:#ffff00">'''Yellow:'''</span> CAEN Dual Timer using Lecroy Disc as input. <span style="color:#00ffff">'''Aqua:'''</span> Lecroy Disc Output. <span style="color:#00ff00">'''Lime:'''</span>: CAEN Update Mode Disc Output. | + | CH NoPolyScaled Borated Poly Atten He3 500uCi ADC 05192022.png|Energy spectrum of He3 tube using peak sensing ADC for the varying thicknesses of borated polyethylene. Note that the 0" of poly (open source) is scaled by <small><math>\frac{1}{3}</math></small>. |
| | | |
| PS_Sim_He3_Energy_Spectrum_Cf252_05112022.png|<span style="color:#800080">'''Purple:'''</span> Input signal from spec amp. <span style="color:#ffff00">'''Yellow:'''</span> CAEN Dual Timer using CAEN Disc as input. <span style="color:#00ffff">'''Aqua:'''</span> Lecroy Disc Output. <span style="color:#00ff00">'''Lime:'''</span>: CAEN Update Mode Disc Output. | | PS_Sim_He3_Energy_Spectrum_Cf252_05112022.png|<span style="color:#800080">'''Purple:'''</span> Input signal from spec amp. <span style="color:#ffff00">'''Yellow:'''</span> CAEN Dual Timer using CAEN Disc as input. <span style="color:#00ffff">'''Aqua:'''</span> Lecroy Disc Output. <span style="color:#00ff00">'''Lime:'''</span>: CAEN Update Mode Disc Output. |
Revision as of 16:59, 14 June 2022
Results
Red: Log plot of scalar measurements using the high activity source with [math]Ae^Bx+C[/math] fit. Blue: Log plot of scalar measurements using the undetermined activity source with [math]Ae^Bx+C[/math]
Energy spectrum of He3 tube using peak sensing ADC for the varying thicknesses of borated polyethylene. Note that the 0" of poly (open source) is scaled by [math]\frac{1}{3}[/math].
Purple: Input signal from spec amp. Yellow: CAEN Dual Timer using CAEN Disc as input. Aqua: Lecroy Disc Output. Lime:: CAEN Update Mode Disc Output.
Attenuation Experiment using New Modules in Electronics Chain and Inside Borated Poly Box
Electronics Chain (Poly Box)
Include schematic of electronics chain
Components
Each component of electronics chain with make/model, function, and brief description of why it's present
Ortec 142PC Pre-amp
Pre-amp for the He3 tube
CAEN Quad Linear Fan In - Fan Out Model N25 Tag #15687
- The fan in/out is used to duplicate the pulse to run the spec amp and watch the signal on the scope for any anomalies.
- The fan in/out is also used to invert the signal from the spec amp to send a negative pulse to the discriminator to trigger the scalar.
Ortec Spec Amp Model 672
- Manual Pole Zero adjustment
LeCroy Discriminator Model 821 Serial #B16303
- Pulse width as measured on scope: ~43ns
Background Measurements
1250V Background Measurements |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec) |
Count Rate (Hz) |
Uncertainty
|
1 |
Detector on 3D printed stands at +1250V |
255 |
No |
|
100 |
|
|
2 |
Detector on 3D printed stands at +1250V |
255 |
No |
|
100 |
|
|
3 |
Detector on 3D printed stands at +1250V |
255 |
No |
|
100 |
|
|
4 |
Detector on 3D printed stands at +1250V |
255 |
No |
|
100 |
|
|
5 |
Detector on 3D printed stands at +1250V |
255 |
No |
|
100 |
|
|
High Activity Source Poly Box Attenuation Measurements
472[math]\mu[/math]Ci at time of measurement.
Open Source
1250V source in |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec)
|
1 |
Detector on poly at +1250V |
80 |
Yes |
34456 |
10
|
2 |
Detector on poly at +1250V |
80 |
Yes |
34920 |
10
|
3 |
Detector on poly at +1250V |
80 |
Yes |
34422 |
10
|
4 |
Detector on poly at +1250V |
80 |
Yes |
34007 |
10
|
5 |
Detector on poly at +1250V |
80 |
Yes |
34420 |
10
|
2" Green poly (one brick) in Box
1250V source in |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec)
|
1 |
Detector on poly at +1250V |
80 |
Yes |
7538 |
10
|
2 |
Detector on poly at +1250V |
80 |
Yes |
7515 |
10
|
3 |
Detector on poly at +1250V |
80 |
Yes |
7613 |
10
|
4 |
Detector on poly at +1250V |
80 |
Yes |
7566 |
10
|
5 |
Detector on poly at +1250V |
80 |
Yes |
7510 |
10
|
4" Green poly (two bricks) in Box
1250V source in |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec)
|
1 |
Detector on poly at +1250V |
80 |
Yes |
4758 |
10
|
2 |
Detector on poly at +1250V |
80 |
Yes |
4798 |
10
|
3 |
Detector on poly at +1250V |
80 |
Yes |
4722 |
10
|
4 |
Detector on poly at +1250V |
80 |
Yes |
4859 |
10
|
5 |
Detector on poly at +1250V |
80 |
Yes |
4672 |
10
|
6" Green poly (three bricks) in Box
1250V source in |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec)
|
1 |
Detector on poly at +1250V |
80 |
Yes |
3664 |
10
|
2 |
Detector on poly at +1250V |
80 |
Yes |
3627 |
10
|
3 |
Detector on poly at +1250V |
80 |
Yes |
3699 |
10
|
4 |
Detector on poly at +1250V |
80 |
Yes |
3688 |
10
|
5 |
Detector on poly at +1250V |
80 |
Yes |
3500 |
10
|
8" Green poly (four bricks) in Box
1250V source in |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec)
|
1 |
Detector on poly at +1250V |
80 |
Yes |
3140 |
10
|
2 |
Detector on poly at +1250V |
80 |
Yes |
3064 |
10
|
3 |
Detector on poly at +1250V |
80 |
Yes |
3030 |
10
|
4 |
Detector on poly at +1250V |
80 |
Yes |
3095 |
10
|
5 |
Detector on poly at +1250V |
80 |
Yes |
3113 |
10
|
Undetermined Activity Source Poly Box Attenuation Measurements
Open Source
1250V source in |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec)
|
1 |
Detector on poly at +1250V |
80 |
Yes |
3873 |
100
|
2 |
Detector on poly at +1250V |
80 |
Yes |
3909 |
100
|
3 |
Detector on poly at +1250V |
80 |
Yes |
3859 |
100
|
4 |
Detector on poly at +1250V |
80 |
Yes |
3944 |
100
|
5 |
Detector on poly at +1250V |
80 |
Yes |
3812 |
100
|
2" Green poly (one brick) in Box
1250V source in |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec)
|
1 |
Detector on poly at +1250V |
80 |
Yes |
764 |
100
|
2 |
Detector on poly at +1250V |
80 |
Yes |
819 |
100
|
3 |
Detector on poly at +1250V |
80 |
Yes |
818 |
100
|
4 |
Detector on poly at +1250V |
80 |
Yes |
815 |
100
|
5 |
Detector on poly at +1250V |
80 |
Yes |
913 |
100
|
4" Green poly (two bricks) in Box
1250V source in |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec)
|
1 |
Detector on poly at +1250V |
80 |
Yes |
486 |
100
|
2 |
Detector on poly at +1250V |
80 |
Yes |
509 |
100
|
3 |
Detector on poly at +1250V |
80 |
Yes |
531 |
100
|
4 |
Detector on poly at +1250V |
80 |
Yes |
544 |
100
|
5 |
Detector on poly at +1250V |
80 |
Yes |
530 |
100
|
6" Green poly (three bricks) in Box
1250V source in |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec)
|
1 |
Detector on poly at +1250V |
80 |
Yes |
415 |
100
|
2 |
Detector on poly at +1250V |
80 |
Yes |
373 |
100
|
3 |
Detector on poly at +1250V |
80 |
Yes |
380 |
100
|
4 |
Detector on poly at +1250V |
80 |
Yes |
397 |
100
|
5 |
Detector on poly at +1250V |
80 |
Yes |
419 |
100
|
8" Green poly (four bricks) in Box
1250V source in |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec)
|
1 |
Detector on poly at +1250V |
80 |
Yes |
371 |
100
|
2 |
Detector on poly at +1250V |
80 |
Yes |
396 |
100
|
3 |
Detector on poly at +1250V |
80 |
Yes |
351 |
100
|
4 |
Detector on poly at +1250V |
80 |
Yes |
366 |
100
|
5 |
Detector on poly at +1250V |
80 |
Yes |
394 |
100
|
Lead Attenuation Measurements
The purpose of these measurements was to determine if photon build up in the He3 tube was a significant effect to consider when trying to determine the activity of the Unknown Source
4/5/22
Lead Open Source
1250V source in |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec)
|
1 |
Detector on poly at +1250V |
80 |
Yes |
3873 |
100
|
2 |
Detector on poly at +1250V |
80 |
Yes |
3909 |
100
|
3 |
Detector on poly at +1250V |
80 |
Yes |
3859 |
100
|
4 |
Detector on poly at +1250V |
80 |
Yes |
3944 |
100
|
5 |
Detector on poly at +1250V |
80 |
Yes |
3812 |
100
|
1" Lead
1250V source in |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec)
|
1 |
Detector on poly at +1250V |
80 |
Yes |
4152 |
100
|
2 |
Detector on poly at +1250V |
80 |
Yes |
3945 |
100
|
3 |
Detector on poly at +1250V |
80 |
Yes |
4141 |
100
|
4 |
Detector on poly at +1250V |
80 |
Yes |
3974 |
100
|
5 |
Detector on poly at +1250V |
80 |
Yes |
4154 |
100
|
2" Lead
1250V source in |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec)
|
1 |
Detector on poly at +1250V |
80 |
Yes |
3509 |
100
|
2 |
Detector on poly at +1250V |
80 |
Yes |
3472 |
100
|
3 |
Detector on poly at +1250V |
80 |
Yes |
3387 |
100
|
4 |
Detector on poly at +1250V |
80 |
Yes |
3488 |
100
|
5 |
Detector on poly at +1250V |
80 |
Yes |
3463 |
100
|
3" Lead
1250V source in |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec)
|
1 |
Detector on poly at +1250V |
80 |
Yes |
2995 |
100
|
2 |
Detector on poly at +1250V |
80 |
Yes |
2921 |
100
|
3 |
Detector on poly at +1250V |
80 |
Yes |
2956 |
100
|
4 |
Detector on poly at +1250V |
80 |
Yes |
2916 |
100
|
5 |
Detector on poly at +1250V |
80 |
Yes |
3015 |
100
|
4" Lead
1250V source in |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec)
|
1 |
Detector on poly at +1250V |
80 |
Yes |
2630 |
100
|
2 |
Detector on poly at +1250V |
80 |
Yes |
2601 |
100
|
3 |
Detector on poly at +1250V |
80 |
Yes |
2558 |
100
|
4 |
Detector on poly at +1250V |
80 |
Yes |
2629 |
100
|
5 |
Detector on poly at +1250V |
80 |
Yes |
2650 |
100
|
Lowest Activity Source Poly Box Attenuation Measurements
Approximately 1.15[math]\mu[/math]Ci at time of measurement, according to ISU Radiation Safety
Open Source
1250V source in |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec)
|
1 |
Detector on poly at +1250V |
80 |
Yes |
31 |
100
|
2 |
Detector on poly at +1250V |
80 |
Yes |
32 |
100
|
3 |
Detector on poly at +1250V |
80 |
Yes |
26 |
100
|
4 |
Detector on poly at +1250V |
80 |
Yes |
23 |
100
|
5 |
Detector on poly at +1250V |
80 |
Yes |
29 |
100
|
2" Green poly (one brick) in Box
1250V source in |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec)
|
1 |
Detector on poly at +1250V |
80 |
Yes |
18 |
100
|
2 |
Detector on poly at +1250V |
80 |
Yes |
27 |
100
|
3 |
Detector on poly at +1250V |
80 |
Yes |
21 |
100
|
4 |
Detector on poly at +1250V |
80 |
Yes |
21 |
100
|
5 |
Detector on poly at +1250V |
80 |
Yes |
22 |
100
|
4" Green poly (two bricks) in Box
1250V source in |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec)
|
1 |
Detector on poly at +1250V |
80 |
Yes |
31 |
100
|
2 |
Detector on poly at +1250V |
80 |
Yes |
18 |
100
|
3 |
Detector on poly at +1250V |
80 |
Yes |
19 |
100
|
4 |
Detector on poly at +1250V |
80 |
Yes |
31 |
100
|
5 |
Detector on poly at +1250V |
80 |
Yes |
30 |
100
|
6" Green poly (three bricks) in Box
1250V source in |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec)
|
1 |
Detector on poly at +1250V |
80 |
Yes |
35 |
100
|
2 |
Detector on poly at +1250V |
80 |
Yes |
24 |
100
|
3 |
Detector on poly at +1250V |
80 |
Yes |
23 |
100
|
4 |
Detector on poly at +1250V |
80 |
Yes |
21 |
100
|
5 |
Detector on poly at +1250V |
80 |
Yes |
22 |
100
|
8" Green poly (four bricks) in Box
1250V source in |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec)
|
1 |
Detector on poly at +1250V |
80 |
Yes |
20 |
100
|
2 |
Detector on poly at +1250V |
80 |
Yes |
28 |
100
|
3 |
Detector on poly at +1250V |
80 |
Yes |
29 |
100
|
4 |
Detector on poly at +1250V |
80 |
Yes |
16 |
100
|
5 |
Detector on poly at +1250V |
80 |
Yes |
22 |
100
|
Previous Page