| Line 293: |
Line 293: |
| ='''Thermistor Reading Log'''= | | ='''Thermistor Reading Log'''= |
|
| |
|
| {| class="wikitable" | | {| class="wikitable sortable mw-collapsible mw-collapsed" border="2" style="text-align:center;" |cellpadding="20" cellspacing="1 |
| | | |- |
| ! Date !! Thermistor Measurement (Ω) !! Detector Thermal Cycled Between Previous Measurement | | ! Date !! Thermistor Measurement (Ω) !! Detector Thermal Cycled Between Previous Measurement |
| |- | | |- |
Revision as of 20:20, 16 December 2021
How to Read HPGe Thermistor
Make sure detector high voltage is turned off
Unplug pre-amp power
Unscrew detector electronics housing (8 screws total)
Arrows pointing to screws on detector electronics housing
Slide electronics housing towards the crystal end of the detector
Slide the sectioned area in the direction of the arrow to expose the thermistor terminals
Turn multimeter to resistance (Ω) setting
Measure resistance across the white/black wires the exit the white nylon 4 pinned terminal
Check resistance in tables below for corresponding detector temperature
Resistance and Corresponding Temperatures
100-109Ω
| Thermistor Resistance (Ω) |
Detector Temperature (Κ)
|
| 100 |
76.5780
|
| 101 |
77.0421
|
| 102 |
77.5063
|
| 103 |
77.9706
|
| 104 |
78.4352
|
| 105 |
78.9000
|
| 106 |
79.3649
|
| 107 |
79.8300
|
| 108 |
80.2952
|
| 109 |
80.7607
|
110-119Ω
| Thermistor Resistance (Ω) |
Detector Temperature (Κ)
|
| 110 |
81.2263
|
| 111 |
81.6921
|
| 112 |
82.1581
|
| 113 |
82.6242
|
| 114 |
83.0906
|
| 115 |
83.5571
|
| 116 |
84.0238
|
| 117 |
84.4906
|
| 118 |
84.9576
|
| 119 |
85.4248
|
120-129Ω
| Thermistor Resistance (Ω) |
Detector Temperature (Κ)
|
| 120 |
85.8922
|
| 121 |
86.3597
|
| 122 |
86.8275
|
| 123 |
87.2954
|
| 124 |
87.7634
|
| 125 |
88.2317
|
| 126 |
88.7001
|
| 127 |
89.1686
|
| 128 |
89.6374
|
| 129 |
90.1063
|
130-139Ω
| Thermistor Resistance (Ω) |
Detector Temperature (Κ)
|
| 130 |
90.5754
|
| 131 |
91.0446
|
| 132 |
91.5141
|
| 133 |
91.9837
|
| 134 |
92.4524
|
| 135 |
92.9234
|
| 136 |
93.3935
|
| 137 |
93.8637
|
| 138 |
94.3342
|
| 139 |
94.8048
|
140-149Ω
| Thermistor Resistance (Ω) |
Detector Temperature (Κ)
|
| 140 |
95.2755
|
| 141 |
95.7465
|
| 142 |
96.2176
|
| 143 |
96.6888
|
| 144 |
97.1603
|
| 145 |
97.6319
|
| 146 |
98.1036
|
| 147 |
98.5756
|
| 148 |
99.0477
|
| 149 |
99.5199
|
150-159Ω
| Thermistor Resistance (Ω) |
Detector Temperature (Κ)
|
| 150 |
99.9923
|
| 151 |
100.4649
|
| 152 |
100.9377
|
| 153 |
101.4106
|
| 154 |
101.8836
|
| 155 |
102.3569
|
| 156 |
102.8303
|
| 157 |
103.3038
|
| 158 |
103.7775
|
| 159 |
104.2514
|
160-169Ω
| Thermistor Resistance (Ω) |
Detector Temperature (Κ)
|
| 160 |
104.7254
|
| 161 |
105.1996
|
| 162 |
105.6740
|
| 163 |
106.1485
|
| 164 |
106.6232
|
| 165 |
107.0980
|
| 166 |
107.5730
|
| 167 |
108.0482
|
| 168 |
108.5235
|
| 169 |
108.9989
|
170-179Ω
| Thermistor Resistance (Ω) |
Detector Temperature (Κ)
|
| 170 |
109.4746
|
| 171 |
109.9503
|
| 172 |
110.4263
|
| 173 |
110.9024
|
| 174 |
111.3786
|
| 175 |
111.8550
|
| 176 |
112.3316
|
| 177 |
112.8083
|
| 178 |
113.2851
|
| 179 |
113.7622
|
180-189Ω
| Thermistor Resistance (Ω) |
Detector Temperature (Κ)
|
| 180 |
114.2393
|
| 181 |
114.7167
|
| 182 |
115.1941
|
| 183 |
115.6718
|
| 184 |
116.1496
|
| 185 |
116.6275
|
| 186 |
117.1056
|
| 187 |
117.5838
|
| 188 |
118.0622
|
| 189 |
118.5408
|
190-199Ω
| Thermistor Resistance (Ω) |
Detector Temperature (Κ)
|
| 190 |
119.0194
|
| 191 |
119.4983
|
| 192 |
119.9773
|
| 193 |
120.4564
|
| 194 |
120.9357
|
| 195 |
121.4152
|
| 196 |
121.8947
|
| 197 |
122.3745
|
| 198 |
122.8544
|
| 199 |
123.3344
|
Thermistor Reading Log
| Date |
Thermistor Measurement (Ω) |
Detector Thermal Cycled Between Previous Measurement
|
|
|
|
|
|
|
|
|
|