=Total Efficiency=
 
=Total Efficiency=
  
This is total efficiency  
+
This is total efficiency using the Gadolinium foil as the source.
  
 
<math> (line\ intensity)\ (incident\ thermal\ neutrons\ per\ second)\ (\sigma_{total})\ (MCNP\ fudge\ factor\ from\ Paul\ ~4)\ =\ Theoretical\ Rate </math>
 
<math> (line\ intensity)\ (incident\ thermal\ neutrons\ per\ second)\ (\sigma_{total})\ (MCNP\ fudge\ factor\ from\ Paul\ ~4)\ =\ Theoretical\ Rate </math>

Revision as of 19:38, 28 September 2021

Eu152 Calibration

  • HPGe calibrated using Eu152 source
CH Eu152 Calibration Spectrum.png

As of 9/28/21

The calibration used to map from ADC channel number to KeV is:

[math]Photon Energy = -36.2928 + 0.449519 x - 0.000000861277 x^2[/math]

Where x = channel 9 of the ADC and is input as PADC.PADC785N[9].

Total Efficiency

This is total efficiency using the Gadolinium foil as the source.

[math] (line\ intensity)\ (incident\ thermal\ neutrons\ per\ second)\ (\sigma_{total})\ (MCNP\ fudge\ factor\ from\ Paul\ ~4)\ =\ Theoretical\ Rate [/math]

Then find detected rate using HPGe

Divide (detected rate) by (theoretical rate) for total detector efficiency


PAS HPGe Detector