Eu152 Calibration

  • HPGe calibrated using Eu152 source

As of 9/28/21

The calibration used to map from ADC channel number to KeV is:

[math]\displaystyle{ Photon Energy = -36.2928 + 0.449519 x - 0.000000861277 x^2 }[/math]

Where x = channel 9 of the ADC and is input as PADC.PADC785N[9].

Theoretical Efficiency

This is total efficiency

[math]\displaystyle{ (line\ intensity)\ (incident\ thermal\ neutrons\ per\ second)\ (\sigma_{total})\ (MCNP\ fudge\ factor\ from\ Paul\ ~4)\ =\ Theoretical\ Rate }[/math]


PAS HPGe Detector