Difference between revisions of "CH HPGe Efficiency"

From New IAC Wiki
Jump to navigation Jump to search
Line 7: Line 7:
 
==As of 9/28/21==
 
==As of 9/28/21==
  
The calibration used to map from ADC channel number to KeV is <math>-36.2928+0.449519x-0.000000861277x^2</math>
+
The calibration used to map from ADC channel number to KeV is:
 +
 
 +
<math>Photon Energy = -36.2928 + 0.449519 x - 0.000000861277 x^2</math>
 +
 
 +
Where x is channel ''9'' of the ADC and is input as PADC.PADC785N[9].
  
  
 
----
 
----
 
[[PAS HPGe Detector]]
 
[[PAS HPGe Detector]]

Revision as of 18:13, 28 September 2021

Eu152 Calibration

  • HPGe calibrated using Eu152 source
CH Eu152 Calibration Spectrum.png

As of 9/28/21

The calibration used to map from ADC channel number to KeV is:

[math]Photon Energy = -36.2928 + 0.449519 x - 0.000000861277 x^2[/math]

Where x is channel 9 of the ADC and is input as PADC.PADC785N[9].



PAS HPGe Detector