Line 19: | Line 19: | ||
| Run Number || Configuration || Discriminator Threshold (mV) || Cs-137 Button Source || Counts || Run Time (Sec) || Count Rate (Hz) | | Run Number || Configuration || Discriminator Threshold (mV) || Cs-137 Button Source || Counts || Run Time (Sec) || Count Rate (Hz) | ||
|- | |- | ||
− | | 1 || Unshielded detector at +750V || 05 || No || | + | | 1 || Unshielded detector at +750V || 05 || No || 2211 || 60 || |
|- | |- | ||
− | | 2 || Unshielded detector at +750V || 15 || No || | + | | 2 || Unshielded detector at +750V || 15 || No || 1125 || 60 || |
|- | |- | ||
− | | 3 || Unshielded detector at +750V || 35 || No || | + | | 3 || Unshielded detector at +750V || 35 || No || 293 || 60 || |
|- | |- | ||
− | | 4 || Unshielded detector at +750V || 55 || No || | + | | 4 || Unshielded detector at +750V || 55 || No || 105 || 60 || |
|- | |- | ||
− | | 5 || Unshielded detector at +750V || 75 || No || | + | | 5 || Unshielded detector at +750V || 75 || No || 80 || 60 || |
|- | |- | ||
− | | 6 || Unshielded detector at +750V || 95 || No || | + | | 6 || Unshielded detector at +750V || 95 || No || 97 || 60 || |
|- | |- | ||
− | | 7 || Unshielded detector at +750V || 115 || No || | + | | 7 || Unshielded detector at +750V || 115 || No || 99 || 60 || |
|- | |- | ||
− | | 8 || Unshielded detector at +750V || 135 || No || | + | | 8 || Unshielded detector at +750V || 135 || No || 89 || 60 || |
|- | |- | ||
− | | 9 || Unshielded detector at +750V || 155 || No || | + | | 9 || Unshielded detector at +750V || 155 || No || 74 || 60 || |
|- | |- | ||
− | | 10 || Unshielded detector at +750V || 175 || No || | + | | 10 || Unshielded detector at +750V || 175 || No || 80 || 60 || |
|- | |- | ||
− | | 11 || Unshielded detector at +750V || 195 || No || | + | | 11 || Unshielded detector at +750V || 195 || No || 80 || 60 || |
|- | |- | ||
− | | 12 || Unshielded detector at +750V || 215 || No || | + | | 12 || Unshielded detector at +750V || 215 || No || 64 || 60 || |
|- | |- | ||
− | | 13 || Unshielded detector at +750V || 235 || No || | + | | 13 || Unshielded detector at +750V || 235 || No || 89 || 60 || |
|- | |- | ||
− | | 14 || Unshielded detector at +750V || 255 || No || | + | | 14 || Unshielded detector at +750V || 255 || No || 69 || 60 || |
|- | |- | ||
|} | |} |
Revision as of 17:59, 19 May 2021
JD-766 Cs-137 Button Source Plateau
Timing filter amp configuration
Course gain: x6
Fine gain: at minimum
Integrate (nS): 500nS
Diff (nS): Out
Invert/Noninverted: Non-Inv
Plateau
Run Number Configuration Discriminator Threshold (mV) Cs-137 Button Source Counts Run Time (Sec) Count Rate (Hz) 1 Unshielded detector at +750V 05 No 2211 60 2 Unshielded detector at +750V 15 No 1125 60 3 Unshielded detector at +750V 35 No 293 60 4 Unshielded detector at +750V 55 No 105 60 5 Unshielded detector at +750V 75 No 80 60 6 Unshielded detector at +750V 95 No 97 60 7 Unshielded detector at +750V 115 No 99 60 8 Unshielded detector at +750V 135 No 89 60 9 Unshielded detector at +750V 155 No 74 60 10 Unshielded detector at +750V 175 No 80 60 11 Unshielded detector at +750V 195 No 80 60 12 Unshielded detector at +750V 215 No 64 60 13 Unshielded detector at +750V 235 No 89 60 14 Unshielded detector at +750V 255 No 69 60
Run Number Configuration Discriminator Threshold (mV) Cs-137 Button Source Counts Run Time (Sec) Count Rate (Hz) 1 Unshielded detector at +800V 05 No 3866 60 2 Unshielded detector at +800V 15 No 2356 60 3 Unshielded detector at +800V 35 No 1091 60 4 Unshielded detector at +800V 55 No 384 60 5 Unshielded detector at +800V 75 No 244 60 6 Unshielded detector at +800V 95 No 167 60 7 Unshielded detector at +800V 115 No 100 60 8 Unshielded detector at +800V 135 No 88 60 9 Unshielded detector at +800V 155 No 89 60 10 Unshielded detector at +800V 175 No 86 60 11 Unshielded detector at +800V 195 No 90 60 12 Unshielded detector at +800V 215 No 81 60 13 Unshielded detector at +800V 235 No 95 60 14 Unshielded detector at +800V 255 No 89 60
Run Number Configuration Discriminator Threshold (mV) Cs-137 Button Source Counts Run Time (Sec) Count Rate (Hz) 1 Unshielded detector at +850V 05 No 6402 60 2 Unshielded detector at +850V 15 No 3867 60 3 Unshielded detector at +850V 35 No 2058 60 4 Unshielded detector at +850V 55 No 1233 60 5 Unshielded detector at +850V 75 No 787 60 6 Unshielded detector at +850V 95 No 391 60 7 Unshielded detector at +850V 115 No 256 60 8 Unshielded detector at +850V 135 No 193 60 9 Unshielded detector at +850V 155 No 151 60 10 Unshielded detector at +850V 175 No 97 60 11 Unshielded detector at +850V 195 No 105 60 12 Unshielded detector at +850V 215 No 97 60 13 Unshielded detector at +850V 235 No 87 60 14 Unshielded detector at +850V 255 No 101 60