Irradiated on Jack by placing samples just above the Radiator as shown in the picture.
Irradiated on Jack by placing samples just above the Radiator as shown in the picture.
Beam parameters:
Irradiation Date : 5/23/18
Start Time: 09:07:30
Stop Time: 11:07:30
Rep Rate=300 Hz
I = 98 mA
Pulse Width = 10 usec
E = 38 MeV


The solar panel current output dropped 1 mV per hour during irradiation.  The cove was too dark to produce a signal from ambient light.  The observed signal was completely due to the electron beam.
The solar panel current output dropped 1 mV per hour during irradiation.  The cove was too dark to produce a signal from ambient light.  The observed signal was completely due to the electron beam.
Line 8: Line 24:
In File name:
In File name:
10uC_Cs_Mn_SolarPan_DetD_55.5cm_002.mpa
10uC_Cs_Mn_SolarPan_DetD_55.5cm_002.mpa
Measurements started at 14:45:38




Line 17: Line 36:
|0 ||  Cs-137 (661 keV line) #64121 55.5 cm away from face
|0 ||  Cs-137 (661 keV line) #64121 55.5 cm away from face
|-
|-
|360 ||  Mn sample (847 keV) 55.5 cm away from face
|360 ||  Mn sample (1.7 g, 1.5 cm x 1.5 cm square, 847 keV) 55.5 cm away from face
|-
|-
|690 ||  Thorium Rods on face of detector
|690 ||  Thorium Rods on face of detector

Revision as of 22:09, 23 May 2018

PAA_Research

Irradiated on Jack by placing samples just above the Radiator as shown in the picture.

Beam parameters: Irradiation Date : 5/23/18

Start Time: 09:07:30

Stop Time: 11:07:30

Rep Rate=300 Hz

I = 98 mA

Pulse Width = 10 usec

E = 38 MeV


The solar panel current output dropped 1 mV per hour during irradiation. The cove was too dark to produce a signal from ambient light. The observed signal was completely due to the electron beam.


In File name: 10uC_Cs_Mn_SolarPan_DetD_55.5cm_002.mpa

Measurements started at 14:45:38


Category Cost
Time (sec) Sample
0 Cs-137 (661 keV line) #64121 55.5 cm away from face
360 Mn sample (1.7 g, 1.5 cm x 1.5 cm square, 847 keV) 55.5 cm away from face
690 Thorium Rods on face of detector
1020 Solar Panel A 55.5 cm away from face
1410 Cs-137 #129792 55.5 cm away from face
1820 Mn sample (847 keV) 55.5 cm away from face
2230 Cs-137 #129792 55.5 cm away from face (0.54% deadtime)
2560 Solar Panel A 55.5 cm away from face (0.55% deadtime, 300 Hz DAQ rate)
2880 Cs-137 (661 keV line) #64121 55.5 cm away from face (1.1% deadtime, 350 Hz)
3070 Mn sample (847 keV) 55.5 cm away from face (13 % deadtime, 4600 Hz)
3330 Ignore (Mn source was too close) Cs-137 #129792 15 cm away from face(6 % deadtime, 2000 Hz)
3600 Cs-137 #129792 15 cm away from face(.5 % deadtime, 200 Hz)
4000 Solar Panel A 15 cm away from face (2 % deadtime, 880 Hz DAQ rate)