Difference between revisions of "LB May 2017 Det A Efficiency"

From New IAC Wiki
Jump to navigation Jump to search
(Created page with " {| border="3" cellpadding="5" cellspacing="0" | Run || Source || Energy (keV) || Position || Expected Rate (Hz) ||ROOT Window || HpGe Rate (Hz) || HpGe Detector A Efficiency …")
 
Line 1: Line 1:
  {| border="3"  cellpadding="5" cellspacing="0"
+
=Theoretical Activities=
 +
 
 +
 
 +
 
 +
Na-22, 10.19 uCi on July 15, 2013, half life 2.602 +/- 0.002 years, 99.937% for 1274.52 keV line and 178.8 for 511 keV line , activity in Feb 4, 2017 = 3.947 uCi
 +
:= <math>\left (0.99937 \right )\left ( 3.947 \times 10^{-6} \mbox{Ci} \right) \left (\frac{ (3.7 \times 10^{10} \mbox{Hz}}{\mbox{Ci}} \right)= 145,946 Hz </math> for the 1274 keV line
 +
 
 +
:= <math>\left (1.788 \right )\left ( 3.947 \times 10^{-6} \mbox{Ci} \right) \left (\frac{ (3.7 \times 10^{10} \mbox{Hz}}{\mbox{Ci}} \right)= 261,117.732 Hz </math> for the 511 keV line
 +
 
 +
Now compute the expected rate for February 5, 2017.
 +
 
 +
The activity on Feb 5, 2017 is 3.944 uCi, so the expected rates for the 2 lines are
 +
 
 +
:= <math>\left (0.99937 \right )\left ( 3.944 \times 10^{-6} \mbox{Ci} \right) \left (\frac{ (3.7 \times 10^{10} \mbox{Hz}}{\mbox{Ci}} \right)= 145,836 Hz </math> for the 1274 keV line
 +
 
 +
:= <math>\left (1.788 \right )\left ( 3.944 \times 10^{-6} \mbox{Ci} \right) \left (\frac{ (3.7 \times 10^{10} \mbox{Hz}}{\mbox{Ci}} \right)=  260919.264 Hz </math> for the 511 keV line
 +
 
 +
Now compute the expected rate for February 9, 2017
 +
 
 +
The activity on Feb 9, 2017 is 3.933 uCi, so the expected rate for the 2 lines are
 +
 
 +
 
 +
:= <math>\left (0.99937 \right )\left ( 3.933 \times 10^{-6} \mbox{Ci} \right) \left (\frac{ (3.7 \times 10^{10} \mbox{Hz}}{\mbox{Ci}} \right)= 145,429.32 Hz </math> for the 1274 keV line
 +
 
 +
:= <math>\left (1.788 \right )\left ( 3.933 \times 10^{-6} \mbox{Ci} \right) \left (\frac{ (3.7 \times 10^{10} \mbox{Hz}}{\mbox{Ci}} \right)=  260,191.55 Hz </math> for the 511 keV line
 +
 
 +
Now compute the expected rate for Feb 12, 2017
 +
 
 +
The activity on Feb 12, 2017 is 3.924 uCi, so the expected rate for the 2 lines are
 +
 
 +
:= <math>\left (0.99937 \right )\left ( 3.924 \times 10^{-6} \mbox{Ci} \right) \left (\frac{ (3.7 \times 10^{10} \mbox{Hz}}{\mbox{Ci}} \right)= 145096.5316 Hz </math> for the 1274 keV line
 +
 
 +
:= <math>\left (1.788 \right )\left ( 3.924 \times 10^{-6} \mbox{Ci} \right) \left (\frac{ (3.7 \times 10^{10} \mbox{Hz}}{\mbox{Ci}} \right)=  259596.144 Hz </math> for the 511 keV line
 +
 
 +
=Actual Efficiency=
 +
 
 +
{| border="3"  cellpadding="5" cellspacing="0"
 
|  Run || Source || Energy (keV) || Position ||  Expected Rate (Hz) ||ROOT Window || HpGe Rate (Hz) || HpGe Detector A Efficiency (%)   
 
|  Run || Source || Energy (keV) || Position ||  Expected Rate (Hz) ||ROOT Window || HpGe Rate (Hz) || HpGe Detector A Efficiency (%)   
 
|-
 
|-

Revision as of 21:48, 22 August 2017

Theoretical Activities

Na-22, 10.19 uCi on July 15, 2013, half life 2.602 +/- 0.002 years, 99.937% for 1274.52 keV line and 178.8 for 511 keV line , activity in Feb 4, 2017 = 3.947 uCi

= [math]\left (0.99937 \right )\left ( 3.947 \times 10^{-6} \mbox{Ci} \right) \left (\frac{ (3.7 \times 10^{10} \mbox{Hz}}{\mbox{Ci}} \right)= 145,946 Hz [/math] for the 1274 keV line
= [math]\left (1.788 \right )\left ( 3.947 \times 10^{-6} \mbox{Ci} \right) \left (\frac{ (3.7 \times 10^{10} \mbox{Hz}}{\mbox{Ci}} \right)= 261,117.732 Hz [/math] for the 511 keV line

Now compute the expected rate for February 5, 2017.

The activity on Feb 5, 2017 is 3.944 uCi, so the expected rates for the 2 lines are

= [math]\left (0.99937 \right )\left ( 3.944 \times 10^{-6} \mbox{Ci} \right) \left (\frac{ (3.7 \times 10^{10} \mbox{Hz}}{\mbox{Ci}} \right)= 145,836 Hz [/math] for the 1274 keV line
= [math]\left (1.788 \right )\left ( 3.944 \times 10^{-6} \mbox{Ci} \right) \left (\frac{ (3.7 \times 10^{10} \mbox{Hz}}{\mbox{Ci}} \right)= 260919.264 Hz [/math] for the 511 keV line

Now compute the expected rate for February 9, 2017

The activity on Feb 9, 2017 is 3.933 uCi, so the expected rate for the 2 lines are


= [math]\left (0.99937 \right )\left ( 3.933 \times 10^{-6} \mbox{Ci} \right) \left (\frac{ (3.7 \times 10^{10} \mbox{Hz}}{\mbox{Ci}} \right)= 145,429.32 Hz [/math] for the 1274 keV line
= [math]\left (1.788 \right )\left ( 3.933 \times 10^{-6} \mbox{Ci} \right) \left (\frac{ (3.7 \times 10^{10} \mbox{Hz}}{\mbox{Ci}} \right)= 260,191.55 Hz [/math] for the 511 keV line

Now compute the expected rate for Feb 12, 2017

The activity on Feb 12, 2017 is 3.924 uCi, so the expected rate for the 2 lines are

= [math]\left (0.99937 \right )\left ( 3.924 \times 10^{-6} \mbox{Ci} \right) \left (\frac{ (3.7 \times 10^{10} \mbox{Hz}}{\mbox{Ci}} \right)= 145096.5316 Hz [/math] for the 1274 keV line
= [math]\left (1.788 \right )\left ( 3.924 \times 10^{-6} \mbox{Ci} \right) \left (\frac{ (3.7 \times 10^{10} \mbox{Hz}}{\mbox{Ci}} \right)= 259596.144 Hz [/math] for the 511 keV line

Actual Efficiency

Run Source Energy (keV) Position Expected Rate (Hz) ROOT Window HpGe Rate (Hz) HpGe Detector A Efficiency (%)