In this experiment, I picked 17 different points on my OSL board to test agreement between two OSLs at the same physical location for the same shot from TriMeV.
 
In this experiment, I picked 17 different points on my OSL board to test agreement between two OSLs at the same physical location for the same shot from TriMeV.
Line 66: Line 67:
  
  
<div style="text-align: center;">='''Shot 179'''=
+
<div style="text-align: center;">
 +
='''Shot 179'''=
  
 
In this experiment, I picked 17 different points on my OSL board to test agreement between two OSLs at the same physical location for the same shot from TriMeV.
 
In this experiment, I picked 17 different points on my OSL board to test agreement between two OSLs at the same physical location for the same shot from TriMeV.

Latest revision as of 21:40, 3 August 2017

Shot 169

In this experiment, I picked 17 different points on my OSL board to test agreement between two OSLs at the same physical location for the same shot from TriMeV.

I have overlaid X and Y profiles as well as created 3D renderings of the beam profile for the front and back sets of OSLs. Note that it is flat due to the distance at which the board was placed from the end of the accelerator.

Diode diameter: 2"

Diode gap: 1.5"

Al vacuum plate thickness: 5/8"

Ta thickness: 0.020"

Distance to Board: 109 cm

Calculated amount of charge: -2.106e07 C


*Unknown calibration on OSL reader
*Unknown calibration on OSL reader
*Unknown calibration on OSL reader
*Unknown calibration on OSL reader
*Unknown calibration on OSL reader
Shot169 Current.png


Shot 170

In this experiment, I picked 17 different points on my OSL board to test agreement between two OSLs at the same physical location for the same shot from TriMeV.

I have overlaid X and Y profiles as well as created 3D renderings of the beam profile for the front and back sets of OSLs. Note that it is flat due to the distance at which the board was placed from the end of the accelerator.

Diode diameter: 2"

gap: 2.5"

Al vacuum plate thickness: 1/8"

Ta thickness: 0.020"

Distance to Board: 102 cm

Calculated amount of charge: -1.679E07


*Unknown calibration on OSL reader
*Unknown calibration on OSL reader
*Unknown calibration on OSL reader
*Unknown calibration on OSL reader
*Unknown calibration on OSL reader
Shot170Current.png


Shot 179

In this experiment, I picked 17 different points on my OSL board to test agreement between two OSLs at the same physical location for the same shot from TriMeV.

I have overlaid X and Y profiles as well as created 3D renderings of the beam profile for the front and back sets of OSLs. Note that it is flat due to the distance at which the board was placed from the end of the accelerator.

Diode diameter: 1" hemispherical

gap: 2.1"

Al vacuum plate thickness: 1/8"

Ta thickness: 0.020"

Distance to Board: 99 cm

Calculated amount of charge: -1.498E07


Shot179Current.png


TriMeV Shot Log