The activity on Feb 12, 2017 is 3.924 uCi, so the expected rate for the 2 lines are | The activity on Feb 12, 2017 is 3.924 uCi, so the expected rate for the 2 lines are | ||
− | := <math>\left (0.99937 \right )\left ( 3.924 \times 10^{-6} \mbox{Ci} \right) \left (\frac{ (3.7 \times 10^{10} \mbox{Hz}}{\mbox{Ci}} \right)= | + | := <math>\left (0.99937 \right )\left ( 3.924 \times 10^{-6} \mbox{Ci} \right) \left (\frac{ (3.7 \times 10^{10} \mbox{Hz}}{\mbox{Ci}} \right)= 145096.5316 Hz </math> for the 1274 keV line |
− | := <math>\left (1.788 \right )\left ( 3.924 \times 10^{-6} \mbox{Ci} \right) \left (\frac{ (3.7 \times 10^{10} \mbox{Hz}}{\mbox{Ci}} \right)= | + | := <math>\left (1.788 \right )\left ( 3.924 \times 10^{-6} \mbox{Ci} \right) \left (\frac{ (3.7 \times 10^{10} \mbox{Hz}}{\mbox{Ci}} \right)= 259596.144 Hz </math> for the 511 keV line |
Below is a table where the actual efficiency will be calculated for position R (farthest position). | Below is a table where the actual efficiency will be calculated for position R (farthest position). |
Revision as of 21:42, 16 February 2017
Below are the theoretical calculations for the theoretical decay frequencies
Na-22, 10.19 uCi on July 15, 2013, half life 2.602 +/- 0.002 years, 99.937% for 1274.52 keV line and 178.8 for 511 keV line , activity in Feb 4, 2017 = 3.947 uCi
- = for the 1274 keV line
- = for the 511 keV line
Now compute the expected rate for February 5, 2017.
The activity on Feb 5, 2017 is 3.944 uCi, so the expected rates for the 2 lines are
- = for the 1274 keV line
- = for the 511 keV line
Now compute the expected rate for February 9, 2017
The activity on Feb 9, 2017 is 3.933 uCi, so the expected rate for the 2 lines are
- = for the 1274 keV line
- = for the 511 keV line
Now compute the expected rate for Feb 12, 2017
The activity on Feb 12, 2017 is 3.924 uCi, so the expected rate for the 2 lines are
- = for the 1274 keV line
- = for the 511 keV line
Below is a table where the actual efficiency will be calculated for position R (farthest position).
Run | Source | Energy (keV) | Expected Rate (Hz) | HpGe Rate (Hz) | HpGe Det D Efficiency (%) |
r5825 | Na-22 | 511 | 79122.6 | (480:540) (4.309-0.065=4.244) | 0.005 |
r5826 | Na-22 | 511 | 28091.2 | (480:540)(1.105-0.02281=1.0821) | 0.004 |
r5827 | Na-22 | 511 | 861.1 | (480:540)(0.04037-0.009123=0.031247) | 0.004 |
r5829 | Na-22 | 511 | 137692 | (480:540)(3.686-0.01939=3.67) | 0.003 |
r5830 | Na-22 | 511 | 44224 | (480:540) (1.073-0.0057=1.0673) | 0.002 |
r5831 | Na-22 | 511 | 139058.52 | (480:540)(3.283-0.05702=3.22598) | 0.002 |