Testing results

Wire Signal Testing

Signal testing performed to all S6's wires for cosmic rays, the testing was simultaneously with the wire burning process. In general the signal is detected when the voltage on the sense wire at least was 1200V, guard's 535V, and field's -750V.

The results are shown in the following table:

Super Layer STP Board Wires without any Signal Hot wires High Noise Wires Notes
SL1 STB1-G1202270 4 24,38,66,87,93 1,17 Sense wire voltage 1380V
SL1 STB2-G1202266 1,49,82 38,40,51,65 N/A Sense wire voltage 1360V or 1500V
SL1 STB3-G1202273 N/A 1,2,39,78 N/A
SL1 STB4-G1202275 N/A 61 61
SL1 STB5-G1202271 N/A 8,60,88,89 8,17,18,33,34,49,50,65,66,81,82,
SL1 STB6-G1202274 N/A N/A 1,2
SL1 STB7-G1202272 6,83 23,33,34,69,78,92 1,2,82 49-64 N/A