The power supply is connected to a spark protection circuit. CAEN N470 has a signal Lemo 00 input to activate a kill option automatically when a discharge signal is created through the detector operation, as the the signal's  is more than  15 us. Kill option will ramp down all the channels' voltages and turn them off.
  
  

Revision as of 20:10, 30 June 2013

Detector operation

Operating the detector occurs by providing voltage to the HV-circuit connected to the detector components mentioned in the previous section. A CAEN N470 four channel power supply is used for this purpose, using a channel of the power supply allow to increase the voltage difference between the top and bottom side of the GEM preamplifier reaches up to 350 V as the circuit board is provided with a voltage of 2.8 kV.

Another channel of the power supply is used to increase the voltage of the Cathode, when the cathode voltage is 3.1 kV, it will create a a drift voltage up to 300V.


Abdel GEM HV distrib 6-19-09.png

The power supply is connected to a spark protection circuit. CAEN N470 has a signal Lemo 00 input to activate a kill option automatically when a discharge signal is created through the detector operation, as the the signal's is more than 15 us. Kill option will ramp down all the channels' voltages and turn them off.



Go Back [1]