Difference between revisions of "Cf252 NaI detector test. 12 Apr 2012"

From New IAC Wiki
Jump to navigation Jump to search
Line 42: Line 42:
 
|}
 
|}
  
rise time of NaI 50 ns
+
 
decay time of NaI 3 us
+
* rise time of NaI 50 ns
 +
* decay time of NaI 3 us
  
  
 
[[File:Cf 2inchBoratedPoly infront.eps]]
 
[[File:Cf 2inchBoratedPoly infront.eps]]

Revision as of 19:35, 12 April 2012

go back

NaI detector, -1030V, 300 mV threshold on gate channel, signal ADC4


Run # source shielding cnts rate Comments
2775 Co-60 some distance away from NaI 2" Pb around, no front shielding
2776 bkgd 2" Pb around, no front shielding
2777 Co-60 on NaI taped 2" Pb around, no front shielding clear 2 big 1.2 and 1.4 MeV peaks, small peak around 4.3 MeV

Calibration 1.4 MeV/575 cnls = 0.00243 MeV/cnl

2778 bkgd 2" Pb around, no front shielding 41.75 Hz
2779 bkgd all around 2" Pb around 15.74 Hz good area around 2000-3800 channels
2780 Cf252 all around 2" Pb around 12 kHz Cf252 just in front of NaI
2781 Cf252 all around 2" Pb around 5.4 kHz Cf252 about 15 cm away from NaI
2782 Cf252 all around 2" Pb around 1.1 kHz Cf252 17" away from NaI
2784 Cf252 all around 2" Pb around 1,372,745 1,140 Hz 20 min run
2785 Cf252 all around 2" Pb around 153,508 129.5 Hz 20 min run. 2" Pb just after Cf252
2787 Cf252 all around 2" Pb around 1,231,537 1,024 Hz 20 min run. 2" Borated Poly just after Cf252
2788 Cf252 all around 2" Pb around 1,225,232 1,020 Hz 20 min run. 2" NoBorated Poly just after Cf252
2790 Cf252 all around 2" Pb around 132,067 111.6 Hz 20 min run. 2" Pb, 2" Borated Poly just after Cf252


  • rise time of NaI 50 ns
  • decay time of NaI 3 us


File:Cf 2inchBoratedPoly infront.eps