Difference between revisions of "05/19/2011 n's detector threshold calibration source/(w/source)"
Line 2: | Line 2: | ||
− | Below are my counts rate measurements as function of threshold voltage for n's detector. The source was CF252. The distance between source and n's PMT was about 50 cm. | + | Below are my counts rate measurements as function of threshold voltage for n's detector. The source was CF252. The distance between source and n's PMT was about 50 cm. It was 1 PMT set up detector and experiment was hold in LDS on May 19 2011. |
+ | |||
+ | First the counts rate was calculated as: | ||
<math>\mbox{Rate} = \frac{N_{\mbox{source}} - N_{\mbox{w/source}}}{\mbox{Time}}</math> | <math>\mbox{Rate} = \frac{N_{\mbox{source}} - N_{\mbox{w/source}}}{\mbox{Time}}</math> | ||
Line 22: | Line 24: | ||
− | '''There are no any plateaus, so what is the threshold should be?''' | + | '''There are no any plateaus, so what is the threshold should be?'''. Or maybe we are need to look at something different. |
− | Let's recalculate data using the formula: | + | Let's recalculate all data using the formula: |
<math>\mbox{Ratio} = \frac{R_s - R_{w/s}}{R_{w/s}}</math> | <math>\mbox{Ratio} = \frac{R_s - R_{w/s}}{R_{w/s}}</math> | ||
− | + | Below my data and calculations for the case of -1500 V power supply voltage: | |
− | |||
[[File:Table 1500V.png | 1000px]] | [[File:Table 1500V.png | 1000px]] | ||
− | + | Below my data and calculations for the case of -1250 V power supply voltage: | |
[[File:Table 1250V.png | 1000px]] | [[File:Table 1250V.png | 1000px]] | ||
− | And below I overlay my results | + | |
+ | And below I overlay my results (linear scale): | ||
[[File:Det ratio lin.png | 1000px]] | [[File:Det ratio lin.png | 1000px]] | ||
− | and for log | + | and for log scale: |
[[File:Det ratio log.png | 1000px]] | [[File:Det ratio log.png | 1000px]] | ||
− | + | ||
+ | There are two important conclusion at least for this type of PMT was used: | ||
# The desirable threshold is about -50 mV | # The desirable threshold is about -50 mV | ||
# The power supply voltage is as expected -1500 V (there are no plateau for -1250 V) | # The power supply voltage is as expected -1500 V (there are no plateau for -1250 V) | ||
+ | |||
+ | And also some difference with |
Revision as of 16:32, 3 June 2011
Below are my counts rate measurements as function of threshold voltage for n's detector. The source was CF252. The distance between source and n's PMT was about 50 cm. It was 1 PMT set up detector and experiment was hold in LDS on May 19 2011.
First the counts rate was calculated as:
First I did my calibration for power supply voltage -1250 V:
Second I did my calibration for power supply voltage -1500 V:
And bellow I overlay both cases (-1250 V and -1500 V)
There are no any plateaus, so what is the threshold should be?. Or maybe we are need to look at something different.
Let's recalculate all data using the formula:
Below my data and calculations for the case of -1500 V power supply voltage:
Below my data and calculations for the case of -1250 V power supply voltage:
And below I overlay my results (linear scale):
and for log scale:
There are two important conclusion at least for this type of PMT was used:
- The desirable threshold is about -50 mV
- The power supply voltage is as expected -1500 V (there are no plateau for -1250 V)
And also some difference with