Difference between revisions of "05/13/2011 coincidence test"

From New IAC Wiki
Jump to navigation Jump to search
 
Line 93: Line 93:
 
|-
 
|-
 
!PS1
 
!PS1
|<math>\pm</math>||60 min||<math>\pm0.05</math>
+
|<math>7700\pm87.7</math>||60 min||<math>2.14\pm0.02</math>
 
|-
 
|-
 
!PS2
 
!PS2
|<math>\pm</math>||60 min||<math>\pm0.05</math>
+
|<math>9060\pm95.2</math>||60 min||<math>2.52\pm0.03</math>
 
|-
 
|-
 
!PS1+PS2+n's  
 
!PS1+PS2+n's  
|<math>\pm</math>||60 min||<math>(\pm)\cdot 10^{-2}</math>
+
|<math>214\pm14.6</math>||60 min||<math>(5.9\pm0.4)\cdot 10^{-2}</math>
 
|}
 
|}
  

Latest revision as of 17:27, 16 May 2011

Go Back


Set up like that:

Detector1.png


The voltages, thresholds and widths are as below. Note: I increased my thresholds for PS1 and PS2 from 80 mV up to 100 mV according to my previous calibration threshold line.

PMT voltage, V threshold, mV discr. width, ns
PS1 -1000 -100 50
PS1 -1000 -100 50
n's -1500 -200 10



First let's do the PS1+PS2 coincidence test.

1. The distance between PS and n's PMT is about 80 cm

counts time counts/sec
PS1 2823±53.1 20 min 2.35±0.04
PS2 3091±55.6 20 min 2.58±0.05
PS1+PS2 93±9.6 20 min (7.7±0.8)102

2. The distance between PS and n's PMT is about 50 cm

counts time counts/sec
PS1 7513±86.5 60 min 2.09±0.02
PS2 9036±95.1 60 min 2.51±0.03
PS1+PS2 251±15.8 60 min (6.9±0.4)102


Second do the PS1+PS2+n's coincidence test

1. The distance between PS and n's PMT is about 80 cm

counts time counts/sec
PS1 3229±56.8 20 min 2.69±0.05
PS2 3192±56.5 20 min 2.66±0.05
PS1+PS2+n's 59±7.7 20 min (4.9±0.6)102

2. The distance between PS and n's PMT is about 50 cm

counts time counts/sec
PS1 7700±87.7 60 min 2.14±0.02
PS2 9060±95.2 60 min 2.52±0.03
PS1+PS2+n's 214±14.6 60 min (5.9±0.4)102



The scope pictures of coincidence events looks like:

PS1+PS2 event looks like: PS1+PS2+n event looks like:
Cosmic coinc PS1 PS2.png Cosmic coinc PS1 PS2 n.png