Set up like that:  
 
Set up like that:  
  

Revision as of 16:35, 16 May 2011

Set up like that:

Detector1.png


The voltages, thresholds and widths are as below. Note: I increased my thresholds for PS1 and PS2 from 80 mV up to 100 mV according to my previous calibration threshold line.

PMT voltage, V threshold, mV discr. width, ns
PS1 -1000 -100 50
PS1 -1000 -100 50
n's -1500 -200 10



First let's do the PS1+PS2 coincidence test. The distance between PS and n's PMT is about 80 cm

counts time counts/sec
PS1 2823±53.1 20 min 2.35±0.04
PS2 3091±55.6 20 min 2.58±0.05
PS1+PS2 93±9.6 20 min (7.7±0.8)102

The distance between PS and n's PMT is about 50 cm

counts time counts/sec
PS1 7513±86.5 60 min 2.09±0.02
PS2 9036±95.1 60 min 2.51±0.03
PS1+PS2 251±15.8 60 min (6.9±0.4)102



The scope picture of PS1+PS2 coincidence event like:

Cosmic coinc PS1 PS2.png



Second tet's do the PS1+PS2+n's coincidence test

counts time counts/sec
PS1 3229±56.8 20 min 2.69±0.05
PS2 3192±56.5 20 min 2.66±0.05
PS1+PS2+n's 59±7.7 20 min (4.9±0.6)102


The scope picture of PS1+PS2 coincidence event like:

Cosmic coinc PS1 PS2 n.png


The ratio of (PS+PS2) rate to (PS+PS2+n's) rate is

[math]\frac{7.7}{4.9} = 1.6[/math]