Let's do first the coincidence test PS1+PS2  
+
 
 +
 
 +
'''Let's do first the coincidence test PS1+PS2'''
 
{| border="1" cellspacing="0" style="text-align: center; width: 800px; height: 100px;"
 
{| border="1" cellspacing="0" style="text-align: center; width: 800px; height: 100px;"
 
! scope="col" width="50" |  
 
! scope="col" width="50" |  
Line 34: Line 36:
 
|-
 
|-
 
!PS1
 
!PS1
|2823<math>\pm</math>||20 min||<math>\pm</math>
+
|2823<math>\pm</math>53.1||20 min||2.35<math>\pm</math>0.04
 
|-
 
|-
 
!PS2
 
!PS2
|3091<math>\pm</math>||20 min||<math>\pm</math>
+
|3091<math>\pm</math>55.6||20 min||2.57<math>\pm</math>0.05
 
|-
 
|-
 
!PS1+PS2  
 
!PS1+PS2  
|93<math>\pm</math>||20 min||<math>\pm</math>
+
|93<math>\pm</math>9.6||20 min||(7.7<math>\pm</math>0.8)\cdot 10^{-2}
 
|}
 
|}
  

Revision as of 21:56, 13 May 2011

Go Back

Set up like that:

Detector1.png

the distance between PS and n's PMT is abot 80 cm


The voltages and thresholds are as below. Note: I increased my thresholds for PS1 and PS2 from 80 mV up to 100 mV according to mu previous calibration threshold line.

PMT voltage, V threshold, mV discr. width, ns
PS1 -1500 -80 100
PS1 -1500 -80 100
n's -1500 -200 10



Let's do first the coincidence test PS1+PS2

counts time counts/sec
PS1 2823[math]\pm[/math]53.1 20 min 2.35[math]\pm[/math]0.04
PS2 3091[math]\pm[/math]55.6 20 min 2.57[math]\pm[/math]0.05
PS1+PS2 93[math]\pm[/math]9.6 20 min (7.7[math]\pm[/math]0.8)\cdot 10^{-2}


The scope picture of PS1+PS2 coincidence event like:

Cosmic coinc PS1 PS2.png