Difference between revisions of "05/13/2011 coincidence test"
Jump to navigation
Jump to search
(Created page with '[https://wiki.iac.isu.edu/index.php/N%27s_detector_calibration Go Back] continue...the same set up as before, the distances are the same: 400px Firs…') |
|||
Line 1: | Line 1: | ||
[https://wiki.iac.isu.edu/index.php/N%27s_detector_calibration Go Back] | [https://wiki.iac.isu.edu/index.php/N%27s_detector_calibration Go Back] | ||
− | + | Set up like that: | |
− | |||
[[File:Detector1.png | 400px]] | [[File:Detector1.png | 400px]] |
Revision as of 20:54, 13 May 2011
Set up like that:
First I change the power supply for HAMAMATSU PMT to the new 2 channel one C.A.E.N N471
The powers are:
PS1 and PS2: -1000 V (the old 2-channel TENNELEC TC952) n's: -1500 V (the new 2-channel C.A.E.N N471)
The thresholds are:
PS1 and PS2: -80 mV n's: -200 mV
Let's do first the coincidence test PS1+PS2
counts/min (1 min test) | counts/min (1 min test) | counts/min (5 min test) | counts/min (5 min test) | avg counts/min | |
---|---|---|---|---|---|
PS1 | 187 | 160 | 166 | 160.8 | 168.4 |
PS2 | 187 | 170 | 180.4 | 184.4 | 180.4 |
n's | 205 | 204 | 201.2 | 202.4 | 203.2 |