05/12/2011 2n correlation test: Difference between revisions
Jump to navigation
Jump to search
Created page with '[https://wiki.iac.isu.edu/index.php/N%27s_detector_calibration Go Back] The powers are: PS1 and PS2: -1000 V (the old 2-channel TENNELEC TC952) n's: -1100 V (the new 2-chan…' |
No edit summary |
||
| Line 21: | Line 21: | ||
Let's do the '''2n correlation test''' with CF252 source. The voltages and thresholds are the same. | Let's do the '''2n correlation test''' with CF252 source. The voltages and thresholds are the same. | ||
First the set up is like: | First the set up is like: | ||
:PS1 and PS2 are parallel to each other and are faced to the source. | :○ PS1 and PS2 are parallel to each other and are faced to the source. | ||
:the CF252 source in the middle | :○ the CF252 source in the middle | ||
:the distance between PS1(PS2) and source is 6 cm | :○ the distance between PS1(PS2) and source is 6 cm | ||
The total number of events (2n fission events + random coincident) are: | The total number of events (2n fission events + random coincident) are: | ||
| Line 45: | Line 45: | ||
Second the set up is like: | Second the set up is like: | ||
:PS1 and PS2 are perpendicular to each other and are faced to the source. | :○ PS1 and PS2 are perpendicular to each other and are faced to the source. | ||
:the CF252 source in the middle | :○ the CF252 source in the middle | ||
:the distance between PS1(PS2) and source is 6 cm | :○ the distance between PS1(PS2) and source is 6 cm | ||
The total number of events (2n fission events + random coincident) are: | The total number of events (2n fission events + random coincident) are: | ||
Revision as of 04:19, 13 May 2011
The powers are:
PS1 and PS2: -1000 V (the old 2-channel TENNELEC TC952) n's: -1100 V (the new 2-channel C.A.E.N N471)
The thresholds are:
PS1 and PS2: -80 mV n's: -200 mV
The discriminator widths are:
PS1: -20 ns PS2: -10 ns n's: -20 ns
Let's do the 2n correlation test with CF252 source. The voltages and thresholds are the same.
First the set up is like:
- ○ PS1 and PS2 are parallel to each other and are faced to the source.
- ○ the CF252 source in the middle
- ○ the distance between PS1(PS2) and source is 6 cm
The total number of events (2n fission events + random coincident) are:
| total counts | time | counts/min | counts/sec | |
|---|---|---|---|---|
| PS1 | [math]\displaystyle{ 255977\pm505.9 }[/math] | 10 min | [math]\displaystyle{ 25597.7\pm50.6 }[/math] | [math]\displaystyle{ 426.63\pm0.84 }[/math] |
| PS1 | [math]\displaystyle{ 421917\pm649.5 }[/math] | 10 min | [math]\displaystyle{ 42191.7\pm64.9 }[/math] | [math]\displaystyle{ 703.19\pm1.08 }[/math] |
| PS1+PS2 | [math]\displaystyle{ 1224\pm34.9 }[/math] | 10 min | [math]\displaystyle{ 122.4\pm3.5 }[/math] | [math]\displaystyle{ 2.04\pm0.06 }[/math] |
Second the set up is like:
- ○ PS1 and PS2 are perpendicular to each other and are faced to the source.
- ○ the CF252 source in the middle
- ○ the distance between PS1(PS2) and source is 6 cm
The total number of events (2n fission events + random coincident) are:
| total counts | time | counts/min | counts/sec | |
|---|---|---|---|---|
| PS1 | [math]\displaystyle{ 250116\pm500.1 }[/math] | 10 min | [math]\displaystyle{ 25011.6\pm50.0 }[/math] | [math]\displaystyle{ 416.86\pm0.83 }[/math] |
| PS1 | [math]\displaystyle{ 414774\pm644.0 }[/math] | 10 min | [math]\displaystyle{ 64.4\pm64.9 }[/math] | [math]\displaystyle{ 691.29\pm1.07 }[/math] |
| PS1+PS2 | [math]\displaystyle{ 1359\pm36.9 }[/math] | 10 min | [math]\displaystyle{ 135.9\pm3.7 }[/math] | [math]\displaystyle{ 2.26\pm0.06 }[/math] |