Revision as of 18:12, 12 May 2011
The powers are:
PS1 and PS2: -1000 V (the old 2-channel TENNELEC TC952) n's: -1100 V (the new 2-channel C.A.E.N N471)
The thresholds are:
PS1 and PS2: -80 mV n's: -200 mV
The discriminator widths are:
PS1: -20 ns PS2: -10 ns n's: -20 ns
Let's do the random coincidence test. PS1 and PS2 is about 30 cm away from each other. And let's look on the coincidence signal between PS1 and PS2:
counts/sec (5 min test), no source | |
---|---|
PS1+PS2 | 19.89 | 0.26 (total 5966 cnts)