HRRL 03-17-2011: Difference between revisions

From New IAC Wiki
Jump to navigation Jump to search
Setisadi (talk | contribs)
Setisadi (talk | contribs)
Line 66: Line 66:
=== 42 mA, Positive Current, X projection ===
=== 42 mA, Positive Current, X projection ===


emit=0.375 +- 0.001 mm*mrad, emit_norm=10.263 +- 0.033 mm*mrad  
emit=0.390 +- 0.001 mm*mrad, emit_norm=10.676 +- 0.034 mm*mrad  


beta=1.250 +- 0.000, alpha=0.936 +-0.001  
beta=1.291 +- 0.000, alpha=0.940 +-0.001  
 
[[File:HRRL 03-17-2011_Results_Q1_42mA_Pos_X.png | 300 px]]


[[File:HRRL 03-17-2011_Results_Q1_42mA_Pos_SupGau_X.png | 300 px]]




=== 42 mA, Negative Current, Y projection ===




[[File:HRRL 03-17-2011_Results_Q1_42mA_Pos_SupGau_Y.png | 300 px]]





Revision as of 16:44, 13 April 2011

Accelerator Tuning

Good Tune Used for good Scan

By Dr. Kim

Solenoid 1 6.8 A
Solenoid 2 10.4 A
Gun Ver -0.2 A
Gun Hor +0.4 A
Output Hor -1.4A
Output Ver -0.5 A
Gun HV +9.75 (Knob Setting)
Gun Grid Voltage 5.25 (Knob Setting)
RF frequency 2855.816 MHz
Modulator HV Power Supply 4.42 (Knob Setting)
RF macro Pulse Length (FWHM) 200 ns
Peak Current on FC 37.2 mAmps
Scope on FC
e- Beam Energy 14 MeV



Calibration

image OTR image cal_h (mm/px) cal_v (mm/px)
0.04308 0.04204
0.04320 0.04228
0.04349 0.04181
0.04323 0.04196
0.04337 0.04212
mean (mm/px) mean (mm/px)
[math]\displaystyle{ 0.04327 \pm 0.00016 }[/math] [math]\displaystyle{ 0.04204 \pm 0.00018 }[/math]


Results

Q1

42 mA, Positive Current, X projection

emit=0.390 +- 0.001 mm*mrad, emit_norm=10.676 +- 0.034 mm*mrad

beta=1.291 +- 0.000, alpha=0.940 +-0.001


42 mA, Negative Current, Y projection




HRRL_Emittance_Measurements_March14-18-2011