The 1 MeV gammas appear near channel 500 like the Na-22 source.  Perhaps I am not amplifying the PMT output enough to see the 0.511 gammas.
+
The 1.x MeV gammas appear near channel 1200. 
 +
 
 +
I'll do a long Na-22 run and think about upping the amplifier gain again to put the Co-60 peak around channel 3000.
  
 
==Gain==
 
==Gain==

Revision as of 16:01, 24 March 2011

Detector setup for a NaI positron detector for PePPO

NaI detector

The PMT is powered by a HV supply which should have a DC ripple less than 200 mV peak-to-peak at 3kV/3mA. A DC ripple will change the gain of the PMT and cause the energy resolution to deteriorate.

The Linearity of the detector system is another way to reduce the energy resolution. The measured PMT output should be linear with the input signal when the gain is constant. This linearity can be measured using 2 radiation sources.


Input HV: +1085 V

Input Current: 2.362 mA


Running CODA: Running CODA

Measurements Using V792 QDC

Run 1979

Start Time : March 23:12:30:26

End Time: March 23: 13:10:47

Elapsed time: 2421 seconds

1979 NaI 1185Volts.png 1979 NaI ADCHist 1185.png

Run 1980

Remove Na-22 source from top of NaI crystal.

Start Time : March 23:13:11:23

End Time: March 23: 15:56:40

Elapsed Time:6317 seconds


NaI ADC1 R1979-R1980.png


Let's lower the gain of the Phillips 777 amplifier so the 511 peak is around channel 500. This way the 1 MeV photons may come on scale.

Run 1982

Overnight Cosmic Run

Start Time : March 23: 16:01:47

End Time: March 24: 09:41:31

Elapsed Time:63558 seconds

Run 1983

Start a run using Co-60

Start Time : March 24: 09:46:25

End Time: March 24:

Elapsed Time: seconds


The 1.x MeV gammas appear near channel 1200.

I'll do a long Na-22 run and think about upping the amplifier gain again to put the Co-60 peak around channel 3000.

Gain

Linearity

Readout

F250

JLab's F250 flash ADC will be used to measure the NaI output

Na-22 source tests

North American Scientific Model CAL2600


Activity = 9.803 \muCi on 5/1/05

Serial Number 64236

Co-60 source tests

North American Scientific Model CAL2600


Activity = 9. \muCi on 5/1/05

Serial Number 64236