Difference between revisions of "TDC Calibration"
Jump to navigation
Jump to search
Line 1: | Line 1: | ||
+ | [http://wiki.iac.isu.edu/index.php/PhotoFission_with_Polarized_Photons_from_HRRL Go Back] | ||
+ | |||
We used: | We used: | ||
Line 42: | Line 44: | ||
The width of photon peak is about '''32.3 ns''' for new calibration. | The width of photon peak is about '''32.3 ns''' for new calibration. | ||
+ | |||
+ | [http://wiki.iac.isu.edu/index.php/PhotoFission_with_Polarized_Photons_from_HRRL Go Back] |
Revision as of 06:18, 5 February 2009
We used:
-detector A to calibrate all the TDCs;
-LeCroy octal discriminator to produce standard negative pulses;
-passive delay line to calibrate TDCs in the region < 100 ns;
-delay amplifier TENNELEC TC215 to calibrate TDCs in the region > 250 ns;
Range of all three TDCs was set to 50nsx100 (upper plot) and 200nsx10 (lower plot).
ADC-TAC pair was chosen in such a way that we used during the experiment.
ORTEC TAC 567-02 and ADC 1D
ORTEC TAC 567-01 and ADC 1A
ORTEC TAC 566-01 and ADC 1C
Comparison of old and new calibration coefficients
We used ADC 1D for the run pf_10_10_08_00040
The width of photon peak is about 10.3 ns for old calibration.
The width of photon peak is about 32.3 ns for new calibration.