TDC Calibration: Difference between revisions

From New IAC Wiki
Jump to navigation Jump to search
Oborn (talk | contribs)
Oborn (talk | contribs)
Line 16: Line 16:


[[Image:567_02.jpg |500px]]
[[Image:567_02.jpg |500px]]
[[Image:567_02_2000.jpg |500px]]


=ORTEC TAC 567-01 and ADC 1A=
=ORTEC TAC 567-01 and ADC 1A=

Revision as of 18:01, 7 November 2008

We used:

-detector A to calibrate all the TDCs;

-LeCroy octal discriminator to produce standard negative pulses;

-passive delay line to calibrate TDCs in the region < 100 ns;

-delay amplifier TENNELEC TC215 to calibrate TDCs in the region > 250 ns;

Range of all three TDCs was set to 50nsx100.

ADC-TAC pair was chosen in such a way that we used during the experiment.

ORTEC TAC 567-02 and ADC 1D

ORTEC TAC 567-01 and ADC 1A

ORTEC TAC 566-01 and ADC 1C


Comparison of old and new calibration coefficients

The width of photon peak is about 10.3 ns for old calibration.

The width of photon peak is about 32.3 ns for new calibration.