2-25-08

Configure 25 Mev Beam line

NaI Detector

PMT HV 1500 Co-60 1microCi source: 1MeV band at about 55mV
Co60 NaI 1500HV.jpg

PMT HV 1500 no source, noisey tube
Background NaI 1500HV.png

IAC NaI Detector Calibrated with 10 [math]\mu[/math] Ci Co60 source with high voltage power supply at 1300 Volt. 1 MeV apprx. equal to 60 mV on the scope
IAC NaI Calibration.jpg

JLAB NaI Detector Calibrated with 10 [math]\mu[/math] Ci Co60 source with high voltage power supply at 1500 Volt. 1 MeV apprx. equal to 70 mV on the scope
JLAB NaI Calibration.png

Beam Profile

1.5 inch FWHM beam profile measured at 8 mA peak current. The units on the beam scan is in inches. so the grid is 2inch by 2 inch (5cm by 5cm). Frequency is 60 Hz.

BeamProfile 2-26-08 1.jpg

_________________________________ 4 mA peak current


400 px

_______________________________________________

400 px



Analysis of exposed glass plate. Glass plate exposed to 4mA of beam current, positioned at exit of straight ahead section on D2.

Raw Image:

Scanned glass image.jpg

Projections and fits:

Plate fits.jpg

Beam on

15:20:

Turned beam on zero degree port and NaI lights up whan positioned just after the beam pipe window.

Must be gammas Rate = 10 count/sec. Anode amplified (x64) then discriminated and put into a scaler. Threshold on discriminator is 2 Volts and results in zero counts when beam off. NaI PMT at 1500 volts.

NaI Pulse 2-24-08.png

15:30:

Now move NaI detector so it is near experimental Cell wall. Just inside the wall

Gamma rate goes to zero!

NaI Pulse 2-24-08 1.png

Now change scope to 5 sec accumulation. 1 Mev electron should give 55 mV pulse on scope.


16:00:

Now try to put electrons into Experimental cell.

We turned on the first dipole to deflect 10 MeV electrons onto the Tungsten target. The quads between the two dipoles were used to minimize the rates on the NaI. Then we turned on the last dipole to deflect 3 MeV electrons into the Experimental cell. And measured the following rates

811 counts /15 sec threshold at a minimim

NaI Pulse 2-24-08 3.png

Now insert 1.2 cm thick Aluminum plate in front of the NaI in order to capture all the electrons and measure the gamma rate.

776 counts /15 sec threshold at a minimim

400 px

2-27-08

Channel 6 : IAC NaI : Co-60 source gave 20 mV pulses on scope when PMT at 1200 Volts

[math]\frac{0.010 V}{ 50 Ohms} \times \frac{200 nsec}{2} * \frac{1 count}{100 \times 10^{-15}Coul} = 200 Channels[/math]

Channel 10: JLab NaI : Co-60 source gave ~40 mV pulses on scope when PMT at 1400 Volts

[math]\frac{0.020 V}{ 50 Ohms} \times \frac{200 nsec}{2} * \frac{1 count}{100 \times 10^{-15}Coul} = 400 Channels[/math]

Positron Runs with DAQ

Run 393: 1400 V JLAB NaI, 1209 Volts on ISU NaI, Trigger: (Gun AND NaI 1 ) OR (Gun AND NaI2), 50 mV Discriminator Threshold

Run 394: 1400 V JLAB NaI, 1209 Volts on ISU NaI, Trigger: (Gun AND NaI 1 ) AND (Gun AND NaI2), 50 mV Discriminator Threshold

Run 395: 1400 V JLAB NaI, 1209 Volts on ISU NaI, Trigger: (Gun AND NaI 1 ) AND (Gun AND NaI2), 30 mV Discriminator Threshold

Run 396: 1400 V JLAB NaI, 1209 Volts on ISU NaI, Trigger: (Gun AND NaI 1 ) AND (Gun AND NaI2), 30 mV Discriminator Threshold, Increase dynode amplier from x20 to x50 to raise trigger threshold.

Run 397: 1500 V JLAB NaI, 1300 Volts on ISU NaI, Trigger: (Gun AND NaI 1 ) AND (Gun AND NaI2), 30 mV Discriminator Threshold, Increase dynode amplier at x50 to raise trigger threshold.

Run 398: 1500 V JLAB NaI, 1300 Volts on ISU NaI, Trigger: (Gun AND NaI 1 ) AND (Gun AND NaI2), 30 mV Discriminator Threshold, Increase dynode amplier from at x50 to raise trigger threshold. Insert 4" thick Aluminum in front of positron stop. SHould see rate die if we have positrons but gammas will come through.

Start time:16:03:25 Stop Time: 16:23:57

47 Count/60 seconds

Run 399: 1500 V JLAB NaI, 1300 Volts on ISU NaI, Trigger: (Gun AND NaI 1 ) AND (Gun AND NaI2), 30 mV Discriminator Threshold, Increase dynode amplier from at x50 to raise trigger threshold. Removed 4" thick Aluminum in front of positron stop.

252 counts/ 60 seconds

Changed first dipole and measured 10 mA beam current after zero degree port


Run 400: 1500 V JLAB NaI, 1300 Volts on ISU NaI, Trigger: (Gun AND NaI 1 ) AND (Gun AND NaI2), 30 mV Discriminator Threshold, Increase dynode amplier from at x50 to raise trigger threshold. Change last dipole polarity. Check electron rate .

172 counts/min


Run 401: 1500 V JLAB NaI, 1300 Volts on ISU NaI, Trigger: (Gun AND NaI 1 ) AND (Gun AND NaI2), 30 mV Discriminator Threshold, Increase dynode amplier from at x50 to raise trigger threshold. Change back to Positron polarity. Check electron rate .

242 counts/min


Histograms

The histograms below are the results of previous runs. Brick means, all the positrons were stopped at the brick and coincidence measruement was made to detect secondaries, background and noise. NOBrick means, the Al brick was removed and only 2"X2" Al absorber exist. The bottom histograms are the subtractions of brick and NOBrick histos for each NaI.

Brick and NO Brick.gif

Channel 10: ISU NaI : Co-60 source gave 60 mV pulses on scope when PMT at 1400 Volts

[math]\frac{0.030 V}{ 50 Ohms} \times \frac{200 nsec}{2} * \frac{1 count}{100 \times 10^{-15}Coul} = 600 Channels[/math]


02-29-08

He tube insterted between exit window and experimental cell. Some ISU/JLAB detector work overnight. Re-established HV and signal connections.

Run 403, 404 and 405 are no good. The ADC gate had shifted relative to the signal and needed to be re-adjusted. Gate width increased to 2[math]\mu[/math]seconds to capture complete NaI signal.

ADC plots now look similar to yesterdays.

Coincidence rate today is about 718 cnts/60secs, factor of three more than yesterday....????....

Run 406:

          Trigger: And
          Threshold: 30mV/30mV  (IAC/JLAB NaI)
          Gain in trigger: 50/50
          HV:   1300/1500

Run 407: Installed permanant magnet after last quadrupole doublet (0.7kGauss field, about 6inches long) to sweep away charged particles (electrons, positrons and pentaquarks) for background studies.

    Trigger:  And
    Threshold: 30mV/30mV
    Gain:   50/50
    HV:  1300/1500
    Coincidence Rate: 360 cnts/60secs

Run 408 : Skipped due to overflow


Run 409 : Changed the setting from coincidence to straight ahead measurement. Jlab NaI was put on the beamline.