Metalica TestPulseoutput 12-20-07: Difference between revisions

From New IAC Wiki
Jump to navigation Jump to search
Oborn (talk | contribs)
No edit summary
Oborn (talk | contribs)
Line 7: Line 7:


The voltage was changed<br>
The voltage was changed<br>
=Operating Conditions=
=PreAmp Voltage=
 
We observed the ouput of a test pule injected into the preamp for several preamp voltage setting.  The top (purple) trace is the injected pulse and the bottom  (torquoise) trace is the output from the POSTamp.
 
 
[[Image:metalica_160mv_input-postamp_out_3.7v.jpg|200px]]<br>
[[Image:metalica_160mv_input-postamp_out_3.7v.jpg|200px]]<br>
[[Image:metalica_160mv_input-postamp_out_5v.jpg|200px]]<br>
[[Image:metalica_160mv_input-postamp_out_5v.jpg|200px]]<br>

Revision as of 17:04, 21 December 2007

Tamuna and I test the DC preamp + postamp output using a pulse generator to create a 40 mV levelpulse that was 10 ns wide in order to simulate a pulse from the drift chamber proto-type.

This time we used Metalica but the results are not so different.

The bottom line appears to be that we get a x4 attenuation and not a net amplification deispite our efforts the change the configuration for the better. We also note that channels 1-11 of the postamp proto type attenuate by a factor of four but channels 12-16 have a signal independent of the input.


The voltage was changed

PreAmp Voltage

We observed the ouput of a test pule injected into the preamp for several preamp voltage setting. The top (purple) trace is the injected pulse and the bottom (torquoise) trace is the output from the POSTamp.













[1]