Difference between revisions of "MiniDC PreampOutput 10 10 07"
Jump to navigation
Jump to search
| Line 33: | Line 33: | ||
[[Image:MINIDC_Sense-Preamp_Output_10-10-07_AmpTrig1.jpg | 200 px]] | [[Image:MINIDC_Sense-Preamp_Output_10-10-07_AmpTrig1.jpg | 200 px]] | ||
| − | |||
[[Image:MINIDC_Sense-Preamp_Output_10-10-07_AmpTrig2.jpg | 200 px]] | [[Image:MINIDC_Sense-Preamp_Output_10-10-07_AmpTrig2.jpg | 200 px]] | ||
| − | |||
[[Image:MINIDC_Sense-Preamp_Output_10-10-07_AmpTrig3.jpg | 200 px]] | [[Image:MINIDC_Sense-Preamp_Output_10-10-07_AmpTrig3.jpg | 200 px]] | ||
| − | |||
[[Image:MINIDC_Sense-Preamp_Output_10-10-07_AmpTrig4.jpg | 200 px]] | [[Image:MINIDC_Sense-Preamp_Output_10-10-07_AmpTrig4.jpg | 200 px]] | ||
[[Image:MINIDC_Sense-Preamp_Output_10-10-07_AmpTrig5.jpg | 200 px]] | [[Image:MINIDC_Sense-Preamp_Output_10-10-07_AmpTrig5.jpg | 200 px]] | ||
| − | |||
[[Image:MINIDC_Sense-Preamp_Output_10-10-07_AmpTrig6.jpg | 200 px]] | [[Image:MINIDC_Sense-Preamp_Output_10-10-07_AmpTrig6.jpg | 200 px]] | ||
| − | |||
[[Image:MINIDC_Sense-Preamp_Output_10-10-07_AmpTrig7.jpg | 200 px]] | [[Image:MINIDC_Sense-Preamp_Output_10-10-07_AmpTrig7.jpg | 200 px]] | ||
| − | |||
[[Image:MINIDC_Sense-Preamp_Output_10-10-07_AmpTrig8.jpg | 200 px]] | [[Image:MINIDC_Sense-Preamp_Output_10-10-07_AmpTrig8.jpg | 200 px]] | ||
Revision as of 00:48, 16 October 2007
Drift chamber High voltage settings
| HV settings | |
| Wire | Volts |
| Sense | 1800 |
| Field | -800 |
| Guard | 800 |
I should have set the Field and Guard wires to 900 volts according to Mac's e-mail reporting his Garfield simulation results. Next time.
Scope pictures triggering on sense wire C2
I set the scope threshold to below -14 mV based on previous measurements.
Scope pictures triggering on preamp output (pin 2)
I found that a trigger threshold of -100 mV discriminated well between noise and pulses consistent with previous ionization events.