Difference between revisions of "June, 7, 2007 Cf rate measurements using Photonics PMT"
Line 40: | Line 40: | ||
Atomic composition for AISI 304) in % | Atomic composition for AISI 304) in % | ||
{| | {| | ||
− | + | ! Element || % | |
|- | |- | ||
− | + | {| | |
+ | Si || 1 | ||
|- | |- | ||
− | + | Cr ||19 | |
|- | |- | ||
− | Fe 68 | + | Mn || 2 |
+ | |- | ||
+ | Fe || 68 | ||
|- | |- | ||
− | Ni 10 | + | Ni|| 10 |
}| | }| | ||
Lead: Z = | Lead: Z = |
Revision as of 18:48, 8 June 2007
Cf source was placed 16.5 cm away from face of PHotonics PMT. The PMT output signal was descriminated by a Lecroy Model 821 descriminator set to pass any voltage below -0.5 mV. The descriminated output was then sent to a counter Model TC 531.
Configuration Definitions
1: Placed the Cf source 16.5 cm from face of Scintillator. neutrons passing through scintillator would also pass through PMT
2: Inserted a 2.5 cm thick steel shielding wall between Scintillator and Cf source. Right up against Scintillator
3: Replaced steel wall with 5 cm thick lead shielding brick
4: Replaced lead brick with 6.7 cm thick paraffin block ( Paraffin N.F. , Household wax made by Parowax. 4 cackes, total net weight 16 oz (454 g))
5: Increased thickness of paraffin block to 13.4 cm
6: No source backround
Measurements
Using Photonics PMT and BC
Configuration | Rate (counts/min) | Statistical Error in Rate (Counts/min) |
---|---|---|
1 | 2.7e3 | 0.09e3 |
2 | 2.2e3 | 0.01e3 |
3 | 0.8e3 | 0.02e3 |
4 | 2.2e3 | 0.08e3 |
5 | 1.8e3 | 0.02e3 |
6 |
Analysis
Steel : Zeff = 26 and
=8.02Atomic composition for AISI 304) in %
Element | % |
---|