Difference between revisions of "CH HPGe Efficiency"

From New IAC Wiki
Jump to navigation Jump to search
Line 17: Line 17:
 
This is total efficiency  
 
This is total efficiency  
  
<math> (line\ intensity)\ (incident\ thermal\ neutrons\ per\ second)\ (\sigma_{total})\ (MCNP\ fudge\ factor\ from\ Paul\ ~4) </math>
+
<math> (line\ intensity)\ (incident\ thermal\ neutrons\ per\ second)\ (\sigma_{total})\ (MCNP\ fudge\ factor\ from\ Paul\ ~4)\ =\ Theoretical\ Rate </math>
 
----
 
----
 
[[PAS HPGe Detector]]
 
[[PAS HPGe Detector]]

Revision as of 19:21, 28 September 2021

Eu152 Calibration

  • HPGe calibrated using Eu152 source
CH Eu152 Calibration Spectrum.png

As of 9/28/21

The calibration used to map from ADC channel number to KeV is:

[math]Photon Energy = -36.2928 + 0.449519 x - 0.000000861277 x^2[/math]

Where x = channel 9 of the ADC and is input as PADC.PADC785N[9].

Theoretical Efficiency

This is total efficiency

[math] (line\ intensity)\ (incident\ thermal\ neutrons\ per\ second)\ (\sigma_{total})\ (MCNP\ fudge\ factor\ from\ Paul\ ~4)\ =\ Theoretical\ Rate [/math]


PAS HPGe Detector