JD-766 Cs-137 Button Source Plateau
Timing filter amp configuration
Course gain: x6
Fine gain: at minimum
Integrate (nS): 500nS
Diff (nS): Out
Invert/Noninverted: Non-Inv
Plateau
700V
Source Out
- Run Number - Configuration - Discriminator Threshold (mV) - Cs-137 Button Source - Counts - Run Time (Sec) - Count Rate (Hz) - 1 - Unshielded detector at +700V - 05 - No - 958 - 60 - 15.97 - 2 - Unshielded detector at +700V - 15 - No - 278 - 60 - 4.63 - 3 - Unshielded detector at +700V - 35 - No - 107 - 60 - 1.78 - 4 - Unshielded detector at +700V - 55 - No - 85 - 60 - 1.42 - 5 - Unshielded detector at +700V - 75 - No - 80 - 60 - 1.33 - 6 - Unshielded detector at +700V - 95 - No - 80 - 60 - 1.33 - 7 - Unshielded detector at +700V - 115 - No - 75 - 60 - 1.25 - 8 - Unshielded detector at +700V - 135 - No - 78 - 60 - 1.3 - 9 - Unshielded detector at +700V - 155 - No - 82 - 60 - 1.37 - 10 - Unshielded detector at +700V - 175 - No - 77 - 60 - 1.28 - 11 - Unshielded detector at +700V - 195 - No - 66 - 60 - 1.1 - 12 - Unshielded detector at +700V - 215 - No - 68 - 60 - 1.13 - 13 - Unshielded detector at +700V - 235 - No - 53 - 60 - 0.88 - 14 - Unshielded detector at +700V - 255 - No - 59 - 60 - 0.98 
Source In
- Run Number - Configuration - Discriminator Threshold (mV) - Cs-137 Button Source - Counts - Run Time (Sec) - Count Rate (Hz) - 1 - Unshielded detector at +700V - 05 - Yes - 1033 - 60 - 2 - Unshielded detector at +700V - 15 - Yes - 291 - 60 - 3 - Unshielded detector at +700V - 35 - Yes - 102 - 60 - 4 - Unshielded detector at +700V - 55 - Yes - 87 - 60 - 5 - Unshielded detector at +700V - 75 - Yes - 88 - 60 - 6 - Unshielded detector at +700V - 95 - Yes - 103 - 60 - 7 - Unshielded detector at +700V - 115 - Yes - 101 - 60 - 8 - Unshielded detector at +700V - 135 - Yes - 84 - 60 - 9 - Unshielded detector at +700V - 155 - Yes - 72 - 60 - 10 - Unshielded detector at +700V - 175 - Yes - 57 - 60 - 11 - Unshielded detector at +700V - 195 - Yes - 72 - 60 - 12 - Unshielded detector at +700V - 215 - Yes - 68 - 60 - 13 - Unshielded detector at +700V - 235 - Yes - 74 - 60 - 14 - Unshielded detector at +700V - 255 - Yes - 52 - 60 
Plateau Plot
750V
Source Out
- Run Number - Configuration - Discriminator Threshold (mV) - Cs-137 Button Source - Counts - Run Time (Sec) - Count Rate (Hz) - 1 - Unshielded detector at +750V - 05 - No - 2211 - 60 - 36.85 - 2 - Unshielded detector at +750V - 15 - No - 1125 - 60 - 18.75 - 3 - Unshielded detector at +750V - 35 - No - 293 - 60 - 4.88 - 4 - Unshielded detector at +750V - 55 - No - 105 - 60 - 1.75 - 5 - Unshielded detector at +750V - 75 - No - 80 - 60 - 1.33 - 6 - Unshielded detector at +750V - 95 - No - 97 - 60 - 1.62 - 7 - Unshielded detector at +750V - 115 - No - 99 - 60 - 1.65 - 8 - Unshielded detector at +750V - 135 - No - 89 - 60 - 1.48 - 9 - Unshielded detector at +750V - 155 - No - 74 - 60 - 1.23 - 10 - Unshielded detector at +750V - 175 - No - 80 - 60 - 1.33 - 11 - Unshielded detector at +750V - 195 - No - 80 - 60 - 1.33 - 12 - Unshielded detector at +750V - 215 - No - 64 - 60 - 1.07 - 13 - Unshielded detector at +750V - 235 - No - 89 - 60 - 1.48 - 14 - Unshielded detector at +750V - 255 - No - 69 - 60 - 1.15 
Source In
- Run Number - Configuration - Discriminator Threshold (mV) - Cs-137 Button Source - Counts - Run Time (Sec) - Count Rate (Hz) - 1 - Unshielded detector at +750V - 05 - Yes - 4055 - 60 - 2 - Unshielded detector at +750V - 15 - Yes - 1160 - 60 - 3 - Unshielded detector at +750V - 35 - Yes - 301 - 60 - 4 - Unshielded detector at +750V - 55 - Yes - 102 - 60 - 5 - Unshielded detector at +750V - 75 - Yes - 82 - 60 - 6 - Unshielded detector at +750V - 95 - Yes - 114 - 60 - 7 - Unshielded detector at +750V - 115 - Yes - 85 - 60 - 8 - Unshielded detector at +750V - 135 - Yes - 93 - 60 - 9 - Unshielded detector at +750V - 155 - Yes - 87 - 60 - 10 - Unshielded detector at +750V - 175 - Yes - 87 - 60 - 11 - Unshielded detector at +750V - 195 - Yes - 79 - 60 - 12 - Unshielded detector at +750V - 215 - Yes - 90 - 60 - 13 - Unshielded detector at +750V - 235 - Yes - 84 - 60 - 14 - Unshielded detector at +750V - 255 - Yes - 73 - 60 
Plateau Plot
800V
Source Out
- Run Number - Configuration - Discriminator Threshold (mV) - Cs-137 Button Source - Counts - Run Time (Sec) - Count Rate (Hz) - 1 - Unshielded detector at +800V - 05 - No - 3866 - 60 - 64.43 - 2 - Unshielded detector at +800V - 15 - No - 2356 - 60 - 39.27 - 3 - Unshielded detector at +800V - 35 - No - 1091 - 60 - 18.18 - 4 - Unshielded detector at +800V - 55 - No - 384 - 60 - 6.4 - 5 - Unshielded detector at +800V - 75 - No - 244 - 60 - 4.07 - 6 - Unshielded detector at +800V - 95 - No - 167 - 60 - 2.78 - 7 - Unshielded detector at +800V - 115 - No - 100 - 60 - 1.67 - 8 - Unshielded detector at +800V - 135 - No - 88 - 60 - 1.47 - 9 - Unshielded detector at +800V - 155 - No - 89 - 60 - 1.48 - 10 - Unshielded detector at +800V - 175 - No - 86 - 60 - 1.43 - 11 - Unshielded detector at +800V - 195 - No - 90 - 60 - 1.5 - 12 - Unshielded detector at +800V - 215 - No - 81 - 60 - 1.35 - 13 - Unshielded detector at +800V - 235 - No - 95 - 60 - 1.58 - 14 - Unshielded detector at +800V - 255 - No - 89 - 60 - 1.48 
Source In
- Run Number - Configuration - Discriminator Threshold (mV) - Cs-137 Button Source - Counts - Run Time (Sec) - Count Rate (Hz) - 1 - Unshielded detector at +800V - 05 - Yes - 10055 - 60 - 2 - Unshielded detector at +800V - 15 - Yes - 5438 - 60 - 3 - Unshielded detector at +800V - 35 - Yes - 1078 - 60 - 4 - Unshielded detector at +800V - 55 - Yes - 375 - 60 - 5 - Unshielded detector at +800V - 75 - Yes - 217 - 60 - 6 - Unshielded detector at +800V - 95 - Yes - 126 - 60 - 7 - Unshielded detector at +800V - 115 - Yes - 100 - 60 - 8 - Unshielded detector at +800V - 135 - Yes - 93 - 60 - 9 - Unshielded detector at +800V - 155 - Yes - 99 - 60 - 10 - Unshielded detector at +800V - 175 - Yes - 96 - 60 - 11 - Unshielded detector at +800V - 195 - Yes - 102 - 60 - 12 - Unshielded detector at +800V - 215 - Yes - 99 - 60 - 13 - Unshielded detector at +800V - 235 - Yes - 62 - 60 - 14 - Unshielded detector at +800V - 255 - Yes - 85 - 60 
Plateau Plot
850V
Source Out
- Run Number - Configuration - Discriminator Threshold (mV) - Cs-137 Button Source - Counts - Run Time (Sec) - Count Rate (Hz) - 1 - Unshielded detector at +850V - 05 - No - 6402 - 60 - 106.7 - 2 - Unshielded detector at +850V - 15 - No - 3867 - 60 - 64.45 - 3 - Unshielded detector at +850V - 35 - No - 2058 - 60 - 34.3 - 4 - Unshielded detector at +850V - 55 - No - 1233 - 60 - 20.55 - 5 - Unshielded detector at +850V - 75 - No - 787 - 60 - 13.12 - 6 - Unshielded detector at +850V - 95 - No - 391 - 60 - 6.52 - 7 - Unshielded detector at +850V - 115 - No - 256 - 60 - 4.27 - 8 - Unshielded detector at +850V - 135 - No - 193 - 60 - 3.22 - 9 - Unshielded detector at +850V - 155 - No - 151 - 60 - 2.52 - 10 - Unshielded detector at +850V - 175 - No - 97 - 60 - 1.62 - 11 - Unshielded detector at +850V - 195 - No - 105 - 60 - 1.75 - 12 - Unshielded detector at +850V - 215 - No - 97 - 60 - 1.62 - 13 - Unshielded detector at +850V - 235 - No - 87 - 60 - 1.45 - 14 - Unshielded detector at +850V - 255 - No - 101 - 60 - 1.68 
Source In
- Run Number - Configuration - Discriminator Threshold (mV) - Cs-137 Button Source - Counts - Run Time (Sec) - Count Rate (Hz) - 1 - Unshielded detector at +850V - 05 - Yes - 15370 - 60 - 2 - Unshielded detector at +850V - 15 - Yes - 10298 - 60 - 3 - Unshielded detector at +850V - 35 - Yes - 2963 - 60 - 4 - Unshielded detector at +850V - 55 - Yes - 1366 - 60 - 5 - Unshielded detector at +850V - 75 - Yes - 842 - 60 - 6 - Unshielded detector at +850V - 95 - Yes - 402 - 60 - 7 - Unshielded detector at +850V - 115 - Yes - 281 - 60 - 8 - Unshielded detector at +850V - 135 - Yes - 196 - 60 - 9 - Unshielded detector at +850V - 155 - Yes - 143 - 60 - 10 - Unshielded detector at +850V - 175 - Yes - 98 - 60 - 11 - Unshielded detector at +850V - 195 - Yes - 88 - 60 - 12 - Unshielded detector at +850V - 215 - Yes - 101 - 60 - 13 - Unshielded detector at +850V - 235 - Yes - 104 - 60 - 14 - Unshielded detector at +850V - 255 - Yes - 101 - 60 
Plateau Plot
1100V
Source Out
- Run Number - Configuration - Discriminator Threshold (mV) - Cs-137 Button Source - Counts - Run Time (Sec) - Count Rate (Hz) - 1 - Unshielded detector at +1300V - 05 - No - 24651 - 60 - 2 - Unshielded detector at +1300V - 10 - No - 22487 - 60 - 3 - Unshielded detector at +1300V - 15 - No - 21029 - 60 - 4 - Unshielded detector at +1300V - 20 - No - 20535 - 60 - 5 - Unshielded detector at +1300V - 25 - No - 19313 - 60 - 6 - Unshielded detector at +1300V - 30 - No - 18947 - 60 - 7 - Unshielded detector at +1300V - 35 - No - 18654 - 60 - 8 - Unshielded detector at +1300V - 45 - No - 18405 - 60 - 9 - Unshielded detector at +1300V - 55 - No - 18078 - 60 - 10 - Unshielded detector at +1300V - 75 - No - 17463 - 60 - 11 - Unshielded detector at +1300V - 95 - No - 17276 - 60 - 12 - Unshielded detector at +1300V - 115 - No - 16801 - 60 - 13 - Unshielded detector at +1300V - 135 - No - 16165 - 60 - 14 - Unshielded detector at +1300V - 155 - No - 15329 - 60 - 15 - Unshielded detector at +1300V - 175 - No - 14452 - 60 - 16 - Unshielded detector at +1300V - 195 - No - 14130 - 60 - 17 - Unshielded detector at +1300V - 215 - No - 12954 - 60 - 18 - Unshielded detector at +1300V - 235 - No - 12155 - 60 - 19 - Unshielded detector at +1300V - 255 - No - 11392 - 60 
Source In
- Run Number - Configuration - Discriminator Threshold (mV) - Cs-137 Button Source - Counts - Run Time (Sec) - Count Rate (Hz) - 1 - Unshielded detector at +1300V - 05 - Yes - 36726 - 60 - 2 - Unshielded detector at +1300V - 10 - Yes - 33627 - 60 - 3 - Unshielded detector at +1300V - 15 - Yes - 31514 - 60 - 4 - Unshielded detector at +1300V - 20 - Yes - 30182 - 60 - 5 - Unshielded detector at +1300V - 25 - Yes - 29448 - 60 - 6 - Unshielded detector at +1300V - 30 - Yes - 28211 - 60 - 7 - Unshielded detector at +1300V - 35 - Yes - 27912 - 60 - 8 - Unshielded detector at +1300V - 45 - Yes - 26890 - 60 - 9 - Unshielded detector at +1300V - 55 - Yes - 26347 - 60 - 10 - Unshielded detector at +1300V - 75 - Yes - 25154 - 60 - 11 - Unshielded detector at +1300V - 95 - Yes - 25104 - 60 - 12 - Unshielded detector at +1300V - 115 - Yes - 24944 - 60 - 13 - Unshielded detector at +1300V - 135 - Yes - 23766 - 60 - 14 - Unshielded detector at +1300V - 155 - Yes - 23158 - 60 - 15 - Unshielded detector at +1300V - 175 - Yes - 22176 - 60 - 16 - Unshielded detector at +1300V - 195 - Yes - 21476 - 60 - 17 - Unshielded detector at +1300V - 215 - Yes - 20375 - 60 - 18 - Unshielded detector at +1300V - 235 - Yes - 19628 - 60 - 19 - Unshielded detector at +1300V - 255 - Yes - 18512 - 60 
Plateau Plot
1200V
Source Out
- Run Number - Configuration - Discriminator Threshold (mV) - Cs-137 Button Source - Counts - Run Time (Sec) - Count Rate (Hz) - 1 - Unshielded detector at +1200V - 05 - No - 24052 - 60 - 2 - Unshielded detector at +1200V - 15 - No - 20684 - 60 - 3 - Unshielded detector at +1200V - 35 - No - 18324 - 60 - 4 - Unshielded detector at +1200V - 55 - No - 17266 - 60 - 5 - Unshielded detector at +1200V - 75 - No - 15643 - 60 - 6 - Unshielded detector at +1200V - 95 - No - 14223 - 60 - 7 - Unshielded detector at +1200V - 115 - No - 12792 - 60 - 8 - Unshielded detector at +1200V - 135 - No - 11691 - 60 - 9 - Unshielded detector at +1200V - 155 - No - 10414 - 60 - 10 - Unshielded detector at +1200V - 175 - No - 9245 - 60 - 11 - Unshielded detector at +1200V - 195 - No - 8574 - 60 - 12 - Unshielded detector at +1200V - 215 - No - 7665 - 60 - 13 - Unshielded detector at +1200V - 235 - No - 7008 - 60 - 14 - Unshielded detector at +1200V - 255 - No - 6386 - 60 
Source In
- Run Number - Configuration - Discriminator Threshold (mV) - Cs-137 Button Source - Counts - Run Time (Sec) - Count Rate (Hz) - 1 - Unshielded detector at +1200V - 05 - Yes - 39450 - 60 - 2 - Unshielded detector at +1200V - 15 - Yes - 34061 - 60 - 3 - Unshielded detector at +1200V - 35 - Yes - 29029 - 60 - 4 - Unshielded detector at +1200V - 55 - Yes - 27672 - 60 - 5 - Unshielded detector at +1200V - 75 - Yes - 25935 - 60 - 6 - Unshielded detector at +1200V - 95 - Yes - 24104 - 60 - 7 - Unshielded detector at +1200V - 115 - Yes - 22223 - 60 - 8 - Unshielded detector at +1200V - 135 - Yes - 20670 - 60 - 9 - Unshielded detector at +1200V - 155 - Yes - 19168 - 60 - 10 - Unshielded detector at +1200V - 175 - Yes - 17920 - 60 - 11 - Unshielded detector at +1200V - 195 - Yes - 16383 - 60 - 12 - Unshielded detector at +1200V - 215 - Yes - 15066 - 60 - 13 - Unshielded detector at +1200V - 235 - Yes - 14130 - 60 - 14 - Unshielded detector at +1200V - 255 - Yes - 13382 - 60 
Plateau Plot
1300V
Source Out
- Run Number - Configuration - Discriminator Threshold (mV) - Cs-137 Button Source - Counts - Run Time (Sec) - Count Rate (Hz) - 1 - Unshielded detector at +1300V - 05 - No - 24651 - 60 - 2 - Unshielded detector at +1300V - 10 - No - 22487 - 60 - 3 - Unshielded detector at +1300V - 15 - No - 21029 - 60 - 4 - Unshielded detector at +1300V - 20 - No - 20535 - 60 - 5 - Unshielded detector at +1300V - 25 - No - 19313 - 60 - 6 - Unshielded detector at +1300V - 30 - No - 18947 - 60 - 7 - Unshielded detector at +1300V - 35 - No - 18654 - 60 - 8 - Unshielded detector at +1300V - 45 - No - 18405 - 60 - 9 - Unshielded detector at +1300V - 55 - No - 18078 - 60 - 10 - Unshielded detector at +1300V - 75 - No - 17463 - 60 - 11 - Unshielded detector at +1300V - 95 - No - 17276 - 60 - 12 - Unshielded detector at +1300V - 115 - No - 16801 - 60 - 13 - Unshielded detector at +1300V - 135 - No - 16165 - 60 - 14 - Unshielded detector at +1300V - 155 - No - 15329 - 60 - 15 - Unshielded detector at +1300V - 175 - No - 14452 - 60 - 16 - Unshielded detector at +1300V - 195 - No - 14130 - 60 - 17 - Unshielded detector at +1300V - 215 - No - 12954 - 60 - 18 - Unshielded detector at +1300V - 235 - No - 12155 - 60 - 19 - Unshielded detector at +1300V - 255 - No - 11392 - 60 
Source In
- Run Number - Configuration - Discriminator Threshold (mV) - Cs-137 Button Source - Counts - Run Time (Sec) - Count Rate (Hz) - 1 - Unshielded detector at +1300V - 05 - Yes - 36726 - 60 - 2 - Unshielded detector at +1300V - 10 - Yes - 33627 - 60 - 3 - Unshielded detector at +1300V - 15 - Yes - 31514 - 60 - 4 - Unshielded detector at +1300V - 20 - Yes - 30182 - 60 - 5 - Unshielded detector at +1300V - 25 - Yes - 29448 - 60 - 6 - Unshielded detector at +1300V - 30 - Yes - 28211 - 60 - 7 - Unshielded detector at +1300V - 35 - Yes - 27912 - 60 - 8 - Unshielded detector at +1300V - 45 - Yes - 26890 - 60 - 9 - Unshielded detector at +1300V - 55 - Yes - 26347 - 60 - 10 - Unshielded detector at +1300V - 75 - Yes - 25154 - 60 - 11 - Unshielded detector at +1300V - 95 - Yes - 25104 - 60 - 12 - Unshielded detector at +1300V - 115 - Yes - 24944 - 60 - 13 - Unshielded detector at +1300V - 135 - Yes - 23766 - 60 - 14 - Unshielded detector at +1300V - 155 - Yes - 23158 - 60 - 15 - Unshielded detector at +1300V - 175 - Yes - 22176 - 60 - 16 - Unshielded detector at +1300V - 195 - Yes - 21476 - 60 - 17 - Unshielded detector at +1300V - 215 - Yes - 20375 - 60 - 18 - Unshielded detector at +1300V - 235 - Yes - 19628 - 60 - 19 - Unshielded detector at +1300V - 255 - Yes - 18512 - 60 
Plateau Plot