=Calculated Dose vs PMT Counts= | =Calculated Dose vs PMT Counts= | ||
− | These fits were created using OSLs I irradiated using the | + | These fits were created using OSLs I irradiated using the 9.3Ci <math> ^{137}Cs </math> Source. I calculated the dose the OSLs should have on them using the dose calculations on the previous page. The PMT counts came straight from the OSL reader. |
Number of degrees of freedom is data points - fit parameters | Number of degrees of freedom is data points - fit parameters |
Revision as of 18:37, 5 January 2018
Initial Fits
These fits are listed dose (from Landauer) vs measured dose (from OSL reader).
Calculated Dose vs PMT Counts
These fits were created using OSLs I irradiated using the 9.3Ci
Source. I calculated the dose the OSLs should have on them using the dose calculations on the previous page. The PMT counts came straight from the OSL reader.Number of degrees of freedom is data points - fit parameters
I should note that the low dose calibration factor from the OSL reader is 9.496494 and 0.6806084 for the high dose. These values are close to the slopes of their respective linear fits.
Up to date fit