These fits were created using OSLs I irradiated using the <math> ^{137}Cs </math> Source. I calculated the dose the OSLs should have on them using the dose calculations on the previous page. The PMT counts came straight from the OSL reader.  
+
These fits were created using OSLs I irradiated using the 20Ci <math> ^{137}Cs </math> Source. I calculated the dose the OSLs should have on them using the dose calculations on the previous page. The PMT counts came straight from the OSL reader.  
  
 
[[File:Low_Dose_OSL_Cal.png|400px|thumb|left|<math> {\chi}^2 = 0.31801</math>, <math> \text{# of Degrees of Freedom} = 2</math> ]]
 
[[File:Low_Dose_OSL_Cal.png|400px|thumb|left|<math> {\chi}^2 = 0.31801</math>, <math> \text{# of Degrees of Freedom} = 2</math> ]]

Revision as of 18:16, 3 November 2017

Initial Fits

These fits are expected vs measured dose.

This fit was calculated using only the OSLs provided by Landauer for calibrating the OSL reader. χ2=359.191 # of Degrees of Freedom=13
This fit was calculated using the OSLs provided by Landauer and 6 Nanodot OSLs dosed by using the 137Cs source. χ2=686.664 # of Degrees of Freedom=19

Calculated Dose vs PMT Counts

These fits were created using OSLs I irradiated using the 20Ci 137Cs Source. I calculated the dose the OSLs should have on them using the dose calculations on the previous page. The PMT counts came straight from the OSL reader.

χ2=0.31801, # of Degrees of Freedom=2