Difference between revisions of "Dose Calculations and 137Cs Shot Log"
Jump to navigation
Jump to search
(→Run 2) |
|||
Line 1: | Line 1: | ||
=Source Calibration and Dose Rate Data= | =Source Calibration and Dose Rate Data= | ||
− | {| border="2" style="text-align:center;" |cellpadding="20" cellspacing="0 | + | :{| border="2" style="text-align:center;" |cellpadding="20" cellspacing="0 |
|- | |- | ||
| Date || Initial Source Strength (Ci) || Current Source Strength (Ci) || Distance to Faceplate (cm) || Exposure Rate (Roentgen/hr)° | | Date || Initial Source Strength (Ci) || Current Source Strength (Ci) || Distance to Faceplate (cm) || Exposure Rate (Roentgen/hr)° | ||
Line 15: | Line 15: | ||
=137Cs Shot Log= | =137Cs Shot Log= | ||
==Dot OSLs== | ==Dot OSLs== | ||
− | {| border="2" style="text-align:center;" |cellpadding="20" cellspacing="0 | + | :{| border="2" style="text-align:center;" |cellpadding="20" cellspacing="0 |
|- | |- | ||
| Date || Run Number || Run Time (s) || Distance to Faceplate || Number of OSLs in Stack || Exposure Rate (R/hr) || Total Exposed Dose (mR) | | Date || Run Number || Run Time (s) || Distance to Faceplate || Number of OSLs in Stack || Exposure Rate (R/hr) || Total Exposed Dose (mR) | ||
Line 35: | Line 35: | ||
==Nanodot OSL Quality Control== | ==Nanodot OSL Quality Control== | ||
===Calibration OSLs=== | ===Calibration OSLs=== | ||
− | {| border="2" style="text-align:center;" |cellpadding="20" cellspacing="0 | + | :{| border="2" style="text-align:center;" |cellpadding="20" cellspacing="0 |
|- | |- | ||
|OSL # || Listed Dose (mRad) || Measured Dose (mRad) || Average Dose (mRad) | |OSL # || Listed Dose (mRad) || Measured Dose (mRad) || Average Dose (mRad) | ||
Line 72: | Line 72: | ||
===QC OSLs=== | ===QC OSLs=== | ||
− | {| border="2" style="text-align:center;" |cellpadding="20" cellspacing="0 | + | :{| border="2" style="text-align:center;" |cellpadding="20" cellspacing="0 |
|- | |- | ||
|OSL # || Listed Dose (mRad) || Measured Dose (mRad) || Average Dose (mRad) | |OSL # || Listed Dose (mRad) || Measured Dose (mRad) || Average Dose (mRad) | ||
Line 96: | Line 96: | ||
===Run 1=== | ===Run 1=== | ||
− | + | :{| border="2" style="text-align:center;"|cellpadding="20" cellspacing="0 | |
|- | |- | ||
| Date || Run Time (s) || Distance to Faceplate || Number of OSLs || Total Exposed Dose (mR) | | Date || Run Time (s) || Distance to Faceplate || Number of OSLs || Total Exposed Dose (mR) | ||
Line 105: | Line 105: | ||
− | + | :{| border="2" style="text-align:center;" |cellpadding="20" cellspacing="0 | |
|- | |- | ||
|OSL # || Listed Dose (mRad) || Measured Dose (mRad) || Average Dose (mRad) | |OSL # || Listed Dose (mRad) || Measured Dose (mRad) || Average Dose (mRad) | ||
Line 119: | Line 119: | ||
===Run 2=== | ===Run 2=== | ||
− | + | :{| border="2" style="text-align:center;" |cellpadding="20" cellspacing="0 | |
|- | |- | ||
| Date || Run Time (s) || Distance to Faceplate || Number of OSLs || Total Exposed Dose (mR) | | Date || Run Time (s) || Distance to Faceplate || Number of OSLs || Total Exposed Dose (mR) | ||
Line 128: | Line 128: | ||
− | + | :{| border="2" style="text-align:center;" |cellpadding="20" cellspacing="0 | |
|- | |- | ||
|OSL # || Listed Dose (mRad) || Measured Dose (mRad) || Average Dose (mRad) | |OSL # || Listed Dose (mRad) || Measured Dose (mRad) || Average Dose (mRad) |
Revision as of 20:00, 16 October 2017
Source Calibration and Dose Rate Data
Date Initial Source Strength (Ci) Current Source Strength (Ci) Distance to Faceplate (cm) Exposure Rate (Roentgen/hr)° 6/27/17 20 9.3 30.5 17.691 9/11/17 20 9.3 30.5 17.606
°Calculated by TSO using
where , ,There is an adjustment distance of 11.2cm which accounts for the distance from the faceplate to the
source137Cs Shot Log
Dot OSLs
Date Run Number Run Time (s) Distance to Faceplate Number of OSLs in Stack Exposure Rate (R/hr) Total Exposed Dose (mR) 9/8/17 1 5660 30cm 11 18.7188 30148.8 9/8/17 1 5660 30cm 11 18.7188 30148.8 9/26/17 2 4500 30cm 11 18.7188 23398.6 9/26/17 2 4500 30cm 11 18.7188 23398.6
Total dose calculated by
In these calculations,
Nanodot OSL Quality Control
Calibration OSLs
OSL # Listed Dose (mRad) Measured Dose (mRad) Average Dose (mRad) DN08276339J Unexposed 71.19332 \ DN08276683L Unexposed 70.39211 72.10899 DN08258554P Unexposed 74.74154 / DN10806640H 500 687.3246 \ DN10608908Z 500 681.0294 683.2422333 DN10806620J 500 681.3727 / DN105057582 3000 3563.558 \ DN104085766 3000 3522.352 3567.258333 DN10608909X 3000 3615.865 / DN108066085 50000 55966.55 \ DN103091798 50000 55420.36 55622.12 DN104086665 50000 55479.45 / DN10606601L 100000 111610.5 \ DN106066095 100000 114271.2 114191.3333 DN10712163M 100000 116692.3 /
QC OSLs
OSL # Listed Dose (mRad) Measured Dose (mRad) Average Dose (mRad) DN08264870R Unexposed 69.36198 \ DN082701640 Unexposed 77.25963 74.74153667 DN08269802P Unexposed 77.603 / DN10712168C 1000 1235.238 \ DN10309170Q 1000 1223.106 1226.119667 DN102095379 1000 1220.015 /
Total dose calculated by
Nanodot OSLS
Run 1
Date Run Time (s) Distance to Faceplate Number of OSLs Total Exposed Dose (mR) 10/12/17 284.04 30.5cm 3 1476.92
OSL # Listed Dose (mRad) Measured Dose (mRad) Average Dose (mRad) DN08252102C Unexposed 1161.413 \ DN08241536W Unexposed 1064.58 1148.9367 DN08229824R Unexposed 1220.817 /
Run 2
Date Run Time (s) Distance to Faceplate Number of OSLs Total Exposed Dose (mR) 10/12/17 284.04 30.5cm 3 1476.92
OSL # Listed Dose (mRad) Measured Dose (mRad) Average Dose (mRad) DN08252102C Unexposed 1161.413 \ DN08241536W Unexposed 1064.58 1148.9367 DN08229824R Unexposed 1220.817 /