=Actual Efficiency= | =Actual Efficiency= | ||
+ | There is a small issue with the calibration when plotting the results in ROOT. The binning along with the calibration create a picket fencing effect that incorrectly displays the results. The effect is shown below. | ||
+ | |||
{| border="3" cellpadding="5" cellspacing="0" | {| border="3" cellpadding="5" cellspacing="0" | ||
| Run || Source || Energy (keV) || Position || Expected Rate (Hz) ||ROOT Window || HpGe Rate (Hz) || HpGe Detector A Efficiency (%) | | Run || Source || Energy (keV) || Position || Expected Rate (Hz) ||ROOT Window || HpGe Rate (Hz) || HpGe Detector A Efficiency (%) | ||
|- | |- | ||
− | ||Ba133_DetA_Farthest_DetB_posN_Efficiency.root || Ba-133 || Farthest Back Position || 71455.51 || | + | ||Ba133_DetA_Farthest_DetB_posN_Efficiency.root || Ba-133 || 80.9979 || Farthest Back Position || 71455.51 || |
|- | |- | ||
|} | |} |
Revision as of 17:58, 25 August 2017
DetB Calibration is A = -2.8392 +/- 0.592828 , B = 0.815444 +/- 0.000630482
Det A Intercept = -12.208800 slope =1.021270
Theoretical Activities
Ba-133, 10.54 uCi on July 1, 2008, half life 10.51 years, 32.9% for 80.9979 keV line, 18.34% for 302.8508 keV line, 62.05% for the 356.0129 keV line. The sample serial # 129791 had an activity of 5.87 uCi on May 27, 2017
- = for the 80.9979 keV line
- = for the 302.8508 keV line
- = for the 302.8508 keV line
Actual Efficiency
There is a small issue with the calibration when plotting the results in ROOT. The binning along with the calibration create a picket fencing effect that incorrectly displays the results. The effect is shown below.
Run | Source | Energy (keV) | Position | Expected Rate (Hz) | ROOT Window | HpGe Rate (Hz) | HpGe Detector A Efficiency (%) |
Ba133_DetA_Farthest_DetB_posN_Efficiency.root | Ba-133 | 80.9979 | Farthest Back Position | 71455.51 |