A observable decrease in the energy resolution of a large HPGe was first seen after the irradiation of 5*10^7 n/cm^2<ref>P. H. Stelson, J. K. Dickens, S. Raman, and R. C. Trammell, “Deterioration of Large Ge(Li) Diodes Caused by Fast Neutrons,” Nuclear Instruments and Methods 98,481 (1972).</ref>, so 5*10^6 n/cm^2 is a conservative number to stay under.  
 
A observable decrease in the energy resolution of a large HPGe was first seen after the irradiation of 5*10^7 n/cm^2<ref>P. H. Stelson, J. K. Dickens, S. Raman, and R. C. Trammell, “Deterioration of Large Ge(Li) Diodes Caused by Fast Neutrons,” Nuclear Instruments and Methods 98,481 (1972).</ref>, so 5*10^6 n/cm^2 is a conservative number to stay under.  
  
The maximum flux occurs right at the center of the detector, where the expression for flux simplifies to <math>n_{rate}\frac{1}{4\pi d^2}</math>.  The number of days it would take to reach an integral flux of <math>5\times 10^6 n/cm^2 </math> as a function of the CF-252 source distance HPGe face is shown below. The graph assumes the activity of the Cf252 source on 01/2017, which is 19066 n/s.  
+
The maximum neutron flux occurs right at the center of the detector, where the expression for flux is simply: <math>n_{rate}\frac{1}{4\pi d^2}</math>.   
 +
 
 +
The number of days it would take to reach an integral flux of 5*10^6 n/cm^2 as a function of the source to HPGe face distance is shown below. The graph assumes the activity of the Cf252 source as of 01/2017, which is 19,066 n/s.  
  
  

Revision as of 06:17, 29 December 2016

A observable decrease in the energy resolution of a large HPGe was first seen after the irradiation of 5*10^7 n/cm^2<ref>P. H. Stelson, J. K. Dickens, S. Raman, and R. C. Trammell, “Deterioration of Large Ge(Li) Diodes Caused by Fast Neutrons,” Nuclear Instruments and Methods 98,481 (1972).</ref>, so 5*10^6 n/cm^2 is a conservative number to stay under.

The maximum neutron flux occurs right at the center of the detector, where the expression for flux is simply: [math]n_{rate}\frac{1}{4\pi d^2}[/math].

The number of days it would take to reach an integral flux of 5*10^6 n/cm^2 as a function of the source to HPGe face distance is shown below. The graph assumes the activity of the Cf252 source as of 01/2017, which is 19,066 n/s.