Difference between revisions of "03-27-2015"
Jump to navigation
Jump to search
(Created page with "Investigation of the nature of the secon peak in the second hit mode.") |
|||
(20 intermediate revisions by the same user not shown) | |||
Line 1: | Line 1: | ||
Investigation of the nature of the secon peak in the second hit mode. | Investigation of the nature of the secon peak in the second hit mode. | ||
+ | |||
+ | 1. LED is in non-updating mode, 30 mV threshold, 40 ns pulse width: | ||
+ | |||
+ | [[File:secon_hit_det1_30mV_LED.png | 400 px]] | ||
+ | |||
+ | 2. LED is in updating mode (in hope to get rid of the second pulse), 30 mV threshold, 40 ns pulse width: | ||
+ | |||
+ | Second pulse is still there. | ||
+ | |||
+ | [[File:secon_hit_det1_30mV_LED_updating.png | 400 px]] | ||
+ | |||
+ | Another case of the second pulse appearance (same conditions): | ||
+ | |||
+ | [[File:secon_hit_det1_30mV_LED_updating2.png | 400 px]] | ||
+ | |||
+ | |||
+ | {| border="1" | ||
+ | | Run # (multihit data)|| Description | ||
+ | |- | ||
+ | |4502 || Cosmic rays test, DAQ is triggered by Det #1 (bottom, multihit TDC9), HV(Det#1) was reduced to -1150 V, LED threshold on the signal coming from the top of Det#1 (multihit TDC8) was set to 255 mV (30 mV all others) and fed into CAEN N93B Dual Timer. The pulse width was increased by the timer to 60 ns. Let's see where in the timing spectrum of the second hits we can observe the spikes due to the ringing. | ||
+ | |||
+ | [[File:cosmic_runtest_increased_stopwidth.png | 400 px]] | ||
+ | |||
+ | First hit (Det1 top): | ||
+ | |||
+ | [[File:cosmic_runtest_first_hit.png | 400 px]] | ||
+ | |||
+ | Second hit (Det1 top): | ||
+ | |||
+ | [[File:cosmic_runtest_second_hit.png | 400 px]] | ||
+ | |||
+ | Second hit (Det1 top) - First hit (Det1 top): | ||
+ | |||
+ | [[File:cosmic_runtest_T8_2_T8_1.png | 400 px]] | ||
+ | |||
+ | |} |
Latest revision as of 00:44, 30 March 2015
Investigation of the nature of the secon peak in the second hit mode.
1. LED is in non-updating mode, 30 mV threshold, 40 ns pulse width:
2. LED is in updating mode (in hope to get rid of the second pulse), 30 mV threshold, 40 ns pulse width:
Second pulse is still there.
Another case of the second pulse appearance (same conditions):
Run # (multihit data) | Description |
4502 | Cosmic rays test, DAQ is triggered by Det #1 (bottom, multihit TDC9), HV(Det#1) was reduced to -1150 V, LED threshold on the signal coming from the top of Det#1 (multihit TDC8) was set to 255 mV (30 mV all others) and fed into CAEN N93B Dual Timer. The pulse width was increased by the timer to 60 ns. Let's see where in the timing spectrum of the second hits we can observe the spikes due to the ringing.
First hit (Det1 top): Second hit (Det1 top): Second hit (Det1 top) - First hit (Det1 top): |