Difference between revisions of "PosRun 10 14 2013"
(48 intermediate revisions by the same user not shown) | |||
Line 7: | Line 7: | ||
Detector 1 is Beam Right and in ADC channel 3, Ortec Dual Spec amp set to course=20, fine=10, Canberra CFD set to 70 mV lower threshold | Detector 1 is Beam Right and in ADC channel 3, Ortec Dual Spec amp set to course=20, fine=10, Canberra CFD set to 70 mV lower threshold | ||
+ | ==In-flight== | ||
{| border="1" |cellpadding="20" cellspacing="0 | {| border="1" |cellpadding="20" cellspacing="0 | ||
Line 12: | Line 13: | ||
| Run # || Run Duration (s) || Configuration | | Run # || Run Duration (s) || Configuration | ||
|- | |- | ||
− | | | + | | 6488 || min s || 2->60Hz rep rate, Looking at NaI in shielded cell at -600 Volts, Magnet on Accelerator side deflects electrons, Tungsten target is out, triggering only on the beam gate and looking at ADC channel 3 |
+ | |- | ||
+ | | 6489 || min s || same as above except: 120Hz rep rate, -500 Volts, | ||
+ | |- | ||
+ | | 6490 || min s || same as above except: -550 Volts, | ||
+ | |- | ||
+ | | 6491 || min s || same as above except: -575 Volts, | ||
+ | |- | ||
+ | | 6492 || min s || same as above except: inserted Aluminum brick to be sure electrons are staying on accelerator side, drops NaI output by factor of 2 | ||
+ | |- | ||
+ | | 6493 || min s || same as above except: -600 Volts | ||
+ | |- | ||
+ | | 6494 || min s || same as above except: -700 Volts | ||
+ | |- | ||
+ | | 6498 || min s || same as above except: -500 Volts, magnet out | ||
+ | |- | ||
+ | | 6499 || min s || same as above except: -700 Volts, very similar ADC spectrum for magnet in | ||
+ | |- | ||
+ | | 6500 || min s (54420 , 120 Hz pulses =453.5 sec)|| same as above except: removed Al brick beam stop, increased current from zero until signal was observed in experimental cell NaI (probably mostly dark current from RF cavity) | ||
+ | |- | ||
+ | | 6501 || min s (54297 pulses) || same as above except: inserted Al brick beam stop, [[File:r6500-01.png | 200 px]] | ||
+ | |- | ||
+ | | 6502 || min s (54297 pulses) || same as above except: REmoved Al brick beam stop | ||
+ | |} | ||
+ | |||
+ | ===40 MeV on Tungsten=== | ||
+ | |||
+ | 40 MeV electrons are hitting the positron production Tungsten target | ||
+ | |||
+ | |||
+ | |||
+ | |||
+ | {| border="1" |cellpadding="20" cellspacing="0 | ||
+ | |- | ||
+ | | Run # || Run Duration (s) || Configuration | ||
+ | |- | ||
+ | | 6503 || 3 min 46 s ( pulses) || Dipoles in positron mode. Al Brick in along 8 inch length | ||
+ | |- | ||
+ | | 6504 || min s ( pulses) || same as previous run, Al brick OUT | ||
+ | |- | ||
+ | | 6505 || min s ( pulses) || same as previous run, Dipoles in Electron mode, [[File:r6500-04-05_ch3.png| 200 px]] | ||
+ | |- | ||
+ | | 6506 || min s ( pulses) || same as previous run, Selection Magnet bending electrons up positioned after 50 micron Ti vacuum window, [[File:r6505-06_ch3.png| 200 px]] | ||
+ | |- | ||
+ | | 6507 || min s ( pulses) || same as previous run, dipoles set to positron mode and selection magnet rotated 180 degrees so positrons bend up, | ||
+ | |- | ||
+ | | 6508 || min s ( pulses) || same as previous run:increase voltage from -700 to -850 | ||
+ | |- | ||
+ | | 6509 || min s ( pulses) || same as previous run, installed 2" thick graphite hardener | ||
+ | |- | ||
+ | | 6510 || min s ( pulses) || same as previous run, removed graphite hardener and selection magnet, returned to dipoles in electron mode, | ||
+ | |} | ||
+ | |||
+ | ===Shielded Cell direct beam with magnet=== | ||
+ | |||
+ | reconfigure with sweep magnet in shielded cell to deflect charged particles towards NaI detector #3, | ||
+ | |||
+ | {| border="1" |cellpadding="20" cellspacing="0 | ||
+ | |- | ||
+ | | 6511 || min s ( pulses) || dipoles in electron mode transporting 8 MeV, NaI at -700 Volts | ||
+ | |- | ||
+ | | 6512 || min s ( pulses) || dipoles in electron mode transporting 6 MeV, Magnet is OUT | ||
+ | |- | ||
+ | | 6513 || min s ( pulses) || dipoles in electron mode transporting 6 MeV, Magnet is IN and deflecting positrons towards NaI | ||
+ | |- | ||
+ | | 6514 || 12 min 3 s ( pulses) || same as last run except: dipoles in positron mode, doubled current from last run | ||
+ | |- | ||
+ | | 6515 || min s ( pulses) || same as last run except: positron dipoles set to 8 MeV, increased current and rep rate (150 Hz) | ||
+ | |- | ||
+ | | 6516 || min s ( pulses) || same as last run except: positron dipoles set to 6 MeV, | ||
+ | |- | ||
+ | | 6517 || 2 min 18 s ( pulses) || same as last run except: Magnet is out | ||
+ | |- | ||
+ | | 6518 || 7 min 10s ( pulses) || same as last run except: positron dipoles set to 8 MeV, sweep Magnet is out | ||
+ | |- | ||
+ | | 6519 || min s ( pulses) || same as last run except: sweep Magnet is In positron bending mode | ||
+ | |- | ||
+ | | 6520 || min s ( pulses) || same as last run except: sweep Magnet is In electron bending mode | ||
+ | |- | ||
+ | | 6521 || min s ( pulses) || same as last run except: dipoles set to electron mode | ||
+ | |} | ||
+ | |||
+ | |||
+ | 38 MeV electrons with 859 pVs/50 Ohms per 100 nsec wide pulse | ||
+ | |||
+ | a scope measurement of the 38 MeV electron pulse impinging the tungsten target suggested that one pulse had | ||
+ | |||
+ | <math>\frac{0.859 \times 10^{-9} V s}{50 Ohm} = 1.7 \times 10^{-11} Coul \frac{1 e^-}{1.6 \times 10^{-19} Coul} = 10^{8}</math> electrons | ||
+ | |||
+ | or given the 100 nsec pulse width the peak current is | ||
+ | |||
+ | <math>\frac{1.7 \times 10^{-11} Coul}{100 \times 10^{-9} sec}</math> = 0.017 Amps | ||
+ | |||
+ | I believe we need to run at 0.01 Amps to keep the ion pumps working. | ||
+ | |||
+ | |||
+ | [[File:PosRes_10-15-2013.png | 200 px]][[File:PosResA_10-15-2013.png | 200 px]] | ||
+ | |||
+ | ==FC -vs- 90A beam stop correlation== | ||
+ | |||
+ | |||
+ | FC = Aluminum brick 2" from exit window near NaI detectors | ||
+ | |||
+ | |||
+ | 50 Ohm scope termination | ||
+ | {| border="1" |cellpadding="20" cellspacing="0 | ||
+ | |- | ||
+ | | 90 A BS (nVs) || FC (nVs) | ||
+ | |- | ||
+ | | 47 || 25 | ||
+ | |- | ||
+ | | 24 || 11.9 | ||
+ | |- | ||
+ | | 13 || 6 | ||
+ | |- | ||
+ | | 6.1 || 2.3 | ||
+ | |- | ||
+ | | 4.8 || 1.8 | ||
|} | |} | ||
+ | |||
[[Positrons_Using_40Mev_Linac]] | [[Positrons_Using_40Mev_Linac]] |
Latest revision as of 20:49, 16 October 2013
Run list
Scaler channel #2 counts the number of beam pulses
Detector 4 is Beam Left and in ADC channel 7, Ortec Dual Spec amp set to course=10, fine=39, Canberra CFD set to 85 mV lower threshold
Detector 1 is Beam Right and in ADC channel 3, Ortec Dual Spec amp set to course=20, fine=10, Canberra CFD set to 70 mV lower threshold
In-flight
40 MeV on Tungsten
40 MeV electrons are hitting the positron production Tungsten target
Shielded Cell direct beam with magnet
reconfigure with sweep magnet in shielded cell to deflect charged particles towards NaI detector #3,
6511 | min s ( pulses) | dipoles in electron mode transporting 8 MeV, NaI at -700 Volts |
6512 | min s ( pulses) | dipoles in electron mode transporting 6 MeV, Magnet is OUT |
6513 | min s ( pulses) | dipoles in electron mode transporting 6 MeV, Magnet is IN and deflecting positrons towards NaI |
6514 | 12 min 3 s ( pulses) | same as last run except: dipoles in positron mode, doubled current from last run |
6515 | min s ( pulses) | same as last run except: positron dipoles set to 8 MeV, increased current and rep rate (150 Hz) |
6516 | min s ( pulses) | same as last run except: positron dipoles set to 6 MeV, |
6517 | 2 min 18 s ( pulses) | same as last run except: Magnet is out |
6518 | 7 min 10s ( pulses) | same as last run except: positron dipoles set to 8 MeV, sweep Magnet is out |
6519 | min s ( pulses) | same as last run except: sweep Magnet is In positron bending mode |
6520 | min s ( pulses) | same as last run except: sweep Magnet is In electron bending mode |
6521 | min s ( pulses) | same as last run except: dipoles set to electron mode |
38 MeV electrons with 859 pVs/50 Ohms per 100 nsec wide pulse
a scope measurement of the 38 MeV electron pulse impinging the tungsten target suggested that one pulse had
electrons
or given the 100 nsec pulse width the peak current is
= 0.017 Amps
I believe we need to run at 0.01 Amps to keep the ion pumps working.
FC -vs- 90A beam stop correlation
FC = Aluminum brick 2" from exit window near NaI detectors
50 Ohm scope termination
90 A BS (nVs) | FC (nVs) |
47 | 25 |
24 | 11.9 |
13 | 6 |
6.1 | 2.3 |
4.8 | 1.8 |