| 07/14/13 ||20:49 || 536 || 1500 || -750 || 0.74 || 0.81 || | | 07/14/13 ||20:49 || 536 || 1500 || -750 || 0.74 || 0.81 || | ||
|- | |- | ||
− | | 07/15/13 ||07:23 || 536 || 1500 || -750 || 0.70 || 0.78 || | + | | 07/15/13 ||07:23 || 536 || 1500 || -750 || 0.70 || 0.78 || There is a guard current when I turned on the power supply for burning, I did not increase the voltage to check if this is ok. Also I checked sector one it is working, i saw signal in all the channels that are connected to the "Dawn" VME. I debugged sector5 's board that has a guard current, none of the boots is responsible for the for the guard curent, as all of them are out of the board, the guard current is still there, the only way to get rid of it is to separate the guard's jumpers from any side of the board. I checked the resistors of the board, they all good. |
|- | |- | ||
− | | | + | | 11/24/13 ||07:23 || 536 || 1500 || -750 || 0.70 || 0.80|| guard current is zero |
− | |||
|} | |} | ||
Latest revision as of 16:53, 24 November 2013
Date | Time (24h) | Guard (V) | Sense (V) | Field (V) | Sense ( | A)Field ( | A)comments |
05/28/13 | 14:08 | 100 | 100 | -100 | 0.02 | 0.00 | Confirmed with initial setting of 30/40/50V that there are no crossed connections between wire types and that voltage supply is connected correctly. |
05/28/13 | 16:34 | 536 | 1500 | -750 | 45.5 | 45.5 | Someone upped the voltage to full the full operating level without recording it. This is the current reading right now. |
05/28/13 | 17:56 | 536 | 1500 | -750 | 41.5 | 41.6 | |
05/29/13 | 07:56 | 536 | 1500 | -750 | 22.5 | 22.6 | |
05/29/13 | 15:02 | 536 | 1500 | -750 | 17.9 | 18.0 | |
05/29/13 | 18:07 | 536 | 1500 | -750 | 16.7 | 16.8 | |
05/30/13 | 07:55 | 536 | 1500 | -750 | 10.6 | 10.7 | |
05/30/13 | 12:51 | 536 | 1500 | -750 | 9.5 | 9.6 | |
05/30/13 | 17:23 | 536 | 1500 | -750 | 8.7 | 8.8 | |
06/03/13 | 08:23 | 536 | 1500 | -750 | 1.7 | 1.8 | |
06/04/13 | 07:42 | 536 | 1500 | -750 | 1.4 | 1.5 | |
06/05/13 | 12:01 | 536 | 1500 | -750 | 1.8 | 1.9 | had to turn off HV to organize cables |
06/05/13 | 17:56 | 536 | 1500 | -750 | 1.8 | 1.9 | |
06/06/13 | 07:56 | 536 | 1500 | -750 | 1.8 | 1.9 | seem to be stuck at this current draw? |
06/10/13 | 17:29 | 536 | 1500 | -750 | 1.5 | 1.6 | I think we ran out of gas on the weekend, the windows were deflated a bit but now they deflected outward |
06/11/13 | 07:50 | 536 | 1500 | -750 | 3.2 | 3.3 | current doubles? Humidity dropped from 40 to 25%. |
06/12/13 | 13:39 | 536 | 1500 | -750 | 3.6 | 3.7 | Humidity rose from 25 to 40%. |
06/13/13 | 11:00 | 536 | 1500 | -750 | 4.5 | 4.6 | Humidity is 40 +/- 5 %. |
06/14/13 | 08:56 | 536 | 1500 | -750 | 3.5 | 3.6 | |
06/17/13 | 07:58 | 536 | 1500 | -750 | 3.6 | 3.7 | |
06/24/13 | 08:37 | 536 | 1500 | -750 | 4.5 | 4.6 | turned off HV since 6/17/13 but continued flowing gas |
06/24/13 | 10:05 | 536 | 1500 | -750 | 5.14 | 5.24 | |
06/24/13 | 12:46 | 536 | 1500 | -750 | 5.10 | 4.99 | |
06/25/13 | 10:36 | 536 | 1500 | -750 | 4.7 | 4.7 | |
07/12/13 | 11:45 | 536 | 1500 | -750 | 1.0 | 1.1 | HV was turned back on to check current levels, I will leave it on over the weekend |
07/13/13 | 9:13 | 536 | 1500 | -750 | 0.76 | 0.84 | |
07/14/13 | 9:04 | 536 | 1500 | -750 | 0.70 | 0.76 | |
07/14/13 | 20:49 | 536 | 1500 | -750 | 0.74 | 0.81 | |
07/15/13 | 07:23 | 536 | 1500 | -750 | 0.70 | 0.78 | There is a guard current when I turned on the power supply for burning, I did not increase the voltage to check if this is ok. Also I checked sector one it is working, i saw signal in all the channels that are connected to the "Dawn" VME. I debugged sector5 's board that has a guard current, none of the boots is responsible for the for the guard curent, as all of them are out of the board, the guard current is still there, the only way to get rid of it is to separate the guard's jumpers from any side of the board. I checked the resistors of the board, they all good. |
11/24/13 | 07:23 | 536 | 1500 | -750 | 0.70 | 0.80 | guard current is zero |
Why is the current increasing?
is the current going to ground ?
Try raising the sense to 1750 and lowering the field to 500 (same voltage difference), and see if the sense current still equals the field current - want to know if the current is inside the chamber from wire to wire or if it's a current to ground, for example ...
If current is from field to sense:
if it is current from field to sense the current should increase exponentially with voltage, doubling with 100V increase; i.e. 2250 -> 2350. You could check to see if this is the case. If so, you could then try to locate where the problem is: which wire(s) are responsible?
Increasing Sense
Date | Time (24h) | Guard (V) | Sense (V) | Field (V) | Sense ( | A)Field ( | A)comments |
06/17/13 | 12:20 | 536 | 1510 | -750 | 8 | 8.1 | |
06/17/13 | 12:16 | 536 | 1520 | -750 | 10.7 | 10.8 | |
06/17/13 | 12:34 | 536 | 1530 | -750 | 22.3 | 22.4 | |
06/17/13 | 12:43 | 536 | 1540 | -750 | 37.3 | 37.4 | |
06/17/13 | 15:53 | 536 | 1540 | -750 | 45.2 | 45.3 | |
06/17/13 | 15:55 | 536 | 1550 | -750 | 54.9 | 55.1 |
Lower the field to -500 V
Date | Time (24h) | Guard (V) | Sense (V) | Field (V) | Sense ( | A)Field ( | A)comments |
06/17/13 | 15:55 | 536 | 1550 | -750 | 54.9 | 55.1 | |
06/17/13 | 15:58 | 536 | 1550 | -740 | 45.9 | 46.1 | |
06/17/13 | 15:59 | 536 | 1550 | -730 | 37.6 | 37.9 | |
06/17/13 | 14:00 | 536 | 1550 | -710 | 24.2 | 24.3 | |
06/17/13 | 14:02 | 536 | 1550 | -690 | 13.9 | 14.0 | |
06/17/13 | 14:03 | 536 | 1550 | -650 | 3.8 | 3.9 | |
06/17/13 | 14:04 | 536 | 1550 | -600 | 1.33 | 1.41 | |
06/17/13 | 14:05 | 536 | 1550 | -550 | 0.88 | 0.97 | |
06/17/13 | 14:06 | 536 | 1550 | -500 | 0.66 | 0.74 |
Increase sense when Field is at -500 V
Date | Time (24h) | Guard (V) | Sense (V) | Field (V) | Sense ( | A)Field ( | A)comments |
06/17/13 | 14:06 | 536 | 1550 | -500 | 0.66 | 0.74 | |
06/17/13 | 14:10 | 536 | 1650 | -500 | 1.5 | 1.6 | |
06/17/13 | 14:13 | 536 | 1750 | -500 | 16.2 | 16.3 |
SL2 current draw search
Date | Time (24h) | Guard (V) | Sense (V) | Field (V) | Sense ( | A)Field ( | A)comments |
06/18/13 | 08:57 | 536 | 1550 | -550 | 0.59 | 0.69 | no voltage to HVB1 |
06/18/13 | 09:00 | 536 | 1550 | -500 | 0.46 | 0.55 | no voltage to HVB1 |
06/18/13 | 09:12 | 536 | 1650 | -500 | 1.1 | 1.2 | no voltage to HVB1 |
06/18/13 | 09:17 | 536 | 1750 | -500 | 9.5 | 9.6 | no voltage to HVB1, less current measured |
06/18/13 | 10:31 | 536 | 1550 | -500 | 0.32 | 0.42 | no voltage to HVB1 and HVB2 |
06/18/13 | 10:40 | 536 | 1650 | -500 | 1.0 | 1.1 | no voltage to HVB1 and HVB2 |
06/18/13 | 10:45 | 536 | 1750 | -500 | 8.8 | 8.9 | no voltage to HVB1 and HVB2 |
06/18/13 | 11:07 | 536 | 1550 | -500 | 0.32 | 0.42 | no voltage to HVB1 and HVB3 |
06/18/13 | 11:11 | 536 | 1650 | -500 | 1.0 | 1.1 | no voltage to HVB1 and HVB3 |
06/18/13 | 11:16 | 536 | 1750 | -500 | 6.8 | 6.9 | no voltage to HVB1 and HVB3 |
06/18/13 | 11:31 | 536 | 1550 | -500 | 0.34 | 0.43 | no voltage to HVB1,HVB3,and HVB4 |
06/18/13 | 11:34 | 536 | 1650 | -500 | 0.82 | 0.92 | no voltage to HVB1,HVB3,and HVB4 |
06/18/13 | 11:38 | 536 | 1750 | -500 | 5.0 | 5.1 | no voltage to HVB1,HVB3,and HVB4 negligible effect |
06/18/13 | 12:12 | 536 | 1550 | -500 | 0.34 | 0.42 | no voltage to HVB1,HVB3,and HVB5 |
06/18/13 | 12:15 | 536 | 1650 | -500 | 0.89 | 0.97 | no voltage to HVB1,HVB3,and HVB5 |
06/18/13 | 12:19 | 536 | 1750 | -500 | 4.9 | 5.0 | no voltage to HVB1,HVB3,and HVB5 negligible effect |
06/18/13 | 14:08 | 536 | 1550 | -500 | 0.0 | 0.055 | no voltage to HVB1,HVB3,and HVB6 |
06/18/13 | 14:24 | 536 | 1650 | -500 | 0.2 | 0.27 | no voltage to HVB1,HVB3,and HVB6 |
06/18/13 | 14:30 | 536 | 1750 | -500 | 4.5 | 4.6 | no voltage to HVB1,HVB3,and HVB6 |
06/18/13 | 14.50 | 536 | 1550 | -500 | 0.335 | 0.420 | no voltage to HVB1,HVB3,and HVB7 |
06/18/13 | 14:53 | 536 | 1650 | -500 | 0.965 | 1.035 | no voltage to HVB1,HVB3,and HVB7 |
06/18/13 | 15:00 | 536 | 1750 | -500 | 5.2 | 5.3 | no voltage to HVB1,HVB3,and HVB7 |
06/26/13 | 17:37 | 536 | 1500 | -750 | 4.6 | 4.3 | Switched to SL2 Kraken HV after burning in SL1 HVB1 with Octopus. SL1 HVB1 burned down to about 1 microamp over ~24 hrs after starting at a high of 4 microamps. When setting up HVB1 for burn-in, noticed a grounding strap attached to the board that caused high guard current even at 30V. Removed the strap and this problem was resolved, burn-in proceeded normally. This is the reading on SL2 + Kraken immediately after reaching 1500V on sense. |
06/26/13 | 17:41 | 536 | 1500 | -750 | 5.7 | 5.3 | Guard current is at 0.295 microamps. Normally is below a level detectable by this HV supply (0.000). |
06/26/13 | 17:55 | 536 | 1500 | -750 | 5.3 | 5.0 | Noticed several of the mounting screws that go into the chamber's metal frame for both STBs and HVBs on SL2 were not torqued to 13 in. lbs / some were obviously loose. Turned off HV and torqued down all mounting screws. Some do not appear flush against STB/HVB surface, but most are flush. Turned back on HV. Guard current is still 0.280 microamps. Torquing screws didn't fix it. |
06/27/13 | 17:25 | 536 | 1500 | -750 | 2.8 | 2.7 | Guard current is down to 0.045 |
Sense Field difference constant
Change sense and field wire voltage but keep the difference 2350 Volts. Record currents over a time interval of minutes. Don't wait long.