Difference between revisions of "2n Neutron signal attenuation"

From New IAC Wiki
Jump to navigation Jump to search
 
(28 intermediate revisions by the same user not shown)
Line 52: Line 52:
 
[[File:Det_I_pic.jpg | 300 px]]
 
[[File:Det_I_pic.jpg | 300 px]]
  
=Na-22 source timing rest=
+
=Na-22 and Co-60 source timing test=
 +
 
 +
Here it was used Det K. DAQ was triggered on the coincidence signal from two PMTs and each PMT signal was delayed and connected to corresponding TDC channel. DAQ was configured to work in the common start mode. Na-22 source was moved over the detector surface.
 +
 
 +
==Raw spectra==
 +
 
 +
The difference of TOF TDC16-TDC17 (i.e. K1(left)-K2(right)) is plotted below for different source position.
  
 
[[File:BKG_timing.jpg | 300 px]]
 
[[File:BKG_timing.jpg | 300 px]]
 +
 +
As can be seen the BKG spectrum is not centered around zero time. That means that delays in the channels are different.
  
 
[[File:Left_close_timing.jpg | 300 px]]
 
[[File:Left_close_timing.jpg | 300 px]]
Line 62: Line 70:
 
[[File:Right_close_timing.jpg | 300 px]]
 
[[File:Right_close_timing.jpg | 300 px]]
  
== Normalized and BKG subtracted data ==
+
==Det K data==
For all the timing spectra it was calculated K1(left)-K2(right).
+
 
 +
[[Normalized and BKG subtracted data]]
 +
 
 +
==Det I data==
 +
 
 +
[[Normalized and BKG subtracted data for Det I]]
 +
 
 +
==Det H data==
 +
 
 +
[[Normalized and BKG subtracted data for Det H]]
 +
 
 +
==Det G data==
 +
 
 +
[[Normalized and BKG subtracted data for Det G]]
 +
 
 +
==Det F data==
  
The source is close to the left PMT (13.86 cm away):
+
[[Normalized and BKG subtracted data for Det F]]
  
[[File:Leftest_minus_bkg.png | 400 px]]
+
==Det M data==
  
The source is a bit further from the left PMT (17.64 cm away):
+
[[Normalized and BKG subtracted data for Det M]]
  
[[File:Left_minus_bkg.png | 400 px]]
+
==Det E data==
  
The source is at the center of the detector:
+
[[Normalized and BKG subtracted data for Det E]]
  
[[File:Center_minus_bkg.png | 400 px]]
+
==Slope data==
  
The source is close to the right PMT (13.86 cm away):
+
[[File:Slope_data.png]]
  
[[File:Rightest_minus_bkg.png | 400 px]]
 
  
==Calibration curve==
+
All the calibrations were obtained for the case when detectors had the following HV applied:
Finally, using the mean values of the fitted Na-22 peaks the calibration curve can be plotted:
 
  
[[File:Calibration_line.jpg | 400 px]]
+
{| class="wikitable"
 +
|-
 +
! Detector !! CFD threshold [mV] !! HV [V]
 +
|-
 +
| F(1,2) || 0 || -1702
 +
|-
 +
| I(1) || 0 || -1654
 +
|-
 +
| I(2) || 0 || -1700
 +
|-
 +
| G(1) || 0 || -1700
 +
|-
 +
| G(2) || 0 || -1650
 +
|-
 +
| H(1) || 0 || -1730
 +
|-
 +
| H(2) || 0 || -1650
 +
|-
 +
| E(1) || 0 || -1650
 +
|-
 +
| E(2) || 0 || -1750
 +
|-
 +
| M(1) || 0 || -1602
 +
|-
 +
| M(2) || 0 || -1750
 +
|-
 +
| K(1) || 0 || -1755
 +
|-
 +
| K(2) || 0 || -1600 (8(60))
 +
|}

Latest revision as of 16:59, 20 November 2012

Cf-252 source attenuation test

Cf-252 source was used to define the attenuation of a signal produced by neutrons in plastic scintillator. Cf-252 source was placed in different positions along the detector surface shielded with 2" of lead and counting rate of the two PMTs was observed via dual counter ORTEC 778. Cf-252 source was surrounded by borated poly to provide collimation of about 5 cm in diameter.

File:Data attenuation.pdf

Would you add to the plot the sum of the rate from each end of the PMT.

In order to get it on the same scale you could divide by 2 or use a scale on the right side of the graph

Data with the counting rate from each tube added together.


Detector E(1,2) neutron signal attenuation data sheet:

File:Detector E data.pdf

Det E pic.jpg

Detector M(1,2) neutron signal attenuation data sheet:

File:Detector M data.pdf

Det M pic.jpg

Detector F(1,2) neutron signal attenuation data sheet:

File:Detector F data.pdf

Det F pic.jpg

Detector G(1,2) neutron signal attenuation data sheet:

File:Detector G data.pdf

Det G pic.jpg

Detector H(1,2) neutron signal attenuation data sheet:

File:Detector H data.pdf

Det H pic.jpg

Detector K(1,2) neutron signal attenuation data sheet:

File:Detector K data.pdf

Det K pic.jpg

Detector I(1,2) neutron signal attenuation data sheet:

File:Detector I data.pdf

Det I pic.jpg

Na-22 and Co-60 source timing test

Here it was used Det K. DAQ was triggered on the coincidence signal from two PMTs and each PMT signal was delayed and connected to corresponding TDC channel. DAQ was configured to work in the common start mode. Na-22 source was moved over the detector surface.

Raw spectra

The difference of TOF TDC16-TDC17 (i.e. K1(left)-K2(right)) is plotted below for different source position.

BKG timing.jpg

As can be seen the BKG spectrum is not centered around zero time. That means that delays in the channels are different.

Left close timing.jpg

Middle timing.jpg

Right close timing.jpg

Det K data

Normalized and BKG subtracted data

Det I data

Normalized and BKG subtracted data for Det I

Det H data

Normalized and BKG subtracted data for Det H

Det G data

Normalized and BKG subtracted data for Det G

Det F data

Normalized and BKG subtracted data for Det F

Det M data

Normalized and BKG subtracted data for Det M

Det E data

Normalized and BKG subtracted data for Det E

Slope data

Slope data.png


All the calibrations were obtained for the case when detectors had the following HV applied:

Detector CFD threshold [mV] HV [V]
F(1,2) 0 -1702
I(1) 0 -1654
I(2) 0 -1700
G(1) 0 -1700
G(2) 0 -1650
H(1) 0 -1730
H(2) 0 -1650
E(1) 0 -1650
E(2) 0 -1750
M(1) 0 -1602
M(2) 0 -1750
K(1) 0 -1755
K(2) 0 -1600 (8(60))