Difference between revisions of "Beam Test Jul 26 2012"
(87 intermediate revisions by 2 users not shown) | |||
Line 44: | Line 44: | ||
= DAQ runs = | = DAQ runs = | ||
+ | |||
+ | 2 MeV positron energy. | ||
TDC CH5 was a delay of (32+16+8+4+2+1+0.5)+ (16+8+4+2+1+0.5) = 95 nano second. | TDC CH5 was a delay of (32+16+8+4+2+1+0.5)+ (16+8+4+2+1+0.5) = 95 nano second. | ||
+ | |||
+ | {| border="1" |cellpadding="20" cellspacing="0 | ||
+ | |- | ||
+ | | run# || T1 || T2 || Sweep Magnet|| logic || HV-D2L || HV-D1R || Reprate || start || stop || Int Rate || run time || Total events | ||
+ | |- | ||
+ | | 3679 || in || in || out|| (DL & GunTrig) or (DR & GunTrig) || 1060V || 1080V || 100 || 08:48:00 || 09:28:01 || 42.3 events/s|| 2401 s || 101466 | ||
+ | |- | ||
+ | | 3680 || in || OUT || out|| (DL & GunTrig) or (DR & GunTrig) || 1060V || 1080V || 100 || 09:29:37 || 09:46:19 || 42.6 events/s|| 1002 s || 42596 | ||
+ | |- | ||
+ | |} | ||
+ | |||
+ | {| border="1" |cellpadding="20" cellspacing="0 | ||
+ | |- | ||
+ | | run# || T1 || T2 || Sweep Magnet|| logic || HV-D2L || HV-D1R || Reprate || start || stop || Int Rate ||comments | ||
+ | |- | ||
+ | | 3679 || in || in || out|| (DL & GunTrig) or (DR & GunTrig) || 1060V || 1080V || 100 || 08:48:00 || 09:28:01 || 42.3 events/s|| ADC12 has Sintilator1 (Sint @ exit of 90 deg). | ||
+ | |- | ||
+ | | 3680 || in || OUT || out|| (DL & GunTrig) or (DR & GunTrig) || 1060V || 1080V || 100 || 09:29:37 || 09:46:19 || 42.6 events/s|| ADC12 has Sintilator1 (Sint @ exit of 90 deg). | ||
+ | |- | ||
+ | | 3682 || in || OUT || out|| (DL & GunTrig) AND (DR & GunTrig) || 1150V || 1150V || 100 || 09:55:20 || 10:33:56 (2316 s) || 4.8884 events/s|| TDC cable was upside down, now it is flipped. Sintilator left was put to CH9. | ||
+ | |- | ||
+ | | 3683 || in || IN || out|| (DL & GunTrig) AND (DR & GunTrig) || 1150V || 1150V || 100 || 10:36:03 || 10:53:48 (1065 s) || 5.9397 events/s|| TDC5->beam left | ||
+ | |- | ||
+ | | 3684 || in || IN || out|| (DL & GunTrig) AND (DR & GunTrig) || 1150V || 1150V || 100 || 10:54:19 || 11:25:01 (1842 s)|| 8.7974 events/s|| TDC3->beam right, TDC5->beam left | ||
+ | |- | ||
+ | | 3685 || in || out || out|| (DL & GunTrig) AND (DR & GunTrig) || 1150V || 1150V || 100 || 11:26:43 || 11:31:01 || 18 events/s|| junk, rate are too high. | ||
+ | |- | ||
+ | |} | ||
+ | |||
+ | We changed the discriminator from Octal discriminator (LeCroy modle 623B) to 8 CH Constant Fraction Discriminator (CFD) (CAEN model N842). The threshold on the detector beam left is 29, and on the detector right is 31. | ||
+ | |||
+ | Following is the image of coincidence of left and right detectors using this CFD. | ||
+ | |||
+ | [[File:hrrl_pos_26jul2012_FADC_coincidence.png | 300 px]] | ||
+ | |||
+ | |||
+ | {| border="1" |cellpadding="20" cellspacing="0 | ||
+ | |- | ||
+ | | run# || T1 || T2 || Sweep Magnet|| logic || HV-D2L || HV-D1R || Reprate || start || stop || Int Rate || Connections ||comments | ||
+ | |- | ||
+ | | 3686 || in || in || out|| (DL & GunTrig) AND (DR & GunTrig) || 1150V || 1150V || 100 || 11:54:44 || 12:15:44 (1260 s)|| 4.8262 events/s|| TDC3->DL, TDC->DR || | ||
+ | |- | ||
+ | | 3687 || in || out || out|| (DL & GunTrig) AND (DR & GunTrig) || 1150V || 1150V || 100 || 12:19:15 || 12:43:14 (1411 s)|| 3.6 events/s|| TDC3->DL, TDC5->DR || | ||
+ | |- | ||
+ | | 3688 || in || in || out|| (DL & GunTrig) AND (DR & GunTrig) || 1150V || 1150V || 100 || 12:44:09 || 13:25:09 (2460 s)|| 4.9743 events/s|| TDC3->DL, TDC12->DR || switched DR from TDC5 to TDC12 | ||
+ | |- | ||
+ | | 3689 || in || out || out|| (DL & GunTrig) AND (DR & GunTrig) || 1150V || 1150V || 100 || 13:27:23 || 14:12:29 (2706 s)|| 3.3156 events/s|| TDC3->DL, TDC12->DR || CFD out-put of DL (to TDC9) and DR (to TDC10) connected to TDC. (Connected to the top row of LeCroy Model 4616.) | ||
+ | |- | ||
+ | | 3690 || in || in || out|| (DL & GunTrig) AND (DR & GunTrig) || 1150V || 1150V || 100 || 14:13:16|| 15:23:44 (4228 s)|| at the beginning was 4.9 then later changed to 13.5111 events/s || TDC3->DL, TDC12->DR || CFD out-put of DL (to TDC9) and DR (to TDC10) connected to TDC. (Connected to the seond row of LeCroy Model 4616.) | ||
+ | |- | ||
+ | | 3691 || in || out || out|| (DL & GunTrig) AND (DR & GunTrig) || 1150V || 1150V || 100 || 16:13:16|| 16:36:19 (1383 s)|| 14.2079 events/s|| TDC3->DL, TDC12->DR || High voltage of scintilator at 90 deg exit set to 1300 V. Scintlator at 90 deg moved to CH13. | ||
+ | |- | ||
+ | | 3692 || in || in || in|| (DL & GunTrig) AND (DR & GunTrig) || 1150V || 1150V || 100 || 16:40:18|| 17:00:56 (1238 s)|| 23.9270 events/s|| TDC3->DL, TDC12->DR || | ||
+ | |- | ||
+ | | 3694 || in || in || in|| (DL & GunTrig) AND (DR & GunTrig) || 1150V || 1150V || 100 || || || 10 events/s|| TDC3->DL, TDC12->DR || | ||
+ | |- | ||
+ | | 3696 || in || in || in|| (DL & GunTrig) AND (DR & GunTrig) || 1150V || 1150V || 100 || 17:30:08||17:31:39 || 10 events/s|| TDC3->DL, TDC12->DR || | ||
+ | |- | ||
+ | |} | ||
+ | |||
+ | == Data Analysis == | ||
+ | |||
+ | r3686 and r3687: [[HRRL_Pos_Test_26Jul2012_Data_Ana_com_r3686_r3687]] | ||
+ | |||
+ | r3688 and r3689: [[HRRL_Pos_Test_26Jul2012_Data_Ana_com_r3688_r3689]] | ||
+ | |||
+ | r3690 and r3691: [[HRRL_Pos_Test_26Jul2012_Data_Ana_com_r3688_r3689]] | ||
+ | |||
+ | Subtracted Sepctrums: [[HRRL_Pos_Test_26Jul2012_Data_Ana_Sub_Spcs]] | ||
+ | |||
+ | = Detector Calibrations = | ||
+ | |||
{| border="1" |cellpadding="20" cellspacing="0 | {| border="1" |cellpadding="20" cellspacing="0 | ||
Line 86: | Line 160: | ||
| 3690 || in || in || out|| (DL & GunTrig) AND (DR & GunTrig) || 1150V || 1150V || 100 || 14:13:16|| 15:23:44 (4228 s)|| at the beginning was 4.9 then later changed to 13.5111 events/s || TDC3->DL, TDC12->DR || CFD out-put of DL (to TDC9) and DR (to TDC10) connected to TDC. (Connected to the seond row of LeCroy Model 4616.) | | 3690 || in || in || out|| (DL & GunTrig) AND (DR & GunTrig) || 1150V || 1150V || 100 || 14:13:16|| 15:23:44 (4228 s)|| at the beginning was 4.9 then later changed to 13.5111 events/s || TDC3->DL, TDC12->DR || CFD out-put of DL (to TDC9) and DR (to TDC10) connected to TDC. (Connected to the seond row of LeCroy Model 4616.) | ||
|- | |- | ||
− | | 3691 || in || out || out|| (DL & GunTrig) AND (DR & GunTrig) || 1150V || 1150V || 100 || 14:13:16|| :: || | + | | 3691 || in || out || out|| (DL & GunTrig) AND (DR & GunTrig) || 1150V || 1150V || 100 || 14:13:16|| 16:36:19 || 14.2079 events/s|| TDC3->DL, TDC12->DR || High voltage of scintilator at 90 deg exit set to 1300 V. Scintlator at 90 deg moved to CH13. |
|- | |- | ||
− | | 3692 || in || in || in|| (DL & GunTrig) AND (DR & GunTrig) || 1150V || 1150V || 100 || ::|| :: || events/s|| TDC3->DL, TDC12->DR || | + | | 3692 || in || in || in|| (DL & GunTrig) AND (DR & GunTrig) || 1150V || 1150V || 100 || 16:40:18|| 17:00:56 || 23.9270 events/s|| TDC3->DL, TDC12->DR || |
+ | |- | ||
+ | | 3694 || in || in || in|| (DL & GunTrig) AND (DR & GunTrig) || 1150V || 1150V || 100 || || || 10 events/s|| TDC3->DL, TDC12->DR || | ||
+ | |- | ||
+ | | 3696 || in || in || in|| (DL & GunTrig) AND (DR & GunTrig) || 1150V || 1150V || 100 || 17:30:08||17:31:39 || 10 events/s|| TDC3->DL, TDC12->DR || | ||
+ | |- | ||
+ | |} | ||
+ | ==Beam Current Calibration of Scint Left== | ||
+ | |||
+ | Determine if the Beam Left Scintillator is linear with beam current. | ||
+ | |||
+ | {| border="1" |cellpadding="20" cellspacing="0 | ||
+ | |- | ||
+ | | run# || Source || HV-D2L || HV-D1R || start || stop || Int Rate || Connections ||comments | ||
+ | |- | ||
+ | | 3699 || Na22 || 1150V || 1150V || 18:20:47 || 18:21:33 || 103.6 events/s|| TDC3->DL, TDC->DR ||A lot of error message saying no data in TDC and ADC. | ||
+ | |- | ||
+ | | 3700 || Na22 || 1150V || 1150V || 18:26:15 || 20:45:47 || 103.6 events/s|| TDC3->DL, TDC->DR || | ||
+ | |- | ||
|} | |} | ||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | + | [[https://wiki.iac.isu.edu/index.php/HRRL_July_2012Run Go Back]] |
Latest revision as of 07:03, 4 October 2013
Tune
Energy Switch | Out (High En) |
Solenoid 1 | 13.9 A |
Solenoid 2 | 4.1 A |
Gun Hor | -0 A |
Gun Ver | +0.2 A |
Output Hor | -0.3 A |
Output Ver | -0 A |
Rep Rate | 17 |
Gun HV | 10 (Knob Setting) |
Gun Grid Voltage | 10 (Knob Setting) |
RF frequency | 2856.321 MHz |
Modulator HV Power Supply | 4.19 (Knob Setting) |
RF macro Pulse Length (FWHM) | 300 ns |
Q1 (A) | -3 |
Q2 (A) | +0.8 |
Q3 (A) | +2 |
Q4 (A) | 0 |
Q5 (A) | 0 |
Q6 (A) | 0 |
Current at FC0 | 96 mA |
DAQ runs
2 MeV positron energy.
TDC CH5 was a delay of (32+16+8+4+2+1+0.5)+ (16+8+4+2+1+0.5) = 95 nano second.
run# | T1 | T2 | Sweep Magnet | logic | HV-D2L | HV-D1R | Reprate | start | stop | Int Rate | run time | Total events |
3679 | in | in | out | (DL & GunTrig) or (DR & GunTrig) | 1060V | 1080V | 100 | 08:48:00 | 09:28:01 | 42.3 events/s | 2401 s | 101466 |
3680 | in | OUT | out | (DL & GunTrig) or (DR & GunTrig) | 1060V | 1080V | 100 | 09:29:37 | 09:46:19 | 42.6 events/s | 1002 s | 42596 |
run# | T1 | T2 | Sweep Magnet | logic | HV-D2L | HV-D1R | Reprate | start | stop | Int Rate | comments |
3679 | in | in | out | (DL & GunTrig) or (DR & GunTrig) | 1060V | 1080V | 100 | 08:48:00 | 09:28:01 | 42.3 events/s | ADC12 has Sintilator1 (Sint @ exit of 90 deg). |
3680 | in | OUT | out | (DL & GunTrig) or (DR & GunTrig) | 1060V | 1080V | 100 | 09:29:37 | 09:46:19 | 42.6 events/s | ADC12 has Sintilator1 (Sint @ exit of 90 deg). |
3682 | in | OUT | out | (DL & GunTrig) AND (DR & GunTrig) | 1150V | 1150V | 100 | 09:55:20 | 10:33:56 (2316 s) | 4.8884 events/s | TDC cable was upside down, now it is flipped. Sintilator left was put to CH9. |
3683 | in | IN | out | (DL & GunTrig) AND (DR & GunTrig) | 1150V | 1150V | 100 | 10:36:03 | 10:53:48 (1065 s) | 5.9397 events/s | TDC5->beam left |
3684 | in | IN | out | (DL & GunTrig) AND (DR & GunTrig) | 1150V | 1150V | 100 | 10:54:19 | 11:25:01 (1842 s) | 8.7974 events/s | TDC3->beam right, TDC5->beam left |
3685 | in | out | out | (DL & GunTrig) AND (DR & GunTrig) | 1150V | 1150V | 100 | 11:26:43 | 11:31:01 | 18 events/s | junk, rate are too high. |
We changed the discriminator from Octal discriminator (LeCroy modle 623B) to 8 CH Constant Fraction Discriminator (CFD) (CAEN model N842). The threshold on the detector beam left is 29, and on the detector right is 31.
Following is the image of coincidence of left and right detectors using this CFD.
run# | T1 | T2 | Sweep Magnet | logic | HV-D2L | HV-D1R | Reprate | start | stop | Int Rate | Connections | comments |
3686 | in | in | out | (DL & GunTrig) AND (DR & GunTrig) | 1150V | 1150V | 100 | 11:54:44 | 12:15:44 (1260 s) | 4.8262 events/s | TDC3->DL, TDC->DR | |
3687 | in | out | out | (DL & GunTrig) AND (DR & GunTrig) | 1150V | 1150V | 100 | 12:19:15 | 12:43:14 (1411 s) | 3.6 events/s | TDC3->DL, TDC5->DR | |
3688 | in | in | out | (DL & GunTrig) AND (DR & GunTrig) | 1150V | 1150V | 100 | 12:44:09 | 13:25:09 (2460 s) | 4.9743 events/s | TDC3->DL, TDC12->DR | switched DR from TDC5 to TDC12 |
3689 | in | out | out | (DL & GunTrig) AND (DR & GunTrig) | 1150V | 1150V | 100 | 13:27:23 | 14:12:29 (2706 s) | 3.3156 events/s | TDC3->DL, TDC12->DR | CFD out-put of DL (to TDC9) and DR (to TDC10) connected to TDC. (Connected to the top row of LeCroy Model 4616.) |
3690 | in | in | out | (DL & GunTrig) AND (DR & GunTrig) | 1150V | 1150V | 100 | 14:13:16 | 15:23:44 (4228 s) | at the beginning was 4.9 then later changed to 13.5111 events/s | TDC3->DL, TDC12->DR | CFD out-put of DL (to TDC9) and DR (to TDC10) connected to TDC. (Connected to the seond row of LeCroy Model 4616.) |
3691 | in | out | out | (DL & GunTrig) AND (DR & GunTrig) | 1150V | 1150V | 100 | 16:13:16 | 16:36:19 (1383 s) | 14.2079 events/s | TDC3->DL, TDC12->DR | High voltage of scintilator at 90 deg exit set to 1300 V. Scintlator at 90 deg moved to CH13. |
3692 | in | in | in | (DL & GunTrig) AND (DR & GunTrig) | 1150V | 1150V | 100 | 16:40:18 | 17:00:56 (1238 s) | 23.9270 events/s | TDC3->DL, TDC12->DR | |
3694 | in | in | in | (DL & GunTrig) AND (DR & GunTrig) | 1150V | 1150V | 100 | 10 events/s | TDC3->DL, TDC12->DR | |||
3696 | in | in | in | (DL & GunTrig) AND (DR & GunTrig) | 1150V | 1150V | 100 | 17:30:08 | 17:31:39 | 10 events/s | TDC3->DL, TDC12->DR |
Data Analysis
r3686 and r3687: HRRL_Pos_Test_26Jul2012_Data_Ana_com_r3686_r3687
r3688 and r3689: HRRL_Pos_Test_26Jul2012_Data_Ana_com_r3688_r3689
r3690 and r3691: HRRL_Pos_Test_26Jul2012_Data_Ana_com_r3688_r3689
Subtracted Sepctrums: HRRL_Pos_Test_26Jul2012_Data_Ana_Sub_Spcs
Detector Calibrations
run# | T1 | T2 | Sweep Magnet | logic | HV-D2L | HV-D1R | Reprate | start | stop | Int Rate | comments |
3679 | in | in | out | (DL & GunTrig) or (DR & GunTrig) | 1060V | 1080V | 100 | 08:48:00 | 09:28:01 | 42.3 events/s | ADC12 has Sintilator1 (Sint @ exit of 90 deg). |
3680 | in | OUT | out | (DL & GunTrig) or (DR & GunTrig) | 1060V | 1080V | 100 | 09:29:37 | 09:46:19 | 42.6 events/s | ADC12 has Sintilator1 (Sint @ exit of 90 deg). |
3682 | in | OUT | out | (DL & GunTrig) AND (DR & GunTrig) | 1150V | 1150V | 100 | 09:55:20 | 10:33:56 | 4.8884 events/s | TDC cable was upside down, now it is flipped. Sintilator left was put to CH9. |
3683 | in | IN | out | (DL & GunTrig) AND (DR & GunTrig) | 1150V | 1150V | 100 | 10:36:03 | 10:53:48 | 5.9397 events/s | TDC5->beam left |
3684 | in | IN | out | (DL & GunTrig) AND (DR & GunTrig) | 1150V | 1150V | 100 | 10:54:19 | 11:25:01 | 8.7974 events/s | TDC3->beam right, TDC5->beam left |
3685 | in | out | out | (DL & GunTrig) AND (DR & GunTrig) | 1150V | 1150V | 100 | 11:26:43 | 11:31:01 | 18 events/s | junk, rate are too high. |
We changed the discriminator from Octal discriminator (LeCroy modle 623B) to 8 CH Constant Fraction Discriminator (CFD) (CAEN model N842). The threshold on the detector beam left is 29, and on the detector right is 31.
Following is the image of coincidence of left and right detectors using this CFD.
run# | T1 | T2 | Sweep Magnet | logic | HV-D2L | HV-D1R | Reprate | start | stop | Int Rate | Connections | comments |
3686 | in | in | out | (DL & GunTrig) AND (DR & GunTrig) | 1150V | 1150V | 100 | 11:54:44 | 12:15:44 (1260 s) | 4.8262 events/s | TDC3->DL, TDC->DR | |
3687 | in | out | out | (DL & GunTrig) AND (DR & GunTrig) | 1150V | 1150V | 100 | 12:19:15 | 12:43:14 (1411 s) | 3.6 events/s | TDC3->DL, TDC5->DR | |
3688 | in | in | out | (DL & GunTrig) AND (DR & GunTrig) | 1150V | 1150V | 100 | 12:44:09 | 13:25:09 (2460 s) | 4.9743 events/s | TDC3->DL, TDC12->DR | switched DR from TDC5 to TDC12 |
3689 | in | out | out | (DL & GunTrig) AND (DR & GunTrig) | 1150V | 1150V | 100 | 13:27:23 | 14:12:29 (2706 s) | 3.3156 events/s | TDC3->DL, TDC12->DR | CFD out-put of DL (to TDC9) and DR (to TDC10) connected to TDC. (Connected to the top row of LeCroy Model 4616.) |
3690 | in | in | out | (DL & GunTrig) AND (DR & GunTrig) | 1150V | 1150V | 100 | 14:13:16 | 15:23:44 (4228 s) | at the beginning was 4.9 then later changed to 13.5111 events/s | TDC3->DL, TDC12->DR | CFD out-put of DL (to TDC9) and DR (to TDC10) connected to TDC. (Connected to the seond row of LeCroy Model 4616.) |
3691 | in | out | out | (DL & GunTrig) AND (DR & GunTrig) | 1150V | 1150V | 100 | 14:13:16 | 16:36:19 | 14.2079 events/s | TDC3->DL, TDC12->DR | High voltage of scintilator at 90 deg exit set to 1300 V. Scintlator at 90 deg moved to CH13. |
3692 | in | in | in | (DL & GunTrig) AND (DR & GunTrig) | 1150V | 1150V | 100 | 16:40:18 | 17:00:56 | 23.9270 events/s | TDC3->DL, TDC12->DR | |
3694 | in | in | in | (DL & GunTrig) AND (DR & GunTrig) | 1150V | 1150V | 100 | 10 events/s | TDC3->DL, TDC12->DR | |||
3696 | in | in | in | (DL & GunTrig) AND (DR & GunTrig) | 1150V | 1150V | 100 | 17:30:08 | 17:31:39 | 10 events/s | TDC3->DL, TDC12->DR |
Beam Current Calibration of Scint Left
Determine if the Beam Left Scintillator is linear with beam current.
run# | Source | HV-D2L | HV-D1R | start | stop | Int Rate | Connections | comments |
3699 | Na22 | 1150V | 1150V | 18:20:47 | 18:21:33 | 103.6 events/s | TDC3->DL, TDC->DR | A lot of error message saying no data in TDC and ADC. |
3700 | Na22 | 1150V | 1150V | 18:26:15 | 20:45:47 | 103.6 events/s | TDC3->DL, TDC->DR |
[Go Back]