Difference between revisions of "PePPO 03242012"
(10 intermediate revisions by the same user not shown) | |||
Line 10: | Line 10: | ||
{| border="1" |cellpadding="20" cellspacing="0 | {| border="1" |cellpadding="20" cellspacing="0 | ||
|- | |- | ||
− | |[[Image:03242012_V895_DoublPuls.png | 200 px]] | + | |[[Image:03242012_V895_DoublPuls.png | 200 px]] || [[Image:Run2905_400Puls.png | 200 px]] |
|- | |- | ||
− | | The V895 "Maj" output is shown on Ch1 along with the NaI signal which was discriminated by it | + | | The V895 "Maj" output is shown on Ch1 along with the NaI signal which was discriminated by it || You can see some pulses are measured after the pulse peak because of multiple hits out of the discriminator |
|} | |} | ||
Line 21: | Line 21: | ||
{| border="1" |cellpadding="20" cellspacing="0 | {| border="1" |cellpadding="20" cellspacing="0 | ||
|- | |- | ||
− | |[[Image:03242012_V895_LongrPuls.png | 200 px]] | + | |[[Image:03242012_V895_LongrPuls.png | 200 px]] || [[Image:Run4079_1000Puls.png | 200 px]] |
|- | |- | ||
− | | I took the "MAJ" output and used a Phillips Gate Generator to extend the pulse length in order to stop multiple triggers | + | | I took the "MAJ" output and used a Phillips Gate Generator to extend the pulse length in order to stop multiple triggers || Extending thepulse length out of the discriminstor by adding a gate generator removes triggers that happened after the pulse output began. |
|} | |} | ||
+ | |||
+ | |||
+ | [[Image:Run4086_1000Puls.png | 200 px]] | ||
=Run list= | =Run list= | ||
Line 95: | Line 98: | ||
T->MakeClass(); | T->MakeClass(); | ||
− | ==Run | + | ==Calibration runs== |
+ | |||
+ | {| border="1" |cellpadding="20" cellspacing="0 | ||
+ | |- | ||
+ | | Run#|| run mode|| Run length (sec)||Source||Trig Logic|| VME discriminator Threshold (mV)|| HV1 (setting/V) A1|| HV2 (setting/V) A3|| PL (setting/ns)|| PTW (setting/ns) | ||
+ | |- | ||
+ | | 4087 || sample || || Co-60 || (Det1 OR Det2) || 50 || 2.4/0.888 || 3.0/1.2 || || | ||
+ | |- | ||
+ | | 4088 || sample || || Na-22 (0.1 uCi)|| (Det1 OR Det2) || 50 || 2.4/0.888 || 3.0/1.2 || || | ||
+ | |- | ||
+ | | 4089 || sample || || Na-22 (10 uCi)|| (Det1 OR Det2) || 50 || 2.4/0.888 || 3.0/1.2 || || | ||
+ | |- | ||
+ | | 4090 || sample || || Na-22 (10 uCi)|| (Det1 AND Det2) || 50 || 2.4/0.888 || 3.0/1.2 || || | ||
+ | |- | ||
+ | | 4090 || sample || || Na-22 (10 uCi)|| (Det1 AND Det2) || 50 || 2.4/0.888 || 3.0/1.2 || || | ||
+ | |- | ||
+ | | 4092 || sample || || Na-22 (10 uCi)|| (Det1 OR Det2) || 50 || 2.2/0.888 || 3.0/1.2 || || | ||
+ | |- | ||
+ | | 4093 || sample || || Co-60 (10 uCi)|| (Det1 OR Det2) || 50 || 2.2/0.888 || 3.0/1.2 || || | ||
+ | |- | ||
+ | | 4094 || sample || || Na-22 (10 uCi)|| (Det1AND Det2) || 50 || 2.2/0.888 || 3.0/1.2 || || | ||
+ | |} | ||
+ | |||
+ | ==Deadtime== | ||
Line 123: | Line 149: | ||
| 4086 || 0.05 || 1454 || (Det1 AND Det2) || 128247/128246 | | 4086 || 0.05 || 1454 || (Det1 AND Det2) || 128247/128246 | ||
|} | |} | ||
+ | |||
+ | ==Energy Spectra -vs- Beam current== | ||
+ | |||
+ | |||
+ | |||
+ | {| border="1" |cellpadding="20" cellspacing="0 | ||
+ | |- | ||
+ | | Run#|| Current || Run length (sec) ||Trig Logic|| Integrated Rate Det 1|| Integrated Rate Det 2 | ||
+ | |- | ||
+ | | 4082 || 0.3 || 1421 || (Det1 AND Det2 AND SYNC) || [[File:r4082_Int15.png | 200 px]] || [[File:r4082_Int16.png | 200 px]] | ||
+ | |- | ||
+ | | 4080 || 1 || 929 ||(Det1 AND Det2 AND SYNC)|| [[File:r4080_Int15.png | 200 px]] || [[File:r4080_Int16.png | 200 px]] | ||
+ | |- | ||
+ | | 4081 || 3 || 523 ||(Det1 AND Det2 AND SYNC)|| [[File:r4081_Int15.png | 200 px]] || [[File:r4081_Int16.png | 200 px]] | ||
+ | |} | ||
+ | |||
+ | |||
+ | I get higher energy photons in the energy spectra as the beam current goes down? | ||
+ | |||
+ | ==Integral pics== | ||
+ | |||
+ | |||
+ | |||
+ | {| border="1" |cellpadding="20" cellspacing="0 | ||
+ | |- | ||
+ | | Run#|| Current || Run length (sec) ||Trig Logic|| Integrated Rate Det 1|| Integrated Rate Det 2 | ||
+ | |- | ||
+ | | 4073 || 0 || 263|| (Det1 OR Det2) || | ||
+ | |- | ||
+ | | 4078 || 0.3 || 166 || (Det1 AND Det2 AND SYNC)|| | ||
+ | |- | ||
+ | | 4079 || 0.3 || 309 ||(Det1 AND Det2) || | ||
+ | |- | ||
+ | | 4080 || 1 || 929 ||(Det1 AND Det2 AND SYNC)|| [[File:r4080_Int15.png | 200 px]] || [[File:r4080_Int16.png | 200 px]] | ||
+ | |- | ||
+ | | 4081 || 3 || 523 ||(Det1 AND Det2 AND SYNC)|| [[File:r4081_Int15.png | 200 px]] || [[File:r4081_Int16.png | 200 px]] | ||
+ | |- | ||
+ | | 4082 || 0.3 || 1421 || (Det1 AND Det2 AND SYNC) || [[File:r4082_Int15.png | 200 px]] || [[File:r4082_Int16.png | 200 px]] | ||
+ | |- | ||
+ | | 4083 || 0.3 ||500 || (Det1 AND Det2) || | ||
+ | |- | ||
+ | | 4084 || 0.15 ||448 || (Det1 AND Det2) || | ||
+ | |- | ||
+ | | 4085 || 0.15 || 1215 || (Det1 AND Det2 AND SYNC) || | ||
+ | |- | ||
+ | | 4086 || 0.05 || 1454 || (Det1 AND Det2) || | ||
+ | |} | ||
+ | |||
+ | |||
+ | [[Positrons]] |
Latest revision as of 16:44, 3 April 2012
Oportunistic Beam studies
Trigger Setup
1.) The V895 was double pulsing and its output pulse maxed out at 40 ns
2.) inserted a gate delay generator to change the V895 output pulse to a microsecond
Run list
Run# | run mode | Run length (sec) | Beam | Target | Trig Logic | VME discriminator Threshold (mV) | HV1 (setting/V) A1 | HV2 (setting/V) A3 | PL (setting/ns) | PTW (setting/ns) |
4073 | sample | 240 | off | empty | (Det1 OR Det2) | 200 | 2.0/0.888 | 2.7/1.2 | ||
4074 | sample | off/tuning | empty | (Det1 AND Det2) | 50 | 2.2/0.888 | 3.0/1.2 | |||
4075 | sample | off/tuning | Chromox | (Det1 AND Det2 AND SYNC) | 50 | 2.2/0.888 | 3.0/1.2 | |||
4078 | sample | e+ (0.3 nA) | Chromox | (Det1 AND Det2 AND SYNC)** | 50 | 2.2/0.888 | 3.0/1.2 | |||
4079 | sample | e+ (0.3nA) | Chromox | (Det1 AND Det2)** | 50 | 2.2/0.888 | 3.0/1.2 | |||
4080 | sample | e+(1 na) | Chromox | (Det1 AND Det2 AND SYNC)** | 50 | 2.2/0.888 | 3.0/1.2 | |||
4081 | sample | 540 sec | e+(3 na) | Chromox | (Det1 AND Det2 AND SYNC)** | 50 | 2.2/0.888 | 3.0/1.2 | ||
4082 | sample | 23 min | e+(0.3 na) | Chromox | (Det1 AND Det2 AND SYNC)** | 50 | 2.2/0.888 | 3.0/1.2 | ||
4083 | sample | sec | e+(0.3 na) | Chromox | (Det1 AND Det2) | 50 | 2.2/0.888 | 3.0/1.2 | ||
4084 | sample | sec | e+(0.150 na) | Chromox | (Det1 AND Det2) | 50 | 2.2/0.888 | 3.0/1.2 | Changed discriminator to 50 mv for both channels it was at 20 the scalers will count differently now but the trigger theshold is still the same 100 mV | |
4085 | sample | sec | e+(0.15 na) | Chromox | (Det1 AND Det2 AND SYNC) | 50 | 2.2/0.888 | 3.0/1.2 | ||
4086 | sample | sec | e+(0.05 na) | Chromox | (Det1 AND Det2) | 50 | 2.2/0.888 | 3.0/1.2 |
- stop using V895 and used a discriminator (Phillips 708 set to ~100 mV) and logic unit (phillips 758)
Scope pict traces for TEK0020 -> TEK00
Ch1(Yellow)=Majority
Ch2(Blue)=Beam SYNC
Ch3(Pink)=Annih. Det. #1
Ch4(Green)=Annih. Det. #2
Data Analysis
Replay
running the decoder using the personal account on the opsmdaq0 failed; perhaps due to problems writing its output to the /data disk
to convert a data file into a rood file I needed to login to opsdaq0 under the username "idaq"
then execute the following command
/work1/fanchini/decoder/fadc_decode_Source /data/annihilation/Annihilation_Sample_4086.dat
This created the summary text file and root file for run 4086
MakeClass
Load up the ntuple and click over to the "T" tree
Then onthe command line type
T->MakeClass();
Calibration runs
Run# | run mode | Run length (sec) | Source | Trig Logic | VME discriminator Threshold (mV) | HV1 (setting/V) A1 | HV2 (setting/V) A3 | PL (setting/ns) | PTW (setting/ns) |
4087 | sample | Co-60 | (Det1 OR Det2) | 50 | 2.4/0.888 | 3.0/1.2 | |||
4088 | sample | Na-22 (0.1 uCi) | (Det1 OR Det2) | 50 | 2.4/0.888 | 3.0/1.2 | |||
4089 | sample | Na-22 (10 uCi) | (Det1 OR Det2) | 50 | 2.4/0.888 | 3.0/1.2 | |||
4090 | sample | Na-22 (10 uCi) | (Det1 AND Det2) | 50 | 2.4/0.888 | 3.0/1.2 | |||
4090 | sample | Na-22 (10 uCi) | (Det1 AND Det2) | 50 | 2.4/0.888 | 3.0/1.2 | |||
4092 | sample | Na-22 (10 uCi) | (Det1 OR Det2) | 50 | 2.2/0.888 | 3.0/1.2 | |||
4093 | sample | Co-60 (10 uCi) | (Det1 OR Det2) | 50 | 2.2/0.888 | 3.0/1.2 | |||
4094 | sample | Na-22 (10 uCi) | (Det1AND Det2) | 50 | 2.2/0.888 | 3.0/1.2 |
Deadtime
Run# | Current | Run length (sec) | Trig Logic | L1A/L1 |
4073 | 0 | 263 | (Det1 OR Det2) | 33895/33896 |
4078 | 0.3 | 166 | (Det1 AND Det2 AND SYNC) | 950/950 |
4079 | 0.3 | 309 | (Det1 AND Det2) | 411867/411851 |
4080 | 1 | 929 | (Det1 AND Det2 AND SYNC) | 9172/9172 |
4081 | 3 | 523 | (Det1 AND Det2 AND SYNC) | 4420/4420 |
4082 | 0.3 | 1421 | (Det1 AND Det2 AND SYNC) | 5184/5185 |
4083 | 0.3 | 500 | (Det1 AND Det2) | 589251/589231 |
4084 | 0.15 | 448 | (Det1 AND Det2) | 304192/304187 |
4085 | 0.15 | 1215 | (Det1 AND Det2 AND SYNC) | 2296/2296 |
4086 | 0.05 | 1454 | (Det1 AND Det2) | 128247/128246 |
Energy Spectra -vs- Beam current
Run# | Current | Run length (sec) | Trig Logic | Integrated Rate Det 1 | Integrated Rate Det 2 |
4082 | 0.3 | 1421 | (Det1 AND Det2 AND SYNC) | ||
4080 | 1 | 929 | (Det1 AND Det2 AND SYNC) | ||
4081 | 3 | 523 | (Det1 AND Det2 AND SYNC) |
I get higher energy photons in the energy spectra as the beam current goes down?