Difference between revisions of "PePPO 03132012"

From New IAC Wiki
Jump to navigation Jump to search
 
(21 intermediate revisions by 2 users not shown)
Line 75: Line 75:
 
| 2904|| sample  ||  452  ||e<math>^-</math>    ||  Empty  ||  Det1 AND Det2 ||        50            ||  2.4/960      ||    3/1200      || 12.2563% <math>\pm</math> 2.32772%    ||     
 
| 2904|| sample  ||  452  ||e<math>^-</math>    ||  Empty  ||  Det1 AND Det2 ||        50            ||  2.4/960      ||    3/1200      || 12.2563% <math>\pm</math> 2.32772%    ||     
 
|-
 
|-
| 2905|| sample  ||  556  || e<math>^-</math> || Chromox  ||  Det1 AND Det2 ||        50          ||  2.4/960      ||    3/1200      ||  3.16901% <math>\pm</math> 4.2301%      ||     
+
| 2905|| sample  ||  556  || e<math>^-</math> || Chromox  ||  Det1 AND Det2 ||        50          ||  2.4/960      ||    3/1200      ||  3.16901% <math>\pm</math> 4.2301%      ||    [[File:peppo-032302012-r2905-A1.png | 300 px]] [[File:peppo-032302012-r2905-A2.png | 300 px]]
 
|-
 
|-
 
| 2908|| sample  || 150 ||        off      ||    out    ||  (Det1 or Det2) ||      50          ||  2.4/960      ||    3/1200      ||  74.0667% <math>\pm</math>  0.157206%  ||     
 
| 2908|| sample  || 150 ||        off      ||    out    ||  (Det1 or Det2) ||      50          ||  2.4/960      ||    3/1200      ||  74.0667% <math>\pm</math>  0.157206%  ||     
Line 84: Line 84:
 
|}
 
|}
  
Why there is negative?
 
  
[[File:peppo-032302012-r2902-int14-negative.png | 300 px]]
+
= data analysis =
 +
 
 +
==Negative Energy ? ==
 +
 
 +
This run 2902, Int_Ch14 histogram. Why there is negative?
 +
 
 +
[[File:peppo-032302012-r2902-int14-negative.png | 700 px]]
 +
 
 +
== Pulses ==
 +
 
 +
=== run 2905 ===
 +
 
 +
1st 100 puleses
 +
 
 +
[[File:PePPO-03132012-pulses-2905-1to100.png | 400 px]]
 +
 
 +
pulse that only partially read-in:
 +
 
 +
[[File:PePPO-03132012-pulses-2905-1-pulse-cut-off.png | 400 px]]
 +
 
 +
[[File:PePPO-03132012-pulses-2905-9th-pulse-cut-off.png | 400 px]]
 +
 
 +
first 10 pulses, some pulses that only partially read-in:
 +
 
 +
[[File:PePPO-03132012-pulses-2905-10-pulse-cut-off.png | 400 px]]
 +
 
 +
 
 +
looks like some pulses are only partially read-in. We need to delay signal more?
 +
 
 +
One possible cut may to determine if the signal is increasing within the first 100 - 300 $Iterations.  This would be like creating a leading edge discriminator in software.
 +
 
 +
 
 +
 
 +
 
 +
 
 +
 
 +
T->Draw("CH14:Iteration$","Iteration$<300 & CH14<100","L",10,1)
 +
 
 +
looks like it just won't draw. I need to know the to way exclude that event does not satisfy "Iteration$<300 & CH14<100" requirement.
 +
 
 +
[[File:PePPO-03132012-pulses-selection-try-1.png | 400 px]]
 +
 
 +
 
 +
 
 +
 
 +
 
 +
 
 +
Why CH14 histogram is all positive, but int_14 histogram has negatives?
 +
 
 +
[[File:PePPO-03132012-pulses-2905-CH14-Hist.png | 400 px]]
 +
 
 +
[[File:PePPO-03132012-pulses-2905-int_14-Hist.png | 400 px]]
 +
 
 +
==TF Analysis==
 +
 
 +
 
 +
Using the following ROOT programs
 +
 
 +
[[Media:03212012_T.C.txt]]
 +
 
 +
[[Media:03212012_T.h.txt]]
 +
 
 +
 
 +
 
 +
{| border="1"  |cellpadding="20" cellspacing="0
 +
|-
 +
| Run#|| run mode|| Run length (sec)||Beam||Target ||Trig Logic||  VME discriminator Threshold (mV)|| HV1 (setting/V) || HV2 (setting/V) || Deadtime || 
 +
|-
 +
| 2902|| sample  ||  0    || e<math>^-</math> ||  Chromox  || (Det1 OR Det2) ||        50            ||  2.4/960      ||    3/1200      ||  93.1818% <math>\pm</math> 10.3874%    ||
 +
|-
 +
| 2903|| sample  || 203 ||e<math>^-</math> ||  Empty    || (Det1 OR Det2) ||        50            ||  2.4/960      ||    3/1200      ||  84.9788% <math>\pm</math>0.106622%    ||   
 +
|-
 +
| 2904|| sample  ||  452  ||e<math>^-</math>    ||  Empty  ||  Det1 AND Det2 ||        50            ||  2.4/960      ||    3/1200      || 12.2563% <math>\pm</math> 2.32772%    ||    [[File:03212012_Ch14r2904.png| 200 px]]
 +
|-
 +
| 2905|| sample  ||  556  || e<math>^-</math> || Chromox  ||  Det1 AND Det2 ||        50          ||  2.4/960      ||    3/1200      ||  3.16901% <math>\pm</math> 4.2301%      ||    [[File:03212012_Ch14r2905.png| 200 px]]
 +
|-
 +
| 2908|| sample  || 150 ||        off      ||    out    ||  (Det1 or Det2) ||      50          ||  2.4/960      ||    3/1200      ||  74.0667% <math>\pm</math>  0.157206%  ||    [[File:03212012_Ch14r2908.png| 200 px]]
 +
|-
 +
| 2909|| sample  || 1118 || off  ||  out  ||    (Det1 AND Det2) ||                      50          ||  2.4/960      ||    3/1200      ||    3.57143% <math>\pm</math>26.9725%    ||   
 +
|-
 +
 
 +
|}
 +
 
 +
Run 2904 and 2905 coincidence overlay (NOT Deadtime corrected)
 +
 
 +
[[File:03212012_Coinc2904-5.png | 200 px]]
 +
 
 +
Normalizing using run time and correcting for DAQ deadtime:
 +
 
 +
[[File:03212012_NormCoinc2904-5.png | 200 px]]

Latest revision as of 06:35, 22 March 2012

Accelerator Configuration

Accelerator Settings
beam energy 6.8 MeV
Beam type: Electron
I_average: 0.3 nA
I_peak: 0.05[math] \mu[/math] A 0.006 duty factor
pulse width: 10 [math]\mu[/math]s (it seems on 200 [math] \mu[/math]s is possible?)
Rate: 60 Hz
PEPPO Spectrometer: 151.3 A

Set HV to Detectors

Signal channel HV dac HV (V)
Detectors #1 A1 2.4 960
Detectors #2 A2 3.0 1200

How to turn HV On


Scope Pictures

CH1: Beam Sync CH2: A1 CH3: A2

Target Pictures
No Target 1 03132012 AnnhCountScopePic.png 2 03132012 AnnhCountScopePic.png
Chromox Target 3 03132012 AnnhCountScopePic.png 4 03132012 AnnhCountScopePic.png

Annihilation Counter Runs

Signal Definitions for Trigger logic
SYNC:=60 Hz sync pulse 200 micro seconds long
Det1:= Annihilation Counter #1 (HV1)
Det2:=Annihilation Counter #2 (HV2)



Run# run mode Run length (sec) Beam Target Trig Logic VME discriminator Threshold (mV) HV1 (setting/V) HV2 (setting/V) Deadtime
2902 sample 0 e[math]^-[/math] Chromox (Det1 OR Det2) 50 2.4/960 3/1200 93.1818% [math]\pm[/math] 10.3874%
2903 sample 203 e[math]^-[/math] Empty (Det1 OR Det2) 50 2.4/960 3/1200 84.9788% [math]\pm[/math]0.106622%
2904 sample 452 e[math]^-[/math] Empty Det1 AND Det2 50 2.4/960 3/1200 12.2563% [math]\pm[/math] 2.32772%
2905 sample 556 e[math]^-[/math] Chromox Det1 AND Det2 50 2.4/960 3/1200 3.16901% [math]\pm[/math] 4.2301% Peppo-032302012-r2905-A1.png Peppo-032302012-r2905-A2.png
2908 sample 150 off out (Det1 or Det2) 50 2.4/960 3/1200 74.0667% [math]\pm[/math] 0.157206%
2909 sample 1118 off out (Det1 AND Det2) 50 2.4/960 3/1200 3.57143% [math]\pm[/math]26.9725%


data analysis

Negative Energy ?

This run 2902, Int_Ch14 histogram. Why there is negative?

Peppo-032302012-r2902-int14-negative.png

Pulses

run 2905

1st 100 puleses

PePPO-03132012-pulses-2905-1to100.png

pulse that only partially read-in:

PePPO-03132012-pulses-2905-1-pulse-cut-off.png

PePPO-03132012-pulses-2905-9th-pulse-cut-off.png

first 10 pulses, some pulses that only partially read-in:

PePPO-03132012-pulses-2905-10-pulse-cut-off.png


looks like some pulses are only partially read-in. We need to delay signal more?
One possible cut may to determine if the signal is increasing within the first 100 - 300 $Iterations.  This would be like creating a leading edge discriminator in software.




T->Draw("CH14:Iteration$","Iteration$<300 & CH14<100","L",10,1)

looks like it just won't draw. I need to know the to way exclude that event does not satisfy "Iteration$<300 & CH14<100" requirement.

PePPO-03132012-pulses-selection-try-1.png




Why CH14 histogram is all positive, but int_14 histogram has negatives?

PePPO-03132012-pulses-2905-CH14-Hist.png

PePPO-03132012-pulses-2905-int 14-Hist.png

TF Analysis

Using the following ROOT programs

Media:03212012_T.C.txt

Media:03212012_T.h.txt


Run# run mode Run length (sec) Beam Target Trig Logic VME discriminator Threshold (mV) HV1 (setting/V) HV2 (setting/V) Deadtime
2902 sample 0 e[math]^-[/math] Chromox (Det1 OR Det2) 50 2.4/960 3/1200 93.1818% [math]\pm[/math] 10.3874%
2903 sample 203 e[math]^-[/math] Empty (Det1 OR Det2) 50 2.4/960 3/1200 84.9788% [math]\pm[/math]0.106622%
2904 sample 452 e[math]^-[/math] Empty Det1 AND Det2 50 2.4/960 3/1200 12.2563% [math]\pm[/math] 2.32772% 03212012 Ch14r2904.png
2905 sample 556 e[math]^-[/math] Chromox Det1 AND Det2 50 2.4/960 3/1200 3.16901% [math]\pm[/math] 4.2301% 03212012 Ch14r2905.png
2908 sample 150 off out (Det1 or Det2) 50 2.4/960 3/1200 74.0667% [math]\pm[/math] 0.157206% 03212012 Ch14r2908.png
2909 sample 1118 off out (Det1 AND Det2) 50 2.4/960 3/1200 3.57143% [math]\pm[/math]26.9725%

Run 2904 and 2905 coincidence overlay (NOT Deadtime corrected)

03212012 Coinc2904-5.png

Normalizing using run time and correcting for DAQ deadtime:

03212012 NormCoinc2904-5.png