=Accelerator Configuration= | |||
;Accelerator Settings | |||
{| border="1" | |||
| beam energy || 6.8 MeV | |||
|- | |||
| Beam type: || Electron | |||
|- | |||
| I_average: || 0.3 nA | |||
|- | |||
| I_peak: || 0.05<math> \mu</math> A 0.006 duty factor | |||
|- | |||
| pulse width:|| 10 <math>\mu</math>s (it seems on 200 <math> \mu</math>s is possible?) | |||
|- | |||
| Rate:|| 60 Hz | |||
|- | |||
| PEPPO Spectrometer:|| 151.3 A | |||
|- | |||
|} | |||
= Set HV to Detectors = | |||
{| border="1" |cellpadding="20" cellspacing="0 | |||
|- | |||
| || Signal channel || HV dac || HV (V) | |||
|- | |||
| Detectors #1 || A1 || 2.4 || 960 | |||
|- | |||
| Detectors #2 || A2 || 3.0 || 1200 | |||
|- | |||
|} | |||
[https://positron.jlab.org/wiki/index.php/PEPPo_Annihilation_Counter_Descriptions#Turning_on_HV_PS How to turn HV On] | |||
=Scope Pictures= | |||
CH1: Beam Sync | |||
CH2: A1 | |||
CH3: A2 | |||
{| border="1" |cellpadding="20" cellspacing="0 | |||
|- | |||
| Target|| Pictures | |||
|- | |||
| No Target || [[File:1_03132012_AnnhCountScopePic.png| 200 px]] [[File:2_03132012_AnnhCountScopePic.png| 200 px]] | |||
|- | |||
| Chromox Target|| [[File:3_03132012_AnnhCountScopePic.png| 200 px]] [[File:4_03132012_AnnhCountScopePic.png| 200 px]] | |||
|} | |||
= Annihilation Counter Runs= | |||
;Signal Definitions for Trigger logic | |||
:SYNC:=60 Hz sync pulse 200 micro seconds long | |||
:Det1:= Annihilation Counter #1 (HV1) | |||
:Det2:=Annihilation Counter #2 (HV2) | |||
{| border="1" |cellpadding="20" cellspacing="0 | |||
|- | |||
| Run#|| run mode|| Run length (sec)||Beam||Target ||Trig Logic|| VME discriminator Threshold (mV)|| HV1 (setting/V) || HV2 (setting/V) || Deadtime || | |||
|- | |||
| 2902|| sample || 0 || e<math>^-</math> || Chromox || (Det1 OR Det2) || 50 || 2.4/960 || 3/1200 || 93.1818% <math>\pm</math> 10.3874% || | |||
|- | |||
| 2903|| sample || 203 ||e<math>^-</math> || Empty || (Det1 OR Det2) || 50 || 2.4/960 || 3/1200 || 84.9788% <math>\pm</math>0.106622% || | |||
|- | |||
| 2904|| sample || 452 ||e<math>^-</math> || Empty || Det1 AND Det2 || 50 || 2.4/960 || 3/1200 || 12.2563% <math>\pm</math> 2.32772% || | |||
|- | |||
| 2905|| sample || 556 || e<math>^-</math> || Chromox || Det1 AND Det2 || 50 || 2.4/960 || 3/1200 || 3.16901% <math>\pm</math> 4.2301% || [[File:peppo-032302012-r2905-A1.png | 300 px]] [[File:peppo-032302012-r2905-A2.png | 300 px]] | |||
|- | |||
| 2908|| sample || 150 || off || out || (Det1 or Det2) || 50 || 2.4/960 || 3/1200 || 74.0667% <math>\pm</math> 0.157206% || | |||
|- | |||
| 2909|| sample || 1118 || off || out || (Det1 AND Det2) || 50 || 2.4/960 || 3/1200 || 3.57143% <math>\pm</math>26.9725% || | |||
|- | |||
|} | |||
= data analysis = | |||
==Negative Energy ? == | |||
This run 2902, Int_Ch14 histogram. Why there is negative? | |||
[[File:peppo-032302012-r2902-int14-negative.png | 700 px]] | |||
== Pulses == | |||
=== run 2905 === | |||
1st 100 puleses | |||
[[File:PePPO-03132012-pulses-2905-1to100.png | 400 px]] | |||
pulse that only partially read-in: | |||
[[File:PePPO-03132012-pulses-2905-1-pulse-cut-off.png | 400 px]] | |||
[[File:PePPO-03132012-pulses-2905-9th-pulse-cut-off.png | 400 px]] | |||
first 10 pulses, some pulses that only partially read-in: | |||
[[File:PePPO-03132012-pulses-2905-10-pulse-cut-off.png | 400 px]] | |||
looks like some pulses are only partially read-in. We need to delay signal more? | |||
One possible cut may to determine if the signal is increasing within the first 100 - 300 $Iterations. This would be like creating a leading edge discriminator in software. | |||
T->Draw("CH14:Iteration$","Iteration$<300 & CH14<100","L",10,1) | |||
looks like it just won't draw. I need to know the to way exclude that event does not satisfy "Iteration$<300 & CH14<100" requirement. | |||
[[File:PePPO-03132012-pulses-selection-try-1.png | 400 px]] | |||
Why CH14 histogram is all positive, but int_14 histogram has negatives? | |||
[[File:PePPO-03132012-pulses-2905-CH14-Hist.png | 400 px]] | |||
[[File:PePPO-03132012-pulses-2905-int_14-Hist.png | 400 px]] | |||
==TF Analysis== | |||
Using the following ROOT programs | |||
[[Media:03212012_T.C.txt]] | |||
[[Media:03212012_T.h.txt]] | |||
{| border="1" |cellpadding="20" cellspacing="0 | |||
|- | |||
| Run#|| run mode|| Run length (sec)||Beam||Target ||Trig Logic|| VME discriminator Threshold (mV)|| HV1 (setting/V) || HV2 (setting/V) || Deadtime || | |||
|- | |||
| 2902|| sample || 0 || e<math>^-</math> || Chromox || (Det1 OR Det2) || 50 || 2.4/960 || 3/1200 || 93.1818% <math>\pm</math> 10.3874% || | |||
|- | |||
| 2903|| sample || 203 ||e<math>^-</math> || Empty || (Det1 OR Det2) || 50 || 2.4/960 || 3/1200 || 84.9788% <math>\pm</math>0.106622% || | |||
|- | |||
| 2904|| sample || 452 ||e<math>^-</math> || Empty || Det1 AND Det2 || 50 || 2.4/960 || 3/1200 || 12.2563% <math>\pm</math> 2.32772% || [[File:03212012_Ch14r2904.png| 200 px]] | |||
|- | |||
| 2905|| sample || 556 || e<math>^-</math> || Chromox || Det1 AND Det2 || 50 || 2.4/960 || 3/1200 || 3.16901% <math>\pm</math> 4.2301% || [[File:03212012_Ch14r2905.png| 200 px]] | |||
|- | |||
| 2908|| sample || 150 || off || out || (Det1 or Det2) || 50 || 2.4/960 || 3/1200 || 74.0667% <math>\pm</math> 0.157206% || [[File:03212012_Ch14r2908.png| 200 px]] | |||
|- | |||
| 2909|| sample || 1118 || off || out || (Det1 AND Det2) || 50 || 2.4/960 || 3/1200 || 3.57143% <math>\pm</math>26.9725% || | |||
|- | |||
|} | |||
Run 2904 and 2905 coincidence overlay (NOT Deadtime corrected) | |||
[[File:03212012_Coinc2904-5.png | 200 px]] | |||
Normalizing using run time and correcting for DAQ deadtime: | |||
[[File:03212012_NormCoinc2904-5.png | 200 px]] | |||
Latest revision as of 06:35, 22 March 2012
Accelerator Configuration
- Accelerator Settings
| beam energy | 6.8 MeV |
| Beam type: | Electron |
| I_average: | 0.3 nA |
| I_peak: | 0.05[math]\displaystyle{ \mu }[/math] A 0.006 duty factor |
| pulse width: | 10 [math]\displaystyle{ \mu }[/math]s (it seems on 200 [math]\displaystyle{ \mu }[/math]s is possible?) |
| Rate: | 60 Hz |
| PEPPO Spectrometer: | 151.3 A |
Set HV to Detectors
| Signal channel | HV dac | HV (V) | |
| Detectors #1 | A1 | 2.4 | 960 |
| Detectors #2 | A2 | 3.0 | 1200 |
Scope Pictures
CH1: Beam Sync CH2: A1 CH3: A2
| Target | Pictures |
| No Target | |
| Chromox Target |
Annihilation Counter Runs
- Signal Definitions for Trigger logic
- SYNC:=60 Hz sync pulse 200 micro seconds long
- Det1:= Annihilation Counter #1 (HV1)
- Det2:=Annihilation Counter #2 (HV2)
data analysis
Negative Energy ?
This run 2902, Int_Ch14 histogram. Why there is negative?
Pulses
run 2905
1st 100 puleses
pulse that only partially read-in:
first 10 pulses, some pulses that only partially read-in:
looks like some pulses are only partially read-in. We need to delay signal more?
One possible cut may to determine if the signal is increasing within the first 100 - 300 $Iterations. This would be like creating a leading edge discriminator in software.
T->Draw("CH14:Iteration$","Iteration$<300 & CH14<100","L",10,1)
looks like it just won't draw. I need to know the to way exclude that event does not satisfy "Iteration$<300 & CH14<100" requirement.
Why CH14 histogram is all positive, but int_14 histogram has negatives?
TF Analysis
Using the following ROOT programs
Run 2904 and 2905 coincidence overlay (NOT Deadtime corrected)
Normalizing using run time and correcting for DAQ deadtime: