05/11/2011 neutron detector threshold calibration: Difference between revisions

From New IAC Wiki
Jump to navigation Jump to search
Shaproma (talk | contribs)
No edit summary
Shaproma (talk | contribs)
No edit summary
 
Line 1: Line 1:
 
[https://wiki.iac.isu.edu/index.php/N%27s_detector_calibration Go Back]





Latest revision as of 21:02, 19 May 2011

Go Back


Because is something wrong I decided to measure the counts rate as function of threshold energy. Maybe my threshold was incorrect.

Below are my results for n's detector with HAMAMATSU PMT (-1500 V):


So it looks like my previous threshold value (-500 mV) was not good. Let's set it up the -200 mV. And let's see does the n's detector sensitive to the source. The same set up and distances as before.

counts/sec (5 min test), no source counts/sec (5 min test), with source on the top of PS1
n's 19.89[math]\displaystyle{ \pm }[/math]0.26 (total 5966 cnts) 58.84[math]\displaystyle{ \pm }[/math]0.29 (total 17652 cnts)

So now we see the source. The ratio are:

[math]\displaystyle{ \frac{Rate_{source}}{Rate_{w/source}} = \frac{58.84}{19.89} = 2.96  }[/math]