Difference between revisions of "05/13/2011 coincidence test"

From New IAC Wiki
Jump to navigation Jump to search
 
(27 intermediate revisions by the same user not shown)
Line 1: Line 1:
 
[https://wiki.iac.isu.edu/index.php/N%27s_detector_calibration Go Back]
 
[https://wiki.iac.isu.edu/index.php/N%27s_detector_calibration Go Back]
  
Set up like that:
+
 
 +
Set up like that:  
  
 
[[File:Detector1.png | 400px]]
 
[[File:Detector1.png | 400px]]
  
the distance between PS and n's PMT is abot 80 cm
 
  
  
The voltages and thresholds  are as below. Note: I increased my thresholds for PS1 and PS2 from 80 mV up to 100 mV according to mu previous calibration threshold line.
+
The voltages, thresholds and widths are as below. Note: I increased my thresholds for PS1 and PS2 from 80 mV up to 100 mV according to my previous calibration threshold line.
 
{| border="1" cellspacing="0" style="text-align: center; width: 800px; height: 100px;"
 
{| border="1" cellspacing="0" style="text-align: center; width: 800px; height: 100px;"
 
! scope="col" width="50" |
 
! scope="col" width="50" |
Line 16: Line 16:
 
|-
 
|-
 
!PS1  
 
!PS1  
| -1500|| -80|| 100
+
| -1000|| -100|| 50
 
|-
 
|-
 
!PS1  
 
!PS1  
| -1500|| -80|| 100
+
| -1000|| -100|| 50
 
|-
 
|-
 
!n's  
 
!n's  
Line 28: Line 28:
  
  
'''Let's do first the coincidence test PS1+PS2'''
+
'''First let's do the PS1+PS2 coincidence test'''.
 +
 
 +
1. The distance between PS and n's PMT is about 80 cm
 
{| border="1" cellspacing="0" style="text-align: center; width: 800px; height: 100px;"
 
{| border="1" cellspacing="0" style="text-align: center; width: 800px; height: 100px;"
 
! scope="col" width="50" |  
 
! scope="col" width="50" |  
Line 45: Line 47:
 
|}
 
|}
  
 +
2. The distance between PS and n's PMT is about 50 cm
 +
{| border="1" cellspacing="0" style="text-align: center; width: 800px; height: 100px;"
 +
! scope="col" width="50" |
 +
! scope="col" width="50" | counts
 +
! scope="col" width="50" | time
 +
! scope="col" width="50" | counts/sec
 +
|-
 +
!PS1
 +
|<math>7513\pm86.5</math>||60 min||<math>2.09\pm0.02</math>
 +
|-
 +
!PS2
 +
|<math>9036\pm95.1</math>||60 min||<math>2.51\pm0.03</math>
 +
|-
 +
!PS1+PS2
 +
|<math>251\pm15.8</math>||60 min||<math>(6.9\pm0.4)\cdot 10^{-2}</math>
 +
|}
  
The scope picture of PS1+PS2 coincidence event like:
 
 
[[File:Cosmic coinc PS1 PS2.png | 800px]]
 
  
  
 +
'''Second do the PS1+PS2+n's coincidence test '''
  
 
+
1. The distance between PS and n's PMT is about 80 cm
'''Let's do second the coincidence test PS1+PS2+n's'''
 
 
{| border="1" cellspacing="0" style="text-align: center; width: 800px; height: 100px;"
 
{| border="1" cellspacing="0" style="text-align: center; width: 800px; height: 100px;"
 
! scope="col" width="50" |  
 
! scope="col" width="50" |  
Line 70: Line 85:
 
|}
 
|}
  
 +
2. The distance between PS and n's PMT is about 50 cm
 +
{| border="1" cellspacing="0" style="text-align: center; width: 800px; height: 100px;"
 +
! scope="col" width="50" |
 +
! scope="col" width="50" | counts
 +
! scope="col" width="50" | time
 +
! scope="col" width="50" | counts/sec
 +
|-
 +
!PS1
 +
|<math>7700\pm87.7</math>||60 min||<math>2.14\pm0.02</math>
 +
|-
 +
!PS2
 +
|<math>9060\pm95.2</math>||60 min||<math>2.52\pm0.03</math>
 +
|-
 +
!PS1+PS2+n's
 +
|<math>214\pm14.6</math>||60 min||<math>(5.9\pm0.4)\cdot 10^{-2}</math>
 +
|}
  
The scope picture of PS1+PS2 coincidence event like:
 
  
[[File:Cosmic coinc PS1 PS2 n.png | 800px]]
 
  
  
 +
The scope pictures of coincidence events looks like:
  
The ratio PS+PS2 rate to PS+PS2+n's rate is
+
{| border="0" cellspacing="0" style="text-align: center; width: 1050px; height: 100px;"
 
+
! scope="col" width="50" |PS1+PS2 event looks like:
<math>\frac{7.7}{4.9} = 1.6</math>
+
! scope="col" width="50" |PS1+PS2+n event looks like:
 +
|-
 +
|[[File:Cosmic coinc PS1 PS2.png | 500px]] || [[File:Cosmic coinc PS1 PS2 n.png | 500px]]
 +
|}

Latest revision as of 17:27, 16 May 2011

Go Back


Set up like that:

Detector1.png


The voltages, thresholds and widths are as below. Note: I increased my thresholds for PS1 and PS2 from 80 mV up to 100 mV according to my previous calibration threshold line.

PMT voltage, V threshold, mV discr. width, ns
PS1 -1000 -100 50
PS1 -1000 -100 50
n's -1500 -200 10



First let's do the PS1+PS2 coincidence test.

1. The distance between PS and n's PMT is about 80 cm

counts time counts/sec
PS1 2823±53.1 20 min 2.35±0.04
PS2 3091±55.6 20 min 2.58±0.05
PS1+PS2 93±9.6 20 min (7.7±0.8)102

2. The distance between PS and n's PMT is about 50 cm

counts time counts/sec
PS1 7513±86.5 60 min 2.09±0.02
PS2 9036±95.1 60 min 2.51±0.03
PS1+PS2 251±15.8 60 min (6.9±0.4)102


Second do the PS1+PS2+n's coincidence test

1. The distance between PS and n's PMT is about 80 cm

counts time counts/sec
PS1 3229±56.8 20 min 2.69±0.05
PS2 3192±56.5 20 min 2.66±0.05
PS1+PS2+n's 59±7.7 20 min (4.9±0.6)102

2. The distance between PS and n's PMT is about 50 cm

counts time counts/sec
PS1 7700±87.7 60 min 2.14±0.02
PS2 9060±95.2 60 min 2.52±0.03
PS1+PS2+n's 214±14.6 60 min (5.9±0.4)102



The scope pictures of coincidence events looks like:

PS1+PS2 event looks like: PS1+PS2+n event looks like:
Cosmic coinc PS1 PS2.png Cosmic coinc PS1 PS2 n.png