We used:
We used:


Line 9: Line 12:
-delay amplifier TENNELEC TC215 to calibrate TDCs in the region > 250 ns;
-delay amplifier TENNELEC TC215 to calibrate TDCs in the region > 250 ns;


Range of all three TDCs was set to 50nsx100.  
Range of all three TDCs was set to '''50nsx100 (upper plot)''' and '''200nsx10 (lower plot)'''.  


ADC-TAC pair was chosen in such a way that we used during the experiment.
ADC-TAC pair was chosen in such a way that we used during the experiment.
Line 16: Line 19:


[[Image:567_02.jpg |500px]]
[[Image:567_02.jpg |500px]]
[[Image:567_02_2000.jpg |500px]]


=ORTEC TAC 567-01 and ADC 1A=
=ORTEC TAC 567-01 and ADC 1A=


[[Image:567_01.jpg |500px]]
[[Image:567_01.jpg |500px]]
[[Image:567_01_2000.jpg |500px]]


=ORTEC TAC 566-01 and  ADC 1C=
=ORTEC TAC 566-01 and  ADC 1C=
Line 25: Line 32:
[[Image:566_01.jpg |500px]]
[[Image:566_01.jpg |500px]]


[[Image:566_01_2000.jpg |500px]]


= Comparison of old and new calibration coefficients =
= Comparison of old and new calibration coefficients =


[[Image:Old_calibration.jpg | 300px ]]
We used ADC 1D for the run pf_10_10_08_00040
 
[[Image:Old_calibration.jpg |500 px]]
 
The width of photon peak is about '''10.3 ns''' for old calibration.
 
[[Image:New_calibration.jpg |500 px]]
 
The width of photon peak is about '''32.3 ns''' for new calibration.


[[Image:New_calibration.jpg | 300px ]]
[http://wiki.iac.isu.edu/index.php/PhotoFission_with_Polarized_Photons_from_HRRL Go Back]

Latest revision as of 06:18, 5 February 2009

Go Back


We used:

-detector A to calibrate all the TDCs;

-LeCroy octal discriminator to produce standard negative pulses;

-passive delay line to calibrate TDCs in the region < 100 ns;

-delay amplifier TENNELEC TC215 to calibrate TDCs in the region > 250 ns;

Range of all three TDCs was set to 50nsx100 (upper plot) and 200nsx10 (lower plot).

ADC-TAC pair was chosen in such a way that we used during the experiment.

ORTEC TAC 567-02 and ADC 1D

ORTEC TAC 567-01 and ADC 1A

ORTEC TAC 566-01 and ADC 1C

Comparison of old and new calibration coefficients

We used ADC 1D for the run pf_10_10_08_00040

The width of photon peak is about 10.3 ns for old calibration.

The width of photon peak is about 32.3 ns for new calibration.

Go Back