Line 201: Line 201:
 
Slope: <math> 6.60047 \pm 0.0179356 </math>
 
Slope: <math> 6.60047 \pm 0.0179356 </math>
  
which gives a detection limit of <math> 0.016 \pm </math>
+
which gives a detection limit of <math> 0.016 \pm 0.008 </math>
  
 
This is about a factor of 10 worse than before. It could be due to the errors in x pulling the fit upward. The 10% error from shifting the concentration makes the uncertainty higher
 
This is about a factor of 10 worse than before. It could be due to the errors in x pulling the fit upward. The 10% error from shifting the concentration makes the uncertainty higher
 +
 +
Let's try fixing the y-intercept instead,
 +
 +
[[File:A Mix Div a Pure NormCorrected Se75 YintFitRestrict XYErrors.png|200px]]
 +
 +
The new parameters are
 +
 +
Intercept: <math> 1.23 \times 10^{-2} \pm 2.82041 \times 10^{-6} </math>
 +
 +
Slope: <math> 1.52756 \times 10^{-1} \pm 5.91359 \times 10^{-3} </math>
 +
 +
So shift the concentrations over by 0.081 +/- 0.003
 +
 +
[[File:N Div B ConcCorrectedbyFix Se75 XYErrors YintFix.png|200px]]
 +
 +
 +
After the concentrations have been shifted, the fit parameters are
 +
 +
Intercept: <math> 1.83184 \pm 0.0200626 </math>
 +
 +
Slope: <math> 7.50744 \pm 0.0148123 </math>
 +
 +
which gives a detection limit of (0.005 +/- 0.003)%
  
 
==X and Y Errors (Se-81m)==
 
==X and Y Errors (Se-81m)==
Line 236: Line 259:
 
[[File:N Div B ConcCorrectedbyFix Se81m XYErrors.png|200px]]
 
[[File:N Div B ConcCorrectedbyFix Se81m XYErrors.png|200px]]
  
Which gives a detection limit of (0.13 +/- )%
+
Which gives a detection limit of (0.13 +/- 0.06 )%
 +
 
 +
Let's try fixing the y-intercept instead of the slope
 +
 
 +
[[File:A Mix Div a Pure NormCorrected Se81m Yint FitRestrict XYErrors.png|200px]]
 +
 
 +
The fit parameters are
 +
 
 +
Intercept: <math> 2.00000 \times 10^{-3} \pm 8.38623 \times 10^{-5} </math>
 +
 
 +
Slope: <math> 2.02521 \times 10^{-2} \pm 6.43888 \times 10^{-4} </math>
 +
 
 +
So shift the concentrations over by
 +
 
 +
<math> 0.099 \pm 0.005 </math>
 +
 
 +
[[File:N Div B ConcCorrectedbyFix Se81m XYErrors YintFix.png|200px]]
 +
 
 +
The new fit parameters are
 +
 
 +
Intercept: <math> 1.33288 \pm 0.0359407 </math>
 +
 
 +
Slope: <math> 0.543333 \pm 0.0116082 </math>
 +
 
 +
These parameters give a detection limit of (0.13 +/- 0.06 )%

Latest revision as of 16:40, 6 August 2018

50% Excluded

For this analysis, begin by using the first measurement of the Se-75 line and using the standard exponential decay equation to correct it and its error back to the time of beam off. Below is a table of the Se-75 (120d) corrected measurements as well as the front nickel foils and the Mn-54 data for each measurement. Note the Mn-54 analysis was weighted by the mass of the soil.

Sample Activated (total beam time) Counted Fully Thesis Corrected Rate @ 1st Measurement (Hz) [math] A_{Beam Off} [/math]
10% Se/Soil Mixture 5/23/17 and 5/24/17 (83 min) 07/18/17 (16:25:44) 3109.68 [math] \pm [/math] 95.39 4280.08 [math] \pm [/math] 131.29
10% Se/Soil Pure Witness selenium 5/23/17 and 5/24/17 (83 min) 8/1/17 (16:05:51) 4161.29 [math] \pm [/math] 127.63 6210.50 [math] \pm [/math] 190.48
10% Se/Soil Front Inner Ni Foil 5/23/17 (1h) 5/23/17 (15:52:21) 385955 [math] \pm [/math] 23488.8 425064.36 [math] \pm [/math] 25868.95
10% Se/Soil Mixture Mn-54 5/23/17 (1h) 07/18/17 (16:25:44) 15.63 [math] \pm [/math] 0.56 17.67 [math] \pm [/math] 0.63
1% Sample Se/Soil Mixture 5/23/17 (1h) 6/23/17 (16:39:03) 2233.33 [math] \pm [/math] 68.52 2675.24 [math] \pm [/math] 82.08
1% Sample Se/Soil Pure Witness selenium 5/23/17 (1h) 7/20/17 (16:39:46) 3840.86 [math] \pm [/math] 117.81 5379.43 [math] \pm [/math] 165.00
1% Sample Se/Soil Front Inner Ni Foil 5/23/17 (1h) 5/23/17 (17:59:39) 261461 [math] \pm [/math] 16045.7 293170.78 [math] \pm [/math] 17991.71
1% Sample Se/Soil Mixture Mn-54 5/23/17 (1h) 6/23/17 (16:39:03) 9.38 [math] \pm [/math] 0.39 10.05 [math] \pm [/math] 0.42
0.1% Sample Se/Soil Mixture 5/23/17 (30 min) 6/19/17 (18:25:18) 849.46 [math] \pm [/math] 26.07 993.59 [math] \pm [/math] 30.49
0.1% Sample Se/Soil Pure Witness Selenium 5/23/17 (30 min) 07/25/17 (16:32:52) 834.41 [math] \pm [/math] 25.61 1201.58 [math] \pm [/math] 36.88
0.1% Sample Se/Soil Front Inner Ni Foil 5/23/17 (30 min) 5/24/17 (14:25:39 118412 [math] \pm [/math] 7280.06 197670.18 [math] \pm [/math] 12152.98
0.1% Sample Se/Soil Mn-54 5/23/17 (30 min) 6/19/17 (18:25:18) 9.38 [math] \pm [/math] 0.33 9.96 [math] \pm [/math] 0.35

Mn-54 Efficiency

A calibrated Mn-54 source was used to find the efficiency of an 834 keV line. The source was serial #J4-348, which had an activity of 9.882 [math] \mu Ci [/math] on 8/01/12, so the activity on 4/18/17 was

[math] A(4/18/17) = 9.882 \times e^{-2.57 \times 10^{-8} * 148694400} = 0.22 \mu Ci [/math]

Now converting to Hz gives

[math]\left (0.999760 \right )\left ( 0.22 \times 10^{-6} \mbox{Ci} \right) \left (\frac{ (3.7 \times 10^{10} \mbox{Hz}}{\mbox{Ci}} \right)= (8138.05 \pm 3\%) Hz [/math]

Source Energy (keV) Position Expected Rate (Hz) ROOT Window (keV) HpGe Rate (Hz) HpGe Detector A Efficiency (%)
Mn-54 (J4-348) 834 10cm (Det A) 8138.05 [math] \pm [/math] 3\% [829,839] 25.87 [math] \pm [/math] 0.28 0.32 [math] \pm [/math] 0.01

Activity Ratio Plots (Se-75)

Below are plots of the activity ratio of the Se/Soil mixture and specific activity of the pure selenium pellet as a reference material. The plots also have no restrictions on the equation of the line.

A Mix Div a Pure NormCorrected Se75 NoFitRestrict.png


Following the analysis in Nate's thesis, the initial concentration should be the intercept on the x-axis, which is

[math] \frac{0.054558 \pm 0.004056}{0.122961 \pm 0.006568} = (0.44 \pm 0.04)\% [/math]

This is much more physical than previous answers, but I was able to detect Se-75 at a 0.1% level. Let's try some other reference materials

Below is a plot where the activity ratio was taken using Ni-57 as a reference material. This was the front inner Nickel foil

A Mix Div Ni NormCorrected Se75 NoFitRestrict.png

This is not very physical because the graph implies that the initial concentration in the soil was 19.18%, but there were no Se-75 lines observed in a pure soil sample, so this cannot be true

Finally let's try Mn-54 as a reference material as it was internal to the sample.

A Mix Div a Mn NormCorrected Se75 NoFitRestrict.png

Now find the initial concentration by finding the x-intercept.

[math] \frac{15.320237 \pm 0.742753}{16.456772 \pm 0.694779} = (0.93 \pm 0.06)\% [/math]

This is still better than the nickel, but still not physical because of the argument presented in the section about the pure selenium ratio.

Now try to fix the linear fit's parameters such that the x-intercept is less than 0.1%. To do this, fix the y-intercept and the slope separately to see if their intercepts agree. If y = A+Bx, when y = 0, then

[math] x = \frac{-A}{B} \gt -0.1 [/math]

The fitting function for the plot with the pure selenium used as a reference material was y(x) = (0.054558 +/- 0.004056) + (0.122961 +/- 0.006568)*x, so by fixing the y-intercept, we have

[math] x=\frac{-0.054558}{B} \gt -0.1, then B\gt 0.54558 [/math]

Below is a plot where the red line is the fixed and bounded fit


A Mix Div a Pure NormCorrected Se75 FitRestrict.png

Even though the y-value was fixed, I kept the error so the fit equation is y(x) = (0.0545580 +/- 0.004056) + (0.545580 +/- 0.0000414)*x, which yields an x intercept of 0.1 +/-0.01 (note that for all plots it seems by bounding the fit the x-intercept will be as close as possible to the closest value before restrictions, so all fixed fits give x-intercepts of 0.1%

Now shift the concentration values by 0.1%, and plot the signal to noise ratio at the highest measured value.

N Div B ConcCorrectedbyFix Se75.png

Now we still have a non-zero y-intercept, but by increasing it by 2 standard deviations gives a value of 0.195189, now finding the value of x for this, we get

[math] x = \frac{Y_{2 \sigma}-A}{B} = (0.001 +/- 0.0004)\% [/math]

where the error is determined by

[math] \sigma_x = \sqrt{(\frac{1}{B})^2 \sigma_A^2 + (\frac{A-Y_{2 \sigma}}{B^2})^2 \sigma_B^2} [/math]

Where [math] Y_{2 \sigma} [/math] is the y intercept increased by 2 standard deviations

Activity Ratio Plots (Se-81m)

Let's repeat the above analysis with Se-81m data. So begin by plotting the ratio [math] \frac{A_{Mix}}{a_{Pure}} [/math] vs the concentration (omitting the 50% sample)

A Mix Div a Pure NormCorrected Se81m NoFitRestrict.png

The equation of the fit gives an initial concentration of (1.75 +/- 0.17)%, but I want to force the intercept to be greater than -0.1%, so force the fit to have the desired intercept

A Mix Div a Pure NormCorrected Se81m FitRestrict.png

Keeping the error in the y-intercept, the equation of the line is

[math] y(x) = (0.0259648 \pm 0.00205475) + (0.259640 \pm 0.0000008)x [/math]

which yields an x-intercept of 0.10000%, so now shift the SNR plot over by 0.1%, and use the values of the highest SNR

N Div B ConcCorrectedbyFix Se81m.png

now

[math] x = \frac{Y_{2 \sigma}-A}{B} = (0.13 \pm 0.06)\% [/math]


Another Method (Se-75)

For this method, I will use the [math] \frac{A_{Mix}}{a_{Pure}} [/math] data and force the activity to be 0 when there is 0% Se in the soil (This was the original goal from gathering the dirt from a national forest) so use the uncorrected concentrations to find how much Se/Soil was "in the dirt to begin with" and force shift the concentration by that amount to force an intercept at (0,0). First, plot the activity ratio as a function of concentration for the previously mentioned ratio. Note the denominator is kept constant, so this plot is measuring how the activity of the mixture behaves.

A Mix Div a Pure NormCorrected Se75 NoFitRestrict.png

Now we have an initial concentration of 0.44%, so shift the concentration by this number

A Mix Div a Pure NormCorrected Se75 SpikantAdded.png

The error in the intercept is high, but it is consistent with zero, so lets increase the intercept by [math] 2 \sigma [/math] to get

[math] Y_{2 \sigma} = 0.012289 [/math]

Now find the corresponding x point

[math] x = \frac{Y_{2 \sigma} - A}{B} = (0.096 \pm 0.05)\% [/math]

This has high error probably due to the uncertainty in the intercept.

Another Method (Se-81m)

Let's repeat this for Se-81m. The plot below will give the initial concentration of Se-81m in the soil.

A Mix Div a Pure NormCorrected Se81m NoFitRestrict.png

The x-intercept is (1.75 +/- 0.17)%, so shift the concentration by this amount.

A Mix Div a Pure NormCorrected Se81m SpikantAdded.png

Now increase the y-intercept by [math] 2 \sigma [/math] to get 0.006213, now find the corresponding x-value

[math] x = \frac{Y_{2 \sigma}-A}{B} = (0.42 \pm 0.21)\% [/math]

Again this error is high. It would seem plotting the SNR vs. Concentration + Spikant works better


X and Y errors (Se-75)

Now lets repeat the method that yielded the best result, but with x and y errors present. I will be using a TGraphErrors() to plot the data, and fitting with the fit panel. Below is a plot of the activity ratio vs. concentration, but note the errors in x are very small (on the order of 0.05%).

A Mix Div a Pure NormCorrected Se75 NoFitRestrict XYErrors.png

The fit parameters are

Intercept = [math] 5.45571 \times 10^{-2} \pm 0.405627 \times 10^{-2} [/math]

Slope = [math] 1.22960 \times 10^{-1} \pm 0.0656862 \times 10^{-1} [/math]

The intercept parameter here is slightly different than the fit without the error bars in x. Now find the "initial concentration"

[math] x = \frac{-A}{B} = - \frac{ 5.45571 \times 10^{-2} \pm 0.405627 \times 10^{-2} }{1.22960 \times 10^{-1} \pm 0.0656862 \times 10^{-1}} = (-0.44 \pm 0.04)\% [/math]

This is the same as the above analysis, but I will continue and fix the fit as described above.

A Mix Div a Pure NormCorrected Se75 FitRestrict XYErrors.png

The fit parameters for the fixed fit are

Intercept: [math] 5.45571 \times 10^{-2} \pm 0.405627 \times 10^{-2} [/math]

Slope : [math] 5.45500 \times 10^{-1} \pm 4.14090 \times 10^{-5} [/math]

Now the "initial concentration" is

[math] x = \frac{-A}{B} = (0.10 \pm 0.01)\% [/math]

Now propagate the errors through when shifting the concentration on the signal to noise ratio plot to get accurate x errors.

N Div B ConcCorrectedbyFix Se75 XYErrors.png

The fit parameters are

Intercept: [math] 3.17916 \pm 0.0520972 [/math]

Slope: [math] 6.60047 \pm 0.0179356 [/math]

which gives a detection limit of [math] 0.016 \pm 0.008 [/math]

This is about a factor of 10 worse than before. It could be due to the errors in x pulling the fit upward. The 10% error from shifting the concentration makes the uncertainty higher

Let's try fixing the y-intercept instead,

A Mix Div a Pure NormCorrected Se75 YintFitRestrict XYErrors.png

The new parameters are

Intercept: [math] 1.23 \times 10^{-2} \pm 2.82041 \times 10^{-6} [/math]

Slope: [math] 1.52756 \times 10^{-1} \pm 5.91359 \times 10^{-3} [/math]

So shift the concentrations over by 0.081 +/- 0.003

N Div B ConcCorrectedbyFix Se75 XYErrors YintFix.png


After the concentrations have been shifted, the fit parameters are

Intercept: [math] 1.83184 \pm 0.0200626 [/math]

Slope: [math] 7.50744 \pm 0.0148123 [/math]

which gives a detection limit of (0.005 +/- 0.003)%

X and Y Errors (Se-81m)

Let's repeat the above analysis for Se-81m. First plot the activity ratio as a function of the concentration of Se added to find the "initial concentration"

A Mix Div a Pure NormCorrected Se81m NoFitRestrict XYErrors.png

The fit parameters are

Intercept: [math] 2.59648 \times 10^{-2} \pm 2.05476 \times 10^{-3} [/math]

Slope: [math] 1.48111 \times 10^{-2} \pm 7.95127 \times 10^{-4} [/math]

which gives an initial concentration of (1.75 +/- 0.17)%. This is the same as the previous answer, so fix the x-intercept to be less than 0.1%

A Mix Div a Pure NormCorrected Se81m FitRestrict XYErrors.png

The equation of the line is

[math] y(x) = (2.59648 \times 10^{-2} \pm 2.05476 \times 10^{-3}) + (2.59630 \times 10^{-1} \pm 8.52899 \times 10^{-7})x [/math]

Now the new x-intercept is (0.1 +/- 0.01)%, now shift the x values and the errors.


The fit parameters become

Intercept: [math] 1.33232 \pm 0.0363379 [/math]

Slope: [math] 0.543336 \pm 0.0116317 [/math]

N Div B ConcCorrectedbyFix Se81m XYErrors.png

Which gives a detection limit of (0.13 +/- 0.06 )%

Let's try fixing the y-intercept instead of the slope

A Mix Div a Pure NormCorrected Se81m Yint FitRestrict XYErrors.png

The fit parameters are

Intercept: [math] 2.00000 \times 10^{-3} \pm 8.38623 \times 10^{-5} [/math]

Slope: [math] 2.02521 \times 10^{-2} \pm 6.43888 \times 10^{-4} [/math]

So shift the concentrations over by

[math] 0.099 \pm 0.005 [/math]

N Div B ConcCorrectedbyFix Se81m XYErrors YintFix.png

The new fit parameters are

Intercept: [math] 1.33288 \pm 0.0359407 [/math]

Slope: [math] 0.543333 \pm 0.0116082 [/math]

These parameters give a detection limit of (0.13 +/- 0.06 )%