Difference between revisions of "Linac Run Plan April 2018, Dr. McNulty"
Jump to navigation
Jump to search
(49 intermediate revisions by the same user not shown) | |||
Line 1: | Line 1: | ||
=Beam Line info= | =Beam Line info= | ||
− | 0 | + | Machine: 25b Linac |
+ | |||
+ | 0<math>°</math> port | ||
− | + | Beam Energy: 8 MeV | |
− | + | Rep Rate: Max (150Hz?) | |
− | + | Pulse Width: 100ns | |
=Run Plan= | =Run Plan= | ||
− | + | ==To find operating point== | |
- Will use 15 OSLs from reproducibility study to make sure that dose is within acceptable range | - Will use 15 OSLs from reproducibility study to make sure that dose is within acceptable range | ||
− | + | ===Table=== | |
− | + | '''Start at 7am''' | |
− | - | + | :{| border="2" style="text-align:center;" |cellpadding="20" cellspacing="0" |
+ | ! scope="col" style="width: 50px;" | Shot # | ||
+ | ! scope="col" style="width: 150px;" | Number of Pulses | ||
+ | ! scope="col" style="width: 150px;" | Number of OSLs | ||
+ | ! scope="col" style="width: 150px;" | Beam Current | ||
+ | ! scope="col" style="width: 150px;" | Dose/pulse | ||
+ | ! scope="col" style="width: 150px;" | Expected Dose/pulse | ||
+ | |- | ||
+ | | 1 || 80 || 1 ||50 mA || || | ||
+ | |- | ||
+ | | 2 || || 1 || mA || || | ||
+ | |- | ||
+ | | 3 || || 1 || mA || || | ||
+ | |- | ||
+ | | 4 || || 1 || mA || || | ||
+ | |- | ||
+ | | 5 || || 1 || mA || || | ||
+ | |- | ||
+ | | 6 || || 1 || mA || || | ||
+ | |- | ||
+ | | 7 || || 1 || mA || || | ||
+ | |- | ||
+ | | 8 || || 1 || mA || || | ||
+ | |- | ||
+ | | 9 || || 1 || mA || || | ||
+ | |- | ||
+ | | 10 || || 1 || mA || || | ||
+ | |- | ||
+ | | 11 || || 1 || mA || || | ||
+ | |- | ||
+ | | 12 || || 1 || mA|| || | ||
+ | |- | ||
+ | | 13 || || 1 || mA|| || | ||
+ | |- | ||
+ | | 14 || || 1 || mA|| || | ||
+ | |- | ||
+ | | 15 || || 1 || mA|| || | ||
+ | |} | ||
− | + | ==Experiment== | |
− | + | 85 Nanodot OSLs for use (dose not include OSLs from reproducibility study) | |
− | + | - upper limit of OSLs is 15Gy, want to be around 7.5Gy with multiple pulses | |
Will talk to engineers about double pulsing and guaranteeing number of pulses and update. | Will talk to engineers about double pulsing and guaranteeing number of pulses and update. | ||
+ | |||
+ | ===Table=== | ||
:{| border="2" style="text-align:center;" |cellpadding="20" cellspacing="0 | :{| border="2" style="text-align:center;" |cellpadding="20" cellspacing="0 | ||
|- | |- | ||
− | | Shot # | + | ! scope="col" style="width: 50px;" | Shot # |
+ | ! scope="col" style="width: 120px;" | Number of Pulses | ||
+ | ! scope="col"| Number of OSLs | ||
+ | ! scope="col" style="width: 150px;" | Beam Current | ||
+ | ! scope="col" style="width: 150px;" | Aluminum Brick | ||
+ | ! scope="col" style="width: 150px;" | Dose/Pulse | ||
+ | ! scope="col" style="width: 150px;" | Expected Dose/Pulse | ||
|- | |- | ||
− | | 1 || | + | | 1 || || 1 || mA || Out || || |
|- | |- | ||
− | | 2 || | + | | 2 || || 1 || mA || Out || || |
|- | |- | ||
− | | 3 || | + | | 3|| || 1 || mA || Out || || |
|- | |- | ||
− | | 4 || | + | | 4 || || 1 || mA || Out || || |
|- | |- | ||
− | | 5 || | + | | 5 || || 1 || mA || Out || || |
|- | |- | ||
− | | 6 || | + | | 6 || || 1 || mA || Out || || |
|- | |- | ||
− | | 7 || | + | | 7 || || 16 || mA || In || || |
|- | |- | ||
− | | 8 || | + | | 8 || || 16 || mA || In || || |
|- | |- | ||
− | | 9 || | + | | 9 || || 16 || mA || Out || || |
|- | |- | ||
− | | 10 || | + | | 10 || || 16 || mA || Out || || |
|- | |- | ||
− | | 11 || | + | | 11 || || 16 || mA || Out || || |
|- | |- | ||
|} | |} | ||
− | |||
− | |||
− | |||
=[[Absorbed Dose Information]]= | =[[Absorbed Dose Information]]= | ||
− | == | + | Titanium window = 0.5mm thick. Radius = 23.813mm |
− | + | Air gap between Titanium window and glass slide = 45mm | |
− | + | Glass slide Thickness = 1mm | |
− | + | ==[[5mA, 100ns pulse width, 100cm from beam pipe with Titanium window]]== | |
− | + | ==[[25mA, 100ns pulse width, 100cm from beam pipe, with Titanium window]]== | |
− | + | ==[[10mA, 100ns pulse width, 100cm from beam pipe, with Titanium window]]== | |
− | == | + | ==[[100mA, 100ns pulse width, 100cm from beam pipe, with Titanium window]]== |
− | + | ==[[25mA, 100ns pulse width, 50cm from beam pipe, tungsten target with aluminum beamstop]]== | |
− | + | ---- | |
− | + | ==[[100mA, 100ns pulse width, 25cm from beam pipe]]== | |
− | + | ==[[25mA, 100ns pulse width, 25cm from beam pipe]]== | |
− | + | ==[[25mA, 100ns pulse width, 50cm from beam pipe]]== | |
− | |||
− | + | ---- | |
− | + | Low Dose Line: Dose = (0.11 +/- 0.01)(PMT Counts) + (119.29 +/- 29.08) | |
− | = | + | High Dose Line: Dose = (1.54 +/- 0.06)(PMT Counts) + (1004.80 +/- 1006.24) |
− | |||
− | + | Peak Current: <math>50\frac{mA}{pulse}</math> | |
− | + | Pulse Width: <math>500ns</math> | |
− | |||
− | + | <math>50\frac{mA}{pulse}=50\frac{mC}{s*pulse}=0.050\frac{C}{s*pulse}</math> | |
− | + | <math>0.050\frac{C}{s*pulse}(500ns)=2.5*10^{-8}\frac{C}{pulse}</math> | |
− | + | <math>2.5*10^{-8}\frac{C}{pulse}*\frac{1\ e-}{1.602*10^{-19}}=1.56055*10^{11}\frac{e-}{pulse}</math> | |
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
Line 246: | Line 154: | ||
---- | ---- | ||
[[Thesis]] | [[Thesis]] | ||
+ | |||
+ | [[May 31st 2018 - 25b 0 degree port]] |
Latest revision as of 01:41, 27 June 2018
Beam Line info
Machine: 25b Linac
0
portBeam Energy: 8 MeV
Rep Rate: Max (150Hz?)
Pulse Width: 100ns
Run Plan
To find operating point
- Will use 15 OSLs from reproducibility study to make sure that dose is within acceptable range
Table
Start at 7am
Shot # Number of Pulses Number of OSLs Beam Current Dose/pulse Expected Dose/pulse 1 80 1 50 mA 2 1 mA 3 1 mA 4 1 mA 5 1 mA 6 1 mA 7 1 mA 8 1 mA 9 1 mA 10 1 mA 11 1 mA 12 1 mA 13 1 mA 14 1 mA 15 1 mA
Experiment
85 Nanodot OSLs for use (dose not include OSLs from reproducibility study)
- upper limit of OSLs is 15Gy, want to be around 7.5Gy with multiple pulses
Will talk to engineers about double pulsing and guaranteeing number of pulses and update.
Table
Shot # Number of Pulses Number of OSLs Beam Current Aluminum Brick Dose/Pulse Expected Dose/Pulse 1 1 mA Out 2 1 mA Out 3 1 mA Out 4 1 mA Out 5 1 mA Out 6 1 mA Out 7 16 mA In 8 16 mA In 9 16 mA Out 10 16 mA Out 11 16 mA Out
Absorbed Dose Information
Titanium window = 0.5mm thick. Radius = 23.813mm
Air gap between Titanium window and glass slide = 45mm
Glass slide Thickness = 1mm
5mA, 100ns pulse width, 100cm from beam pipe with Titanium window
25mA, 100ns pulse width, 100cm from beam pipe, with Titanium window
10mA, 100ns pulse width, 100cm from beam pipe, with Titanium window
100mA, 100ns pulse width, 100cm from beam pipe, with Titanium window
25mA, 100ns pulse width, 50cm from beam pipe, tungsten target with aluminum beamstop
100mA, 100ns pulse width, 25cm from beam pipe
25mA, 100ns pulse width, 25cm from beam pipe
25mA, 100ns pulse width, 50cm from beam pipe
Low Dose Line: Dose = (0.11 +/- 0.01)(PMT Counts) + (119.29 +/- 29.08)
High Dose Line: Dose = (1.54 +/- 0.06)(PMT Counts) + (1004.80 +/- 1006.24)
Peak Current:
Pulse Width: